Clinical Chemistry Link to Randox Laboratories Web Site spacer gif
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 0: clinchem.2007.101014v1, 2008; 10.1373/clinchem.2007.101014
This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
clinchem.2007.101014v1
54/6/1060    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Karon, B. S.
Right arrow Articles by Jaffe, A. S.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Karon, B. S.
Right arrow Articles by Jaffe, A. S.

Received on November 26, 2007
Accepted on March 5, 2008

Hemostasis and Thrombosis

Aspirin Responsiveness in Healthy Volunteers Measured with Multiple Assay Platforms

Brad S. Karon 1*, Amy Wockenfus 1, Renee Scott 1, Stacy J. Hartman 1, Joseph P. McConnell 1, Paula J. Santrach 1, Allan S. Jaffe 1

1 Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN

* To whom correspondence should be addressed. E-mail: karon.bradley{at}mayo.edu.

BACKGROUND: We evaluated the sensitivity, precision, and concordance of 4 assays designed to detect aspirin responsiveness or resistance.

METHODS: Twenty-nine healthy laboratory volunteers took 80 mg aspirin for 7 days, and a subset of volunteers took 325 mg aspirin for an additional 7 days. We measured platelet function by light transmission aggregometry with arachidonic acid, PFA-100, and VerifyNow. PFA-100 and VerifyNow assays were performed in duplicate to assess method imprecision. Some volunteers had samples taken within 2–4 h of the final dose of aspirin and again within 20–24 h of the final dose. We measured urinary 11-dehydro-thromboxane B2 at baseline and after 80 or 325 mg aspirin.

RESULTS: No volunteers were nonresponsive to aspirin therapy as measured by the PFA-100. One of 29 participants demonstrated lack of response to aspirin as measured by VerifyNow and urinary 11-dehydro-thromboxane B2; 2 of 29 demonstrated lack of response as measured by light transmission aggregometry. Imprecision was <10% for the PFA-100 and VerifyNow. Concordance was high (>90%) between all assays. Neither aspirin dose (80 vs 325 mg) nor timing between final dose of aspirin and blood draw (2–4 vs 20–24 h) affected any of the assays.

CONCLUSIONS: Light transmission aggregometry, PFA-100, VerifyNow, and urinary 11-dehydro-thromboxane B2 are all sensitive to the effects of aspirin in healthy individuals. Variables such as aspirin dose, timing between final dose of aspirin and blood collection, and imprecision do not affect the ability of the assays to detect aspirin effect on platelet function.