|
|
||||||||
Articles |
1
Department of Clinical Chemistry, Vejle County Central Hospital, DK-7100 Vejle, Denmark.
2
Department of General Practice, University of Aarhus,
Aarhus, Denmark.
3
Department of Clinical Chemistry, Odense University
Hospital, Denmark.
a Author for correspondence. Fax + 45 75 82 18 14.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The efficacy and safety of a new near-patient test should ideally be demonstrated before the introduction of the test in routine situations in general practice. A guideline for implementation of near-patient tests emphasized that (3): "It is likely that the equipment will be operated by staff who are not trained as analysts and particular importance should be attached to the robustness of the analytical system: i.e., not what its performance is in the best conditions of operation, but rather what will be achieved in the conditions in which it will actually be used." The term robustness is defined as the ability to yield acceptable results of measurements in spite of deviation from details of the measurement procedure (4).
In general practice, patients with infectious diseases constitute a major part of all consultations (5). General practitioners (GPs) use laboratory tests when assessing these patients.1 One such test is for C-reactive protein (CRP), a marker of the acute-phase response (6)(7)(8). Because the plasma concentration of CRP increases rapidly after stimulation and decreases rapidly with a short half-life, CRP can be a very useful tool in diagnosing and monitoring infections and inflammatory diseases (6)(7)(8).
In Denmark, all CRP measurements from primary and secondary healthcare are performed at hospital laboratories. In the service area of Vejle County Central Hospital (Vejle Hospital), GPs use a CRP measurement in one of 20 consultations and request a CRP measurement for one in three blood samples mailed to the laboratory (9).
Now simple and rapid methods for CRP measurement have been developed, suitable as near-patient tests in general practice. Having access to a rapid CRP result while the patient is still in the doctor's office could be valuable, avoiding unnecessary prescription of an antibiotic, by helping the doctor distinguish between viral and bacterial infection (1)(10)(11). The technical performance of some of these tests is well evaluated when used in a laboratory by experienced technicians (12)(13) but only few studies have been carried out with the test in routine situations in general practice (14)(15). These studies have compared the CRP results with the CRP result from a laboratory nearby as the "true" CRP value. But external quality assessments for central laboratories have shown a major interlaboratory variability of CRP measurements (16). Reference laboratories participating in studies evaluating the technical performance of near-patient tests must be able to document their methods being traceable to BCR/CAP/IFCC CRM 470 (international reference preparation) to ensure the analytical quality for the "true" CRP value. Moreover, the robustness of the analytical system in terms of "importance of the analytical experience for the person operating the test" or "time needed for personnel before proper test performance" must be evaluated before implementation (3). None of the previous studies evaluating near-patient tests for CRP measurements, when used in general practice, fulfills these requirements (14)(15).
The aim of this study was to evaluate the technical performance, especially to test the robustness, of a near-patient test for CRP measurements when used in daily routine in general practice, with measurements carried out by technician staff or unskilled laboratory personnel, and compare results with a documented high-quality laboratory method. We aimed to make a clear presentation of the results by use of difference plots, with a 95% prediction interval based on estimated and assumed obtainable analytical imprecision.
| Materials and Methods |
|---|
|
|
|---|
patients
Patients were included during a 2-month period in February and
March 1995. All patients for whom a blood test for CRP was ordered by
the GP for a clinical purpose were included in the study. Of the 909
patients included, 551 (60.6%) were female, with a median age (range)
of 58.5 years (697), and 358 (39.4%) were male, with a median
age of 59.3 years (1098).
crp measurements in gpcs
All blood samples were drawn by venipuncture with the closed
Monovette system (Sarstedt) with heparin as anticoagulant. With a
capillary straw, 25 µL of whole blood for near-patient testing was
removed from the test tube and the tube was then routinely mailed to
the laboratory at Vejle Hospital for CRP measurements.
NycoCard® CRP Whole Blood (Nycomed Pharma) was investigated. The test system is based on an immunometric principle and consists of (a) a liquid for sample dilution and lysis of blood cells, (b) a test card with six holes containing CRP-specific monoclonal antibodies coated to a membrane, (c) a conjugate solution with monoclonal CRP antibodies coupled to small gold particles, and (d) a washing solution.
