Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 26: 111-114, 1980;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Cham, B. E.
Right arrow Articles by Rowland, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cham, B. E.
Right arrow Articles by Rowland, M.

Clinical Chemistry, Vol 26, 111-114, Copyright © 1980 by American Association for Clinical Chemistry

Simultaneous liquid-chromatographic quantitation of salicylic acid, salicyluric acid, and gentisic acid in urine

BE Cham, F Bochner, DM Imhoff, D Johns and M Rowland

We have developed a specific and sensitive method for the determination of salicylic acid, salicyluric acid, and gentisic acid in urine. Any proteins present are precipitated with methyl cyanide. After centrifugation, an aliquot of the supernate is directly injected into an octadecyl silane reversed-phase chromatographic column, then eluted with a mixture of water, butanol, acetic acid, and sodium sulfate, and quantitated at 313 nm by ultraviolet detection according to peak-height ratios (with internal standard, o-methoxybenzoic acid) or peak heights (no internal standard). The method allows estimates within 25 min. Sensitivity was 0.2 mg/L for gentisic acid, and 0.5 mg/L for both salicyluric and salicylic acid (20-micro L injection volume); response was linear with concentration to at least 2.000 g/L for salicylic acid and metabolites. Analytical recovery of salicylic acid and metabolites from urine is complete. Intra-assay precision (coefficient of variation) is 5.52% at 7.5 mg/L for salicylic acid, 5.01% at 9.33 mg/L for salicyluric acid, and 3.07% at 7.96 mg/L for gentisic acid. Interassay precision is 7.32% at 7.51 mg/L for salicylic acid, 5.52% at 8.58 mg/L for salicyluric acid, and 3.97% at 8.32 mg/L for gentisic acid. We saw no significant interference in urine from patients being treated with various drugs other than aspirin.


The following articles in journals at HighWire Press have cited this article:


Home page
Hum Exp ToxicolHome page
D.K. Patel, A. Hesse, A. Ogunbona, L.J. Notarianni, and P.N. Bennett
Metabolism of Aspirin after Therapeutic and Toxic Doses
Human and Experimental Toxicology, January 1, 1990; 9(3): 131 - 136.
[Abstract] [PDF]


Home page
Hum Exp ToxicolHome page
D.K. Patel, A. Ogunbona, L.J. Notarianni, and P.N. Bennett
Depletion of Plasma Glycine and Effect of Glycine by Mouth on Salicylate Metabolism During Aspirin Overdose
Human and Experimental Toxicology, January 1, 1990; 9(6): 389 - 395.
[Abstract] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1980 by the American Association for Clinical Chemistry.