Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 40: 48-57, 1994;
This Article
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Meyer, E.
Right arrow Articles by De Leenheer, A. P.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Meyer, E.
Right arrow Articles by De Leenheer, A. P.

Clinical Chemistry, Vol 40, 48-57, Copyright © 1994 by American Association for Clinical Chemistry

Simultaneous determination of endogenous retinoic acid isomers and retinol in human plasma by isocratic normal-phase HPLC with ultraviolet detection

E Meyer, WE Lambert and AP De Leenheer
Laboratory of Medical Biochemistry and of Clinical Analysis, Faculty of Pharmaceutical Sciences, University of Ghent, Belgium.

We describe a rapid and simple procedure for the simultaneous quantitation of endogenous 13-cis-retinoic acid, all-trans-retinoic acid, and retinol by isocratic normal-phase HPLC with ultraviolet detection, in 0.5 mL of human plasma. A silica adsorption column was eluted with n-hexane:2-propanol:acetic acid (200:0.7:0.135 by vol) at 0.9 mL/min, and the effluent monitored at 350 nm. The arotinoid ethylsulfonic acid Ro 15-1570 was used as the internal standard. High sensitivity, allowing quantitation of physiological concentrations, was achieved, particularly for the retinoic acid isomers. The detection limits were 0.5 microgram/L in plasma for both 13-cis- and trans- retinoic acid, and 10 micrograms/L for retinol. The CVs for between-day determinations of the lowest quality-control concentration (n = 12) were 4.8% for 13-cis-retinoic acid, 3.4% for trans-retinoic acid, and 3.0% for retinol. The mean (+/- SD) concentrations of 13-cis-retinoic acid (1.79 +/- 0.56 microgram/L), trans-retinoic acid (1.35 +/- 0.42 microgram/L), and retinol (533 +/- 58 micrograms/L) measured in plasma from 22 healthy volunteers agreed well with those previously reported.


The following articles in journals at HighWire Press have cited this article:


Home page
Am. J. Physiol. Endocrinol. Metab.Home page
E. A. Werner and H. F. Deluca
Retinoic acid is detected at relatively high levels in the CNS of adult rats
Am J Physiol Endocrinol Metab, March 1, 2002; 282(3): E672 - E678.
[Abstract] [Full Text] [PDF]


Home page
Proc. Natl. Acad. Sci. USAHome page
J. C. White, V. N. Shankar, M. Highland, M. L. Epstein, H. F. DeLuca, and M. Clagett-Dame
Defects in embryonic hindbrain development and fetal resorption resulting from vitamin A deficiency in the rat are prevented by feeding pharmacological levels of all-trans-retinoic acid
PNAS, November 10, 1998; 95(23): 13459 - 13464.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1994 by the American Association for Clinical Chemistry.