The CRP measurement was performed by diluting the 25 µL of whole blood in the capillary straw in 1000 µL of dilution liquid. After mixing and 45 s of lysis time, 25 µL of this diluted sample was applied to a test hole on the test card. After allowing the liquid to soak into the membrane, one drop of the conjugate solution was applied and afterwards one drop of the washing solution was applied to remove conjugate solution in surplus. The gold particles coupled to the CRP antibodies in the conjugate give the membrane a purple-reddish color, and the CRP value is measured from the intensity of the color at the membrane. The color is either measured quantitatively with a color densitometer (NycoCard Reader) or semiquantitatively by visual comparison with a reference color chart corresponding to CRP values of 10, 25, 50, 100, and 200 mg/L. In this study, visual reading of the test results was evaluated.
The person performing the test in GPCs was asked to state the results as "best guess" within 5 mg/L from 1050 mg/L, within 10 mg/L from 50100 mg/L, and within 25 mg/L for values >100 mg/L. Possible results were <10, 10, 15, 20, 25, 30, 35, 40, 45, 50, 60, 70, 80, 90, 100, 125, 150, 175, 200, 225, and >225 mg/L.
For estimation of CVs of the test kit, three of the GPCs were asked to make the CRP measurements as duplicate measurements. The GPCs were asked to arrange the testing with two different employees, ensuring the independency of the double measurements. The CV for 117 duplicated measurements was 0.7% (SD = 0.14, mean CRP value = 21.12 mg/L). In a previously published laboratory evaluation of this test, the CV was from 10% to 15% between series for values at 25 mg/L (n = 27) and 130 mg/L (n = 27), when all measurements were performed by 11 experienced technicians and the reading of the color response was measured with a color densitometer (13). Thus, the CV found in the three GPCs is surprisingly low, and it is likely that the readings were not made independently, but rather were a result of a consensus view between two employees. For the further calculations in this study, we have used an approximation for the CV at 15%, as obtainable when the test kit is used in general practice, based on the laboratory evaluation (13). It seems that this is the best that can be expected in general practice.
The test allows the measurements to be carried out in whole blood, but presents the results as serum values assuming a hematocrit of 0.40. CRP is then slightly overestimated when hematocrit is low and slightly underestimated when hematocrit is high (13). The manufacturer recommends to correct for hematocrit when value is >0.55. However, when an infectious patient is seen by a GP, values of hematocrit are only seldom available, and if the GP has to perform a hematocrit every time he orders a bedside CRP, the idea of using bedside tests will disappear. So in daily routine use we do not expect the GP to correct for hematocrit, and as we wanted to evaluate the test kit while used in daily routine in general practice, we have made the same decision.
crp measurements at the laboratory
CRP was determined by turbidimetry with a Hitachi 717 analyzer
with antibodies (cat. no. 67128) and buffers (cat. no. 67179) from
Orion Diagnostica. The calibrator was from Dako (code no. X 0923). The
accuracy of the calibration was checked with the international
reference preparation for immunochemical measurements, BCR/CAP/IFCC CRM
470 (Commission of the European Communities, Bruxelles, Belgium)
(17). The reference preparation has an assigned value of
39.2 mg/L [confidence interval (CI) 95%: 37.341.2], and when
checking the calibration four different times in the study period, we
found 38 mg/L. As control samples we used a pool of patient serum with
a low CRP value and a pool of patient serum with a high CRP value. In
the study period the following means and coefficient of total
analytical variation (CV) for a specified number of control
measurements were found: low pool: mean = 29.7 mg/L and CV =
5% (n = 185), and high pool: mean = 136.1 mg/L and CV =
5% (n = 51). Results were given quantitatively except for CRP
values <10 mg/L.
statistical analysis
CRP results from general practice and from the laboratory were
compared by using four different methodologies:
1) Results as "best guess" from general practice were compared with the quantitative results from the laboratory divided in intervals matching the answer from the general practice.
