Clinical Chemistry, Vol 20, 188-191, Copyright © 1974 by the American Association for Clinical Chemistry
1 Department of Pathology, Hutzel Hospital, 432 E. Hancock,
Detroit, Mich. 48201 (H.Y.Y.), and Detroit Health Department,
1151 Taylor, Detroit, Mich. 48202 (J.F.G.).
Copper and iron may be sequentially measured in a single aliquot of a protein-free filtrate by use of bis (1 -piperidylthiocarbonyl) disulfide and disodium 3-(2-pyridyl)-5,6-bis(4-phenylsulfonic acid)-1,2,4-triazine (also known as ferrozine). Recovery studies showed excellent results for both metals in various concentrations. Copper interference has been obviated by use of thiourea in mercaptoacetic acid. Simultaneous equations are not needed in calculating results and high-reading blanks are eliminated. Comparative results are given with methods in which each metal was measured singly.
Submitted on November 24, 1973
Accepted on November 27, 1973
The following articles in journals at HighWire Press have cited this article:
![]() |
M. L. Sampson, A. Aubry, G. Csako, and A. T. Remaley Triple Lipid Screening Test: A Homogeneous Sequential Assay for HDL-Cholesterol, Total Cholesterol, and Triglycerides Clin. Chem., March 1, 2001; 47(3): 532 - 539. [Abstract] [Full Text] [PDF] |
||||