To the Editor:
When glucose reacts with hemoglobin (Hb), the first products are Schiff bases that slowly undergo Amadori rearrangement to produce stable ketoamines (glycohemoglobin). The relative fraction of the ketoamines to total Hb can be measured as hemoglobin A1c (Hb A1c), which constitutes ~60% of the bound glucose (1). The Schiff base (or labile fraction) is unstable and is fairly readily hydrolyzed, especially at acid pH. Effective removal of the labile fraction is essential for the accurate determination of Hb A1c because the labile fraction cannot be separated from the true (ketoamine) fraction by simple ion-exchange procedures and because its concentration varies acutely with the plasma glucose concentration (2). Partial and variable removal of the labile fraction adds to assay imprecision and reduces the correlation (3) between measured Hb A1c and mean plasma glucose.
The Bio-Rad Labs. Variant analyzer is a dedicated HPLC system designed
to measure variant hemoglobins or Hb A1c. For Hb
A1c the hemolyzing reagent is a pH 5.0 citrate buffer that
also serves to remove the labile fraction. The Bio-Rad instrument
manual advises users to "let samples stand at 1828 °C for at
least 10 min to allow Schiff base removal" (this was increased
recently to 15 min). The manual also describes how to run the
instrument essentially in batch mode. At start-up, or on restarting
after running samples, the recommended protocol is an 8-min wash, a
3-min prime with hemolysate, followed by 3-min runs with two
calibrators and two controls (total 23 min) before patients' samples
are processed. With this protocol, patients' hemolysates are tested
after variable periods of incubation at room temperature and at 8 °C
(the temperature of
References
Bio-Rad Labs., Hercules, CA
a Address correspondence to this author, at: 17 Runnymede Rd., Toronto, ON, Canada, M6S 2Y1.
To the Editor:
a Address correspondence to this author, at: 17 Runnymede Rd., Toronto, ON, Canada, M6S 2Y1.
To the Editor:
The following articles in journals at HighWire Press have cited this article:
![]() |
M. P. Olmsted, P. A. Colton, D. Daneman, A. C. Rydall, and G. M. Rodin Prediction of the Onset of Disturbed Eating Behavior in Adolescent Girls With Type 1 Diabetes Diabetes Care, October 1, 2008; 31(10): 1978 - 1982. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Colton, M. Olmsted, D. Daneman, A. Rydall, and G. Rodin Disturbed Eating Behavior and Eating Disorders in Preteen and Early Teenage Girls With Type 1 Diabetes: A case-controlled study Diabetes Care, July 1, 2004; 27(7): 1654 - 1659. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Gibb, A. Parnham, M. Fonfrede, and F. Lecock Multicenter Evaluation of Tosoh Glycohemoglobin Analyzer Clin. Chem., October 1, 1999; 45(10): 1833 - 1841. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |