Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 21: 1253-1257, 1975;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Klein, P. D.
Right arrow Articles by Hachey, D. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Klein, P. D.
Right arrow Articles by Hachey, D. L.

Clinical Chemistry, Vol 21, 1253-1257, Copyright © 1975 by the American Association for Clinical Chemistry

Stable Isotope Ratiometer-Multiple Ion Detector Unit for Quantitative and Qualitative Stable Isotope Studies by Gas Chromatography-Mass Spectrometry

Peter D. Klein 1, Joseph R. Haumann 1, and David L. Hachey 1

1 Division of Biological and Medical Research, Argonne National Laboratory, Argonne, Ill. 60439.

We have designed and constructed a stable isotope ratiometer-multipie ion detector unit, which can drive existing gas chromatograph-quadrupole or magnetic sector mass spectrometers to monitor up to six ions in turn. Each of the three pairs of ions can be selected for quantitation; thus three different or successive components can be analyzed in a single gas-chromatographic run. A background subtraction option permits the ion intensity in the absence of sample to be subtracted automatically during sample measurement. Displays of accumulated counts and isotope ratio are updated twice per second during the measurement and can be printed out at its conclusion. All six ions can be monitored in the analog mode by parallel outputs to a multipen recorder. Experience gained in the construction of this prototype indicates that such units could be commercially available for $10 000, or about a third to a sixth of the cost of even an inexpensive computer system.

Submitted on March 4, 1975
Accepted on April 30, 1975