Background: To simplify International Sensitivity Index (ISI) calibration, the possibility of substituting fresh plasma for fresh whole-blood samples with point-of-care testing (POCT) whole-blood monitors was investigated in a three-center study of three different POCT systems.
Methods: A modified full WHO calibration procedure based on 20 healthy controls and 60 coumarin-treated patients was performed on three monitoring systems with whole-blood and plasma samples against plasma tested using the European Concerted Action on Anticoagulation (ECAA) rabbit reference plain thromboplastin and the manual prothrombin time (PT) method.
Results: With one of the three systems, the mean ISI was 1.51 for whole blood and 1.49 for plasma; with the second system, the mean ISI was 1.08 for both whole blood and plasma. With the third system, however, the difference between the mean ISI for whole blood and that for plasma was greater (1.15 and 1.01, respectively). Overall, the precision of the calibrations was less than with traditional manual plasma PT testing.
Conclusions: Provided that an appropriate calcium chloride concentration is used, the plasma PT results can be used for accurate ISI calibration of two of these three whole-blood POCT systems. Precision criteria need to be modified for POCT monitors.
The following articles in journals at HighWire Press have cited this article:
![]() |
L. Poller, J. Jespersen, M. Keown, S. Ibrahim, and A. Tripodi The authors of the article cited above respond: Clin. Chem., August 1, 2007; 53(8): 1556 - 1557. [Full Text] [PDF] |
||||
![]() |
L. Poller, M. Keown, N. Chauhan, A. M. van den Besselaar, A. Tripodi, C. Shiach, and J. Jespersen Reliability of international normalised ratios from two point of care test systems: comparison with conventional methods BMJ, July 3, 2003; 327(7405): 30. [Abstract] [Full Text] [PDF] |
||||
![]() |
L Poller, M Keown, N Chauhan, A M H P van den Besselaar, A Tripodi, C Shiach, and J Jespersen European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors J. Clin. Pathol., February 1, 2003; 56(2): 114 - 119. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Poller, M. Keown, N. Chauhan, A. M.H.P. van den Besselaar, A. Tripodi, C. Shiach, and J. Jespersen European Concerted Action on Anticoagulation. Evaluation of a Method for International Sensitivity Index Calibration of Two Point-of-Care Prothrombin Time (PT) Monitoring Systems (CoaguChek Mini and TAS PT-NC) with Fresh Plasmas Based on Whole-Blood Equivalent PT Clin. Chem., October 1, 2002; 48(10): 1672 - 1680. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |