Clinical Chemistry Email Content Delivery
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 20: 188-191, 1974;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yee, H. Y.
Right arrow Articles by Goodwin, J. F.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yee, H. Y.
Right arrow Articles by Goodwin, J. F.

Clinical Chemistry, Vol 20, 188-191, Copyright © 1974 by the American Association for Clinical Chemistry

Simultaneous Determination of Copper and Iron in a Single Aliquot of Serum

Hugh Y. Yee 1 and Jesse F. Goodwin 1

1 Department of Pathology, Hutzel Hospital, 432 E. Hancock, Detroit, Mich. 48201 (H.Y.Y.), and Detroit Health Department, 1151 Taylor, Detroit, Mich. 48202 (J.F.G.).

Copper and iron may be sequentially measured in a single aliquot of a protein-free filtrate by use of bis (1 -piperidylthiocarbonyl) disulfide and disodium 3-(2-pyridyl)-5,6-bis(4-phenylsulfonic acid)-1,2,4-triazine (also known as ferrozine). Recovery studies showed excellent results for both metals in various concentrations. Copper interference has been obviated by use of thiourea in mercaptoacetic acid. Simultaneous equations are not needed in calculating results and high-reading blanks are eliminated. Comparative results are given with methods in which each metal was measured singly.


Key Words: trace elements in serum • ferrozine

Submitted on November 24, 1973
Accepted on November 27, 1973




The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
M. L. Sampson, A. Aubry, G. Csako, and A. T. Remaley
Triple Lipid Screening Test: A Homogeneous Sequential Assay for HDL-Cholesterol, Total Cholesterol, and Triglycerides
Clin. Chem., March 1, 2001; 47(3): 532 - 539.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1974 by the American Association for Clinical Chemistry.