Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 26: 1813-1815, 1980;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Berger, R. L.
Right arrow Articles by Sapoff, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Berger, R. L.
Right arrow Articles by Sapoff, M.

Clinical Chemistry, Vol 26, 1813-1815, Copyright © 1980 by American Association for Clinical Chemistry

A digitally linearized thermistor thermometer referenced to IPTS-68

RL Berger, T Clem, C Gibson, W Siwek and M Sapoff

We describe a digitally linearized thermistor thermometer that is accurate to +/- 0.01 degrees C from 0 to 60 degrees C at a wattage of 50 microW and a sampling time of 1 s. It is traceable to the International Practical Temperature Scale of 1968 (IPTS-68). Analog and digital outputs are provided, the latter as ASCII RS-232C. The thermistor is a 10-cm long immersible probe, 1.5 mm in diameter. Only 0.5 cm of the probe need be immersed, with proper heat sinking of at least 1 cm of the remaining 9.5 cm, for accurate readings. Error from self-heating of the thermistor is less than 5X 10(-3) degrees C at all power settings from 50 to 1 microW.





HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1980 by the American Association for Clinical Chemistry.