Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 45: 1889-1897, 1999;
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Roberts, T.
Right arrow Articles by Roberts, B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Roberts, T.
Right arrow Articles by Roberts, B.
Related Collections
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 1999;45:1889-1897.)
© 1999 American Association for Clinical Chemistry, Inc.


Articles

Observations on Centrifugation: Application to Centrifuge Development

Thomas Roberts1,a, Mathew Smith1 and Barry Roberts2

1 Laboratory Automation Developments and Services, 141 St. James Rd., Glasgow G4 0LT, Scotland.

2 Biochemistry Department, Gartnavel General Hospital, Glasgow G12 0YN, Scotland.
a Author for correspondence. Fax 44-141-552-3886; e-mail info{at}ladsltd.demon.co.uk

This report outlines the background to the development of an automated, serial, discrete centrifuge, reporting on the criteria considered essential in such an instrument. We established the criteria by examining the detailed logistics of centrifuge operation in a hospital laboratory. The mean sample load per run, using six centrifuges, was 13.6 samples, and the user-selectable cycle time ranged from 00:01:10 to 00:12:33 (hours:minutes:seconds) with a fixed g value of 1050. During the laboratory working window, (0900–1700), only 50% of the centrifuge capacity was utilized and more than one-third of the sample workload was delayed for >5 min because the centrifuges were not emptied promptly. In addition, 35% of the sample workload was centrifuged for less than the time prescribed in the operational specifications. Based on these findings, we designed a new continuous, serial centrifuge to overcome some of the deficiencies noted in the logistics study. The centrifuge operates continuously, nominally treating 150 samples/h, with a cycle time of 5 min at 1000g. The cycle time and g value are variable between limits, and their selection governs the throughput rate. Each sample is centrifuged separately in individual rotors mounted in a sturdy carousel with a periphery that traverses a load/unload station. There is no sample delay because of operator absence, and the capacity is fully utilized. The centrifuge can operate in a stand-alone capacity or has the capability of being integrated into a sample preparation system or as a direct front end for high-throughput analyzers.




The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
J. W. Holman, T. E. Mifflin, R. A. Felder, and L. M. Demers
Evaluation of an Automated Preanalytical Robotic Workstation at Two Academic Health Centers
Clin. Chem., March 1, 2002; 48(3): 540 - 548.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1999 by the American Association for Clinical Chemistry.