2) The mean of difference, defined as the mean difference between CRP
measurements performed with the near-patient test in general practice
and CRP measurements performed at the laboratory (near-patient test
result - laboratory results; unit: mg/L), and the standard
deviation for the mean of differences, were calculated. A 95% CI for
the mean of differences was calculated by using the
t-distribution, and a 95% CI for the SD of differences was
calculated by using the
2-distribution. These values
were calculated for all participating GPCs, for five intervals of
laboratory CRP values (<10, 1025, 2650, 51100, and >100 mg/L),
for GPCs with or without technician staff, for GPCs with more or less
than 50 tests performed in the study period, and for test results
performed within the first week or two of the study period compared
with results from the rest of the period. CRP values read as <10 mg/L
in the GPCs and measured <10 mg/L at the laboratory were defined as 9
mg/L in the calculations.
3) Linear regression was used to compare the paired results from our study with results from other studies.
4) The paired results were evaluated with difference plots. A 95%
prediction interval is calculated, expressing the interval within which
we would expect 95% of the data points to be found, on the basis of
knowledge about the analytical variation (CV) of the two tests
(CV2difference =
CV2lab + CV2test).
Knowing the CVs for the two tests, it is possible before actually
performing the practical part of the test evaluation to set up a
prediction interval for which a certain fraction of the difference
points are expected to be distributed. Such a 95% prediction interval
is in contrast to the 95% limits of agreement, described by Bland and
Altman, which is based on calculations of the actually measured
difference between the two methods and thereby describes the 95%
interval for measured differences (18)(19).
The calculation of the 95% prediction interval is based upon the
analytical variation for the laboratory method, CVlab =
5%, and the CV for the near-patient test, CVtest, composed
by the analytical variation CVtest-analytic = 15%,
approximated from the laboratory study (13), and a
variation for the readings, CVtest-read. When using a
semiquantitative test where test results are reported in intervals, we
normally have no specific knowledge about the possible values of the
test within the interval, but we can assume the value to lie anywhere
within the interval, which gives us a rectangular distribution of the
test results (20). The CRP results from general practice
was given within a "best guess" interval (within 5 mg/L in the
interval from 10 to 50 mg/L, within 10 mg/L from 50 to 100 mg/L, and
within 25 mg/L for values >100 mg/L). As example: If a "true
value" of CRP is within the interval of 17.522.5 mg/L, the answer
20 mg/L is considered correct for all CRP values within this interval,
but as we do not know the specific value within the interval, a reading
variation is then introduced. Assuming the rectangular distribution for
this reading variation, a standard deviation can be transformed from
the formula: a/(2 x 3
) where a is
the length of the whole interval (here 5 mg/L) (20).
(Further considerations and calculations of the 95% prediction
interval are described in detail in the Appendix.) The
between-method differences are plotted both with the result for the
laboratory method as abscissa (Fig. 1
A) and with the average results for the two methods as abscissa
(Fig. 1B
), the latter as suggested by Bland and Altman
(18)(19). Further difference plots are
presented logarithm-transformed (ln-transformed) (Fig. 2
).
|
|
ethics
The procedures followed in this study were in accordance with the
second Declaration of Helsinki (amended in 1989), the Danish law on
biomedical research, and the Scientific Ethical Committee system of
October 1, 1992. According to this, for research on existing data
without intervention and with the purpose of technical and medical
quality assurance, an approval by a scientific ethical committee is not
needed. The local scientific ethical committee was informed about the
study. All included patients had a CRP measurement ordered routinely by
their GP, and blood from this routine test was used to evaluate the
quality of the near-patient tests.
| Results |
|---|
|
|
|---|
compared test results
Results from all test samples are shown in Table 1
. Quantitative results from the laboratory are divided into
intervals matching the "best guess" results for the test kit. In
61% (n = 549) the results of both methods were within the same
"best guess" interval and in 82% (n = 736) the results were
within the same interval or in an adjacent interval. Approximately 70%
(n = 634) of all CRP values measured in the laboratory were within
the normal reference interval (
10 mg/L), 20% (n = 176) were
>20 mg/L, and ~10% (n = 92) were high (
50 mg/L).
|
Results from all 13 GPCs are shown in Table 2
. The overall mean of differences (95% CI) between the paired
CRP measurements was -0.3 mg/L (-0.9 to 0.3 mg/L) and the SD of the
differences was 9.6 mg/L. A tendency towards a higher SD of difference
for GPCs with a higher mean value of CRP is seen [r =
0.50 between the SD of difference and the mean value of CRP found in
the GPCs (n = 13)(Table 2
)].
|
Using linear regression to evaluate the overall number of split-sample measurements (n = 898), we found r = 0.94.
