Clinical Chemistry, Vol 28, 75-78, Copyright © 1982 by American Association for Clinical Chemistry
WH Porter and A Auansakul
We describe a new gas-chromatographic procedure for ethylene glycol in serum, determined as the cyclic phenylboronate ester. The method is rapid, requires only 100 microL of serum, and either an OV-1 or an OV- 17 column is used. The linear dynamic range extends to at least 5000 mg/L, with a lower limit of detection of 10 mg/L. Analytical recovery of ethylene glycol from serum averaged 97% (95-101%) over the concentration range of 250-5000 mg/L. The method demonstrates good precision, with CVs less than 2.5%, over the concentration range 500- 2000 mg/L.
The following articles in journals at HighWire Press have cited this article:
![]() |
J. Brent, K. McMartin, S. Phillips, K. K. Burkhart, J. W. Donovan, M. Wells, K. Kulig, and The Methylpyrazole for Toxic Alcohols Study Group Fomepizole for the Treatment of Ethylene Glycol Poisoning N. Engl. J. Med., March 18, 1999; 340(11): 832 - 838. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Aabakken, K. S. Johansen, E.-B. Rydningen, J. E. Bredesen, S. Øvrebø, and D. Jacobsen Osmolal and Anion Gaps in Patients Admitted to an Emergency Medical Department Human and Experimental Toxicology, January 1, 1994; 13(2): 131 - 134. [Abstract] [PDF] |
||||
![]() |
E. Bodd, D. Jacobsen, E. Lund, Ås. Ripel, Jør. Mørland, and E. Wiik-Larsen Morphine-6-Glucuronide might Mediate the Prolonged Opioid Effect of Morphine in Acute Renal Failure Human and Experimental Toxicology, January 1, 1990; 9(5): 317 - 321. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |