Clinical Chemistry, Vol 28, 2264-2268, Copyright © 1982 by American Association for Clinical Chemistry
SM Sax, JJ Moore, A Oley, JS Amenta and JA Silverman
The "high-performance" liquid-chromatographic method we report for measuring palmitate derived from saturated phospholipids in amniotic fluid is sensitive and precise and possesses a wide dynamic range. Palmitate from nonsurfactant sources is largely excluded. The procedure involves oxidation of the chloroform/methanol-extracted lipids with osmium tetroxide, precipitation of the resulting phosphatides with cold acetone, mild alkaline hydrolysis, derivatization of the liberated fatty acids to phenacyl esters, and reversed-phase chromatography. The method is no less convenient or more time-consuming than measurement of the lecithin/sphingomyelin ratio by thin-layer chromatography. The clinical usefulness of the procedure is indicated by results of a correlative study of 26 amniotic fluids evaluated by "lung profile" tests in a second laboratory. A value for dipalmitoyl lecithin of 13 mg/L or greater indicates fetal lung maturity.