In Table 2
the results from all measurements are shown, grouped in five
intervals of CRP values (<10, 1025, 2650, 51100, and >100
mg/L). For all five CRP intervals the mean of differences is within
-4.5 and 0.6 mg/L. The SD of difference depends upon the numerical
value of the CRP results, which means that it is not correct to use a
single value of SD to describe the test.
effect of analytical experience
Cumulated data from the five GPCs with technician staff and from
the eight GPCs with unskilled laboratory personnel are shown in Table 2
. No statistical difference was found between the two groups according
to the mean of difference, but the SD of difference was, with
statistical significance, slightly higher among the nontechnicians than
among technicians.
effect of a learning period
To assess the importance of a learning period before routine use
of the test, results obtained within the first week or two were
compared with test results obtained in the rest of the study period
(Table 2
). No statistical differences were demonstrated between the two
groups comparing the mean of difference. When we compared the SD of
difference, measurements performed in the first weeks of the study
agreed, with statistical significance, a little more precisely than in
the rest of the study period. When we compared results for GPCs with
>50 CRP measurements with GPCs with <50 CRP measurements in the study
period, no statistical differences were demonstrated on the mean of
difference. When we compared the SD of the difference, measurements
performed in GPCs with a low number of CRP measurements agreed, with
statistical significance, slightly better (Table 2
).
difference plot
The measured difference between the two methods (the
between-method difference) is plotted against the results of the
laboratory method in Fig. 1A
and the average of the two measurements in
Fig. 1B
. In both figures, an upper and lower limit for the 95%
prediction interval is shown. The 95% prediction interval becomes
wider for higher CRP values, illustrating the effect of an inconstant
standard deviation, which is also seen in Table 2
. The values are
therefore ln transformed. In Fig. 2A
the differences of ln values are
plotted against the laboratory method and in Fig. 2B
against the
average of the two measurements. In the ln-transformed figures, the
95% prediction interval and the between-method differences are nearly
constant across the range of measurements. Of the data points, 8.2%
(n = 74) are found outside the 95% prediction interval in the
normal difference plot (Fig. 1B
) and 8.8% (n = 79) in the
ln-transformed plot. Both in the normal plot and in the ln-transformed
plot the number of results outside the prediction interval and the
number of results with a high deviation from zero are higher for the
lower values of CRP.
The lines for the upper and lower limits of the 95% prediction
interval are not perfectly linear (Figs. 1
and 2
). This is because the
accumulated CV for the two methods is not constant, but varies from
16% to 21% depending on the value of the CRP (see Appendix
for the CV calculations). In Fig. 1A
, all differences are positive when
the laboratory CRP is 10 mg/L. This is because the lowest possible
result for the near-patient test is <10 mg/L, which is defined as 9
mg/L in the calculations, and then a difference <-1 mg/L is not
possible. In Fig. 1B
this effect is balanced.
| Discussion |
|---|
|
|
|---|
As in other studies from general practice
(9)(14)(15), a majority of
measured CRP values was found within the normal reference interval
(i.e., <10 mg/L). Still, 20% (n = 176) of all test results were
>20 mg/L, and 10% (n = 92) were high (
50 mg/L). The large
sample size makes evaluation of higher values of CRP possible. The
percentage is in accordance with other studies in this field where 21%
and 22% of all CRP measurements were >20 mg/L
(14)(15), but the total number is higher
because of the considerable sample size in the present investigation.
The percentage of higher CRP values (
50 mg/L) in our study population
is nearly the same as in the other studies, indicating some
homogenicity in the study populations and in the threshold to do the
test among GPs (14)(15).
