Clinical Chemistry AACC Online Job Center
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 49: 2094-2096, 2003; 10.1373/clinchem.2003.026336
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (4)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Herzum, I.
Right arrow Articles by Wahl, H. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Herzum, I.
Right arrow Articles by Wahl, H. G.
Related Collections
Right arrow General Clinical Chemistry
Right arrow Proteomics and Protein Markers
(Clinical Chemistry. 2003;49:2094-2096.)
© 2003 American Association for Clinical Chemistry, Inc.


Technical Briefs

Reliability of IFCC Method for Lactate Dehydrogenase Measurement in Lithium-Heparin Plasma Samples

Ileana Herzum1, Robert Bünder1, Harald Renz1 and Hans Günther Wahl1,a

1 Klinikum der Philipps–Universität Marburg, Department of Clinical Chemistry and Molecular Diagnostics, 35033 Marburg, Germany

aauthor for correspondence: fax 49-6421-2865594, e-mail hg.wahl@med.uni-marburg.de

The first 300 words of the full text of this article appear below.

In a recent Technical Brief on the IFCC-recommended method for lactate dehydrogenase (LD) measurement, Bakker et al. (1) questioned the reliability of this method for heparin samples. According to their findings, the IFCC method produced an excessive frequency of significantly discordant duplicate measurements (17.8%; 27 of 152 samples) compared with the formerly applied French Society (SFBC)-recommended method (1.4%; 2 of 140 samples). Neither increased centrifugation time and speed or the use of plasma separators to reduce the plasma contamination with blood cells, nor LD measurements without concurrent measurements of other analytes to eliminate possible interferences lowered the reported high rate of duplicate errors. The problem was also not attributable to the clinical chemistry analyzer because measurements on both Hitachi 911 and 717 analyzers (Roche Diagnostics GmbH) showed results similar to those obtained with the previously used Modular analyzer (Roche Diagnostics GmbH). Only when serum samples or secondary tubes for centrifuged plasma samples were used was there a significantly lower frequency of duplicate errors: 2.6% (3 of 114) for serum and 1.1% (1 of 94) for secondary plasma tubes. The authors pointed out that the differences in buffer composition of the IFCC (pH 9.40 ± 0.05) (2) and SFBC (pH 7.4–7.8) methods influenced the integrity of platelets and erythrocytes. In this case, the presence of cells might play a role in the high frequency of duplicate errors.

We have been measuring LD activity on three Roche Hitachi 917 analyzers according to the recommendations of the IFCC (2) in our laboratory since February 2003. The within-run imprecisions (CV) for a heparin-plasma sample (n = 20) were 0.8% (197.4 U/L) and 0.7% (321.3 U/L), and the between-run CV were 1.4% for the Roche Precipath U (178.3 U/L; n = 28) and 1.1% for the Roche Precinorm U . . . [Full Text of this Article]




The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
A. J. Bakker, A. Bakker, A. Bierma-Ram, J. T. Dijkstra, H. Renting-Wiering, H. Syperda, and A. Zijlstra
Improved Reliability of Measurement of Lactate Dehydrogenase by IFCC Method in Heparin Plasma
Clin. Chem., January 1, 2005; 51(1): 215 - 217.
[Full Text] [PDF]


Home page
Clin. Chem.Home page
R. R. Miles, R. F. Roberts, A. R. Putnam, and W. L. Roberts
Comparison of Serum and Heparinized Plasma Samples for Measurement of Chemistry Analytes
Clin. Chem., September 1, 2004; 50(9): 1704 - 1706.
[Full Text] [PDF]