Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 34: 404-407, 1988;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Peake, M. J.
Right arrow Articles by White, G. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Peake, M. J.
Right arrow Articles by White, G. H.

Clinical Chemistry, Vol 34, 404-407, Copyright © 1988 by American Association for Clinical Chemistry

Laboratory evaluation of the Beckman Synchron CX3 clinical chemistry analyzer

MJ Peake, M Pejakovic and GH White
Department of Biochemistry and Chemical Pathology, Flinders Medical Centre, Bedford Park, South Australia.

In this evaluation of the Beckman Synchron CX3, the multi-analyte clinical chemistry analyzer exhibited high precision, good linearity, and no carryover for each of the eight analytes measured. Results obtained correlated well with those produced by our routine instrumentation (Beckman Astra, Varian atomic absorption spectrophotometer). The instrument can process up to 75 samples per hour (600 tests per hour if all tests available are requested) and, after calibration, can provide urgent results for the complete panel of tests within 2 1/2 min. The performance characteristics of this instrument make it ideal as a routine or a "stat" analyzer for commonly requested tests in the clinical chemistry laboratory.


The following articles in journals at HighWire Press have cited this article:


Home page
Annals of Clinical & Laboratory ScienceHome page
Z.K. Shihabi, R.P. Schwartz, and M.J. Pugia
Decreasing the Variability Observed in Urine Analysis
Ann. Clin. Lab. Sci., January 1, 2001; 31(1): 99 - 102.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1988 by the American Association for Clinical Chemistry.