In a laboratory evaluation of the test comprising 230 paired
measurements performed by experienced technicians and evaluated with a
color densitometer, the results were grouped in the following
intervals: 110, 1125, 2650, 51100, 101200, and >200 mg/L
(13). Of the results of both methods, 79.5% were found in
the same groups and 99.6% were found within the same group or an
adjacent group. If we make the same rough grouping from the data shown
in Table 1
, we find 92.6% within the same group and 99.6% within the
same or an adjacent group, indicating that the test performance made in
general practice is as good as obtained by experienced technicians
performing the test in a laboratory.
Studies comparing two analytical methods (procedures) for measuring the same component often report results as either a correlation coefficient or as the number of "false-normal" and "false-elevated" values when compared with a reference method. These two test statistics are easy to use but not always informative. A correlation coefficient measures the relation between two variables (and not necessarily the agreement between the two variables) and is dependent on the range of the measured quantity (18). If testing for significance is made, a high level of significance will nearly always be found, but it would otherwise be a surprise if two methods designed to measure the same quantity were not related (18).
NycoCard CRP is available for whole blood as well as for serum or plasma. In 1991 a Norwegian group tested the NycoCard CRP Serum/Plasma test kit in 194 samples and reported their results as r = 0.85 when the test was operated in 10 primary healthcare centers (14). We found an overall r of 0.94. This difference may be due to fewer analytical steps in the whole-blood test than in the serum test, reflecting that few analytical steps improve the analytical quality. But the difference can also be explained by the fact that more data points for high values of CRP also gives a higher correlation, as we found 27 CRP measurements >100 mg/L compared with 910 CRP measurements in the Norwegian study.
In a study from The Netherlands, NycoCard CRP Whole Blood was evaluated
in general practice comparing 439 samples (15). In 88% of
all measurements, corresponding results were found between a laboratory
method and the test, with CRP results at 25 mg/L as the cutoff value in
general practice and CRP results at 20 mg/L as the cutoff value for the
laboratory. Furthermore, the results were reported as the frequency of
"false normal" and "false elevated," which were 8% and 28%
respectively. The study group concluded: "The reliability of the
NycoCard CRP measurement in whole blood is disappointing. In particular
the 'false elevated' rate is unacceptably high ... ." In
general, it is necessary to choose the same cutoff value for both
methods to avoid misinterpretion of results when a cutoff value is used
as evaluation strategy. Using data from Table 1
for reporting our
results as the frequency of "false normal" and "false
elevated," we found 2.9% and 6.6% respectively with a cutoff value
of 25 mg/L and 1.2% and 4.6% respectively with a more clinically
meaningful cutoff value of 50 mg/L.
When a test evaluation is made by comparing the test results with results from a laboratory, the evaluation is entirely dependent on the analytical quality of the laboratory. External quality assessments among clinical chemical laboratories have demonstrated major interlaboratory variation in CRP measurements. In a Belgian study the interlaboratory CV was between 19% and 23% when mailing two samples for CRP measurements (median values of 17.0 mg/L and 40.8 mg/L) to 345 laboratories (16). So if a laboratory measures CRP too high or too low, it will have a major influence when using these results as "true values" in a test evaluation, especially if the evaluation is made only as a cutoff evaluation. The laboratory must be able to document its analytical results to be traceable to international reference preparations (i.e., BCR/CAP/IFCC CRM 470). In the two mentioned studies evaluating CRP measurement in general practice, no information is given ensuring the traceability of the laboratory results (14)(15).
A near-patient test in general practice will be operated by nurses,
secretary staff, or by GPs themselves, and not so often by technicians.
The robustness of the near-patient test is therefore an important
factor to evaluate (3). When comparing results from GPCs
served by technicians with results from GPCs without technicians, we
found no significant differences in the test quality regarding the bias
(mean of difference) but a slight difference regarding the precision
(SD of the differences) (Table 2
). In the Norwegian study all serum
samples were frozen and retested by one technician at the laboratory,
and r then increased from 0.85 to 0.95 (14).
The Norwegian study group considered this effect to be due to an
elimination of the interpersonal variation as well as a difference in
the analytical experiences between the personnel in general practice
and the technician. Comparing these results with our results, it seems
likely that the difference is mostly due to elimination of the
interpersonal variation.
