|
|
||||||||
Articles |
1
Department of Materials Science & Engineering, University of Utah, Salt Lake City, UT 84112.
2
Department of Pathology and Associated Regional and
University Pathologists, 500 Chipeta Way, Salt Lake City, UT 84108.
a Author for correspondence. Fax 801-583-2712; e-mail ASHKO{at}ARUP-lab.com
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Analysis of trace elements in biological fluids demands a versatile and reliable technique. The analytical method used must be sensitive, precise, accurate, and relatively fast. Since the first inductively coupled plasma mass spectrometry (ICP-MS) instrumentation was introduced in 1980 (16), this hybrid technique has become an important method for the analysis of trace elements in biological fluids.1 ICP-MS is capable of direct analysis of solution samples with coverage of most elements in the periodic table (17). Moreover, compared with inductively coupled plasma atomic emission spectrometry (ICP-AES), ICP-MS provides much lower detection limits, simpler spectral interpretation, and reliable isotopic analysis (18). Yet, several factors are still of concern in the analysis of trace elements in biological matrices with ICP-MS. Most biological fluids contain large amounts of organic compounds and inorganic salts, which can lead to spectral and nonspectral interferences. Spectral interference occurs because typical ICP-MS instruments use quadrupoles as mass analyzers, which limits the resolution to approximately unit mass. Signals between the analyte and any interferent with a mass larger than the resolution thus cannot be distinguished. Analysis for some elements of biological interest, such as first-row transition metals, as well as As and Se, are thus compromised. Nonspectral interference is a change of signal intensity that cannot be accounted for by recognizable spectral interferences. The interference can be due to either enhancement or suppression. The concentration of the analyte is thus over- or underdetected.
Nonspectral and spectral interferences can seriously affect the analytical performance of ICP-MS. To obtain accurate results for biological fluids, the influences from both interferences must be investigated and eliminated. External calibration and calibrator addition coupled with internal calibration were used to correct for nonspectral interferences. External calibration is based on a set of external calibrators containing elements of interest and internal calibrators in a simple acid. Success in correction of nonspectral interferences depends on the effectiveness of internal calibration. Selection of internal calibrators can be crucial for the accuracy and precision of trace element analysis in biological fluids (19). Calibrator addition is performed by adding increasing quantities of the elements of interest to multiple aliquots of the sample to be analyzed. The calibration set has an identical sample matrix and hence corrects for nonspectral interferences. Internal calibration is preferably combined with calibrator additions to correct for instrumental drift. To eliminate spectral interferences, a careful selection of the analyte isotope is of prime importance. By correctly selecting an isotope with minimal interferences, accurate results can be obtained with no tedious sample pretreatment. Examples of isotope selection for analysis of Cu and Zn are discussed. The feasibility of ICP-MS for direct, quantitative analysis of trace and ultratrace elements in biological fluids is demonstrated by analyzing two reference materials, Bio-Rad Lyphochek urine and Kaulson Contox sera.
| Materials and Methods |
|---|
|
|
|---|
sample preparation
Serum.
The preparation of serum specimens for ICP-MS
usually requires a simple dilution (20)(21)(22)(23)(24). However, a
precipitation method with nitric acid at room temperature to
precipitate protein is preferred in our laboratory; this is important
to minimize the occurrence of permanent blockage of the nebulizer, the
torch, and the sampling orifice resulting from high concentrations of
proteins, as well as to reduce polyatomic ion interferences such as
13C14N on 27Al. The specimen is
treated with nitric acid to precipitate proteins and quantitatively
release the trace elements, and then diluted with deionized water. The
precipitate is removed by centrifugation and the supernatant is used
for analysis. Deproteinization should be processed with care to prevent
the coprecipitation of trace metals of interest with residues. For some
elements, like Fe, precipitation of proteins in plasma or serum with
nitric acid is not appropriate because of the low recovery.
Freeze-dried lyophilized serum metals reference materials obtained from Kaulson Lab. were reconstituted with 5 mL of deionized water and 1-mL aliquots placed into 10-mL PP vials with metal-free PP pipette tips. All aliquots were then frozen until used. For analysis, serum was defrosted and a 1-mL stock solution of internal calibrators in 5% nitric acid was added. Protein precipitation occurred upon the addition of internal calibrators, owing to the high nitric acid concentration. The solution was then adjusted to 5 mL with deionized water and centrifuged. The clear supernatant was transferred to another PP vial through a filter for analysis.
Urine.
The specimen preparation required for urine is a
simple dilution with nitric acid; then the sample is ready for
analysis. The urine is the product of a set of complex processes in the
kidneys that removes low-molecular-mass molecules without loss of
proteins. Thus, no deproteinization is required unless protein
concentration is high; in this case, the precipitate should be removed
by centrifugation.
Freeze-dried lyophilized urine metals reference materials obtained from Bio-Rad were reconstituted with 25 mL of deionized water and 1-mL aliquots placed into 10-mL PP vials with metal-free PP pipette tips. All aliquots were then frozen until used. For analysis, urine was defrosted at room temperature. After the addition of 1 mL of stock solution of internal calibrators in 5% nitric acid, the solution was adjusted to 5 mL with deionized water to a final nitric acid concentration of 1%.
instrumentation
The instrument used is a Perkin-Elmer Sciex Elan 5000a ICP-MS
equipped with an AS 90 autosampler. A cross-flow nebulizer, a standard
Elan torch, and a Scott-type spray chamber were used. Platinum sampling
and skimmer cones were used for all studies. Solution uptake was
controlled by a peristaltic pump.
To obtain optimum signal intensity for multielement analysis at the
trace and ultratrace concentrations, it is necessary to consider the
optimization for the entire mass range because optimum ICP-MS
instrumental parameters vary from element to element (25).
Therefore, three elements, Mg at low mass, Rh at medium mass, and Pb at
high mass, at 10 µg/L in a calibrator from Perkin-Elmer, were used to
optimize ICP-MS instrumental parameters, including the nebulizer gas
flow rate, ion optics voltage, and aerosol injector orifice position
relative to the sampling cone, at 1.0 kW power, on a daily basis. By
using this three-element optimization, general signal optimization for
multielement analysis can be achieved. The instrument operating
parameters and the data acquisition parameters are listed in Table 1
.
|
| Results and Discussion |
|---|
|
|
|---|
nonspectral interferences
Nonspectral interferences in ICP-MS refer to analyte signal
intensity changes where the change cannot be accounted for by a
recognizable spectral overlap. The analyte signal intensity change can
be either a suppression (26)(27) or an
enhancement (27)(28)(29)(30)(31)(32). The causes of nonspectral
interferences or so-called matrix effects are due to the presence of
matrix components. The magnitude of the matrix effects depends on the
mass and ionization energy of the matrix elements. It has been reported
that the higher the mass of the matrix element, the larger the matrix
effects (29)(30)(31)(32). However, high mass matrix elements are
rare in biological fluids. For ionization dependence, Olivares and Houk
(28) found that the trend of matrix effects was in the
order of most easily ionized matrix element, i.e., Na (5.319)>Mg
(7.646)>I (10.451)>Br (11.814)>Cl (13.618), where the first
ionization energy in eV is noted in parentheses. In contrast,
biological specimens contain relatively large concentrations of easily
ionized matrix elements, such as Na, K, Ca, and Mg, which makes
analysis of these samples difficult. Such matrix effects need to be
eliminated to obtain accurate results. Errors associated with
matrix-induced signal variation can be corrected by means of an
appropriate calibration method. The merits of external calibration with
internal calibration and sample dilution, and the alternative strategy
of calibrator addition in correction for matrix effects were compared.
external calibration
There are several approaches to overcome matrix effects
(33)(34). The most widely used calibration
method is by a set of external calibrators containing elements of
interest and internal calibrators in a simple acid. Several calibrator
solutions are needed to cover the range of expected analyte
concentrations. The success of external calibration to correct for
matrix effects depends on the effectiveness of internal calibration and
dilution of samples.
internal calibration
The calibration or correction of one element by using a second as
a reference point has been used in a variety of analytical atomic
spectrometry and is termed "internal calibration." An element with
a known concentration is added to all solutions, including the blank,
calibrators, and unknowns. The analyte signal is then normalized to the
signal of the internal calibrator. The effectiveness of an internal
calibrator requires that its behavior accurately reflects that of the
elements of interest to be measured. Selection of an internal
calibrator is of great importