Clinical Chemistry, Vol 21, 770-773, Copyright © 1975 by the American Association for Clinical Chemistry
1 Department of Pathology & Laboratory Medicine, Emory University School of Medicine, Atlanta, Ga. 30322.
We report a simplified radioimmunoassay procedure for determination of plasma cortisol. The method offers the advantage that tracer, antibody, and assay buffer are added in a single step with the use of semi-automated equipment. Thus, critical pipetting procedures are minimized, and assay time is reduced to 4 h. No prior extraction or chromatographic purification of cortisol is required. The procedure is simple, reliable, and accurate. For either between-assay or within-assay determinations, the coefficients of variation are about 6%. Our results compare well with those obtained from a procedure that included extraction and chromatographic purification of cortisol in several plasma samples. Conditions for optimizing the assay are also discussed.
Submitted on January 23, 1975