Another critical demand for a near-patient test is the time needed from
the introduction of the test until the staff members are experienced as
operators. In the study from The Netherlands the staff in general
practice was given 2 weeks to practice CRP measurement before the
evaluation was performed (15). In our study the staff in
general practice was given one introductory visit, but no time for
practicing the test in advance of the study, as we wanted to evaluate
the analytical quality during the learning period. We found no
differences in the test results obtained within the first or second
week compared with test results from the rest of the period, except for
a slightly lower SD of difference in the first week of the study (Table 2
). When comparing clinics with a high and clinics with a low request
frequency (more or less than 50 tests in the 2-month study period), we
found no differences regarding the mean of difference and only a small
difference in the SD in favor of the clinics with a low request
frequency (Table 2
). These results indicate that the use of the test is
not technically demanding and the test can be operated sufficiently
after just one introductory visit.
application of difference plots in test evaluation
Bland and Altman have suggested using a difference plot when
evaluating a new method with a standard method, and they recommend that
the between-method difference be plotted against the average value of
the two methods, thus avoiding a negative correlation between the
differences and the standard method (18)(19).
In Fig. 1A
we have plotted the between-method difference against the
laboratory method. Calculating the expected negative correlation
between the differences and the laboratory method by using the formula
given by Bland and Altman (19), we find r
= -0.19. The correlation between the actually measured differences and
results of the laboratory method was r = -0.16. Making
the same calculations between the differences and the average results
of the two methods, we find an expected r = 0.01 and a
measured r = 0.01. So, plotting the differences against
the laboratory method gives a slightly negative correlation. In our
test evaluation, we compare a bedside method with a known high
variation (CVtest = 1520%) with a laboratory
method with a known very low analytical imprecision (CVlab
= 5%) (see Appendix for CV calculations). If the
between-method differences are plotted with the laboratory method as
the abscissa, the uncertainty because of known analytical variation for
the abscissa will be low compared with a plot with the average of the
two methods as the abscissa. This fact is illustrated in Fig. 3
. In the Figure
, boxes for the 95% analytical uncertainty of
CRP values of 20, 50, 100, and 200 mg/L are shown. The height of the
boxes illustrates the 95% analytical uncertainty for the difference of
zero between the two methods. The width of the boxes illustrates the
95% analytical uncertainty of the abscissa. In Fig. 3A
the laboratory
method is plotted as the abscissa and in Fig. 3B
the average of the two
methods is plotted as the abscissa.
|
In Fig. 3A
and B the 95% analytical uncertainty of the differences is
calculated as:
![]() |
Example: With a true CRP value of 20 mg/L and a measured
between-method difference of zero, the uncertainty of the difference
is:
![]() |
![]() |
Example: With a true CRP value of 20 mg/L, the uncertainty of
the value is:
![]() |
![]() |
![]() |
Example: With a true CRP value of 20 mg/L, the uncertainty of
the value is:
![]() |
![]() |
Bland and Altman suggest making the 95% limits of agreement on the basis of plus or minus 2 SDs of the differences (or more precisely 1.96) with the limits based upon the results of the test evaluation (18)(19). Thus, a 95% limit of agreements will show the interval where 95% of all data points always will be within when the distribution is gaussian. If the agreement between the two tests is high, the interval will be narrow; if the agreement is low, the interval will be wide.
Instead, we have tried to predict the interval within which we would expect 95% of the data points to be found, having knowledge about the variances (CV) of the two test methods. In other words: Before actually performing the practical part of the study, we are predicting a kind of acceptance limits for the test evaluation, according to the known impressions of the two methods obtained during professional testing in the laboratory. Thus, in this study the 95% prediction interval expresses the best obtainable results possible when the test is operated by staff in general practice not trained as analysts.
general application of evaluations of near-patient tests
When evaluating a near-patient test it is important to ensure that
the measurements are performed under circumstances as near daily
routine as possible, and when planning the study to ensure that the
sample size is sufficient for an evaluation of all intervals of test
results. Moreover, the analytical quality at the reference laboratory
is of major importance to the result of the evaluation. Comparison of
semiquantitative methods with quantitative methods gives statistical
problems that demand some reservations, some of which are set out in
this paper. We have tried to fulfill these major assumptions in this
evaluation by:
Making the evaluation in the field of different GPCs, some with technician staff and others with nurses, secretary staff, or GPs themselves operating the tests.
Not introducing a training period, but as in "real life" start performing measurements just after the introductory visit from the manufacturer.
Having a patient population sample size large enough to ensure validity for test results above the normal reference interval.
Having a reference laboratory that has values traceable to an international reference preparation (i.e., BCR/CAP/IFCC CRM 470).
Developing a statisticalgraphical method, making semiquantitative methods comparable with quantitative methods.
NycoCard CRP for whole blood with visual evaluation of test results is as good when used in GPCs as could be expected from laboratory testing in terms of bias and precision as a semiquantitative test for near-patient testing. The technical performance is nearly equal for technician and nontechnician staff, and for clinics with frequent and nonfrequent use of the test kit, indicating that use of the kit is not technically demanding. Before introducing the test in daily routine in general practice, we recommend an evaluation of the clinical effectiveness of introducing the test, as introduction of a new test should be determined by medical needs rather than by availability (21).
| Appendix 1 |
|---|
|
|
|---|
![]() |
CVtest-analytical is the CV for the near patient test, CVtest = 15%, approximated from the laboratory study (13) where results were performed by 11 experienced technicians and the reading of the color response was measured with a color densitometer (see Materials and Methods).
CVtest-read is the CV for the readings. Using
semiquantitative tests, reporting the test results in intervals, we
normally have no specific knowledge about a possible value within an
interval, and we can assume the value to lie anywhere within the
interval, which gives us a rectangular distribution. Results from
general practice were reported as "best guess" (within 5 mg/L from
1050 mg/L, within 10 mg/L from 50100 mg/L, and within 25 mg/L for
values >100 mg/L). Thus, if a "true value" of CRP is within the
interval of 17.522.5 mg/L, the midpoint answer 20 mg/L is considered
correct for all CRP values within this interval, introducing a reading
variation. This reading variation can be estimated by using the formula
describing the distribution of uncertain values within a rectangle
(20):
![]() |
Example 1A.
At a CRP value of 20 mg/L, µ = 20 mg/L and
a = 5 mg/L.:
![]() |
Example 2A.
At a CRP value of 175 mg/L, µ = 175 mg/L
and a = 25 mg/L:
![]() |
Example 1B.
At a CRP value of 20 mg/L:
![]() |
Example 2B.
At a CRP value of 175 mg/L:
![]() |
![]() |
calculation of the 95% prediction interval when the differences
are ln transformed
A ln-transformed CV is calculated as (22):
![]() |
![]() |
![]() |
Example 1c.
At a CRP value of 20 mg/L,
CVtotal = 17%:
![]() |
Example 2c.
At a CRP value of 175 mg/L,
CVtotal = 16%:
![]() |
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
G. Falk and T. Fahey C-reactive protein and community-acquired pneumonia in ambulatory care: systematic review of diagnostic accuracy studies Fam. Pract., December 12, 2008; (2008) cmn095v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Skeie, G. Thue, K. Nerhus, and S. Sandberg Instruments for Self-Monitoring of Blood Glucose: Comparisons of Testing Quality Achieved by Patients and a Technician Clin. Chem., July 1, 2002; 48(7): 994 - 1003. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Soletormos, P. Hyltoft Petersen, and P. Dombernowsky Progression Criteria for Cancer Antigen 15.3 and Carcinoembryonic Antigen in Metastatic Breast Cancer Compared by Computer Simulation of Marker Data Clin. Chem., July 1, 2000; 46(7): 939 - 949. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. S. Dahler-Eriksen, T. Lauritzen, J. F. Lassen, E. D. Lund, and I. Brandslund Near-Patient Test for C-Reactive Protein in General Practice: Assessment of Clinical, Organizational, and Economic Outcomes Clin. Chem., April 1, 1999; 45(4): 478 - 485. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Andersen, P. H. Petersen, O. Blaabjerg, J. Hangaard, and C. Hagen Evaluation of growth hormone assays using ratio plots Clin. Chem., May 1, 1998; 44(5): 1032 - 1038. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Gambino C-Reactive Protein—Undervalued, Underutilized Clin. Chem., November 1, 1997; 43(11): 2017 - 2018. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||