Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 26: 1677-1682, 1980;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Roda, A.
Right arrow Articles by Barbara, L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Roda, A.
Right arrow Articles by Barbara, L.

Clinical Chemistry, Vol 26, 1677-1682, Copyright © 1980 by American Association for Clinical Chemistry

Results with six "kit" radioimmunoassays for primary bile acids in human serum intercompared

A Roda, E Roda, R Aldini, M Capelli, D Festi, C Sama, G Mazzella, AM Morselli and L Barbara

We examined six radioimmunoassay procedures for measuring primary bile acids in human serum (two 3H-labeled and four 125I-labeled). A significant (p < 0.01) correlation was observed between measurements in the assay both for cholic acid and chenodeoxycholic acid, at low and high concentrations of serum bile acids. All kits were acceptable with respect to accuracy, precision, stability, and analytical recovery. All six procedures gave similar results for chenodeoxycholic and cholic acid in sera of 80 healthy subjects; the agreement was also close when the two primary bile acids were compared with their sum in serum. Normal values ranged from 0.4 to 2.5 mumol/L for conjugated chenodeoxycholic acid and from 0.3 to 1.5 mumol/L for conjugated cholic acid. The 125I assays do not require liquid-scintillation equipment but 125I induces a decrease in the affinity constant of antibody. The sensitivity of the assays was still adequate for measuring bile acids in the serum of healthy fasting persons and liver-disease patients.