Clinical Chemistry Email Content Delivery
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 20: 1416-1421, 1974;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hellbron, D. C.
Right arrow Articles by Kelly, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hellbron, D. C.
Right arrow Articles by Kelly, D.

Clinical Chemistry, Vol 20, 1416-1421, Copyright © 1974 by the American Association for Clinical Chemistry

Warning Methods for Quality Control in Determinations of Serum Calcium

D. C. Hellbron 1, J. W. Eastman 1, and D. Kelly 1

1 Office of Information Systems and The Department of Clinical Pathology and Laboratory Medicine, University of California, San Francisco, Calif. 94143.

We have examined quantitative warning methods based on the measurement of controls (reference samples) to signal day-to-day displacements in determinations of serum calcium. A warning procedure that uses the mean of two control values (Method 2) was compared to a procedure that tested the two control values individually (Method 1). A simulated baseline shift was used to approximate the mechanism of day-to-day displacement. The two methods compared were calibrated to signal a warning with equal frequency in the absence of shift; then the method that signaled with highest frequency under baseline shift was judged most effective. Some statistical models were evolved for the distribution of serum calcium control values. Theoretical comparisons based on these models showed Method 2 to be more effective on the average under a broad range of conditions. However, in an empirical comparison based on 253 pairs of control values, Methods 1 and 2 were about equally effective.


Key Words: statistical treatment of quality-control data

Submitted on November 29, 1973
Accepted on August 2, 1974