Technical Briefs |
1
Lab. Univ. de Phys. Biol., URA CNRS 1173, Service de Méd. Interne, CHRU, Strasbourg, France;
2
Service Central de Méd. Nucléaire, Clin. Marc Linquette, CHRU, Lille, France;
3
Boehringer Mannheim, Tutzing, Germany;
a address for
correspondence: Inst. de Phys. Biol., Faculté de Méd., F-67085 Strasbourg Cedex, France: fax 33 3 88 14 48 79, e-mail sapin@alsace.u-strasbg.fr
Immunoassays on the fully automated Elecsys® 2010 analyzer (Boehringer Mannheim, Meylan, France) involve the electrochemiluminescent reaction of Ru(II) Tris(bipyridyl) with tripropylamine combined with a magnetic microparticle separation after short incubation times with low sample volumes [1]. The thyrotropin (TSH) assay, which requires a high sensitivity [2], has an 18-min incubation time and 50-µL sample volume. In the present study we assessed the analytical and clinical performances of this TSH assay during 5 weeks in June and July of 1996. Assays were performed in singleton according to the manufacturer's instructions with two different lot numbers. Results were compared with those of the Immulite TSH3G assay (Behring Diagnostic, Rueil Malmaison, France). The procedures were performed in accordance with the Helsinki Declaration of 1975, as revised in 1983.
Interassay reproducibility was assessed by repeated analysis (n =
12) of 12 patients' frozen sera and of the TSH3G Immulite control
serum. CVs were <5% at TSH >0.06 mU/L, 10.4% at TSH of 0.018
mU/L, and 14.9% at TSH of 0.014 mU/L. "Functional sensitivity"
(CV = 20%) was <0.014 mU/L; therefore Elecsys TSH can be
considered
Acknowledgments
References
The following articles in journals at HighWire Press have cited this article:
![]() |
J. Kratzsch, G. M. Fiedler, A. Leichtle, M. Brugel, S. Buchbinder, L. Otto, O. Sabri, G. Matthes, and J. Thiery New Reference Intervals for Thyrotropin and Thyroid Hormones Based on National Academy of Clinical Biochemistry Criteria and Regular Ultrasonography of the Thyroid Clin. Chem., August 1, 2005; 51(8): 1480 - 1486. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Hendriks, W. Kortlandt, and W. M. Verweij Standardized Comparison of Processing Capacity and Efficiency of Five New-Generation Immunoassay Analyzers Clin. Chem., January 1, 2000; 46(1): 105 - 111. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Navarro, T. Palencia, M. R. Bonnin, and P. Rosel Daily BlankCell Procedure in Elecsys 2010 Improves the Reproducibility of Thyrotropin, Free Thyroxine, and Triiodothyronine Clin. Chem., August 1, 1998; 44(8): 1772 - 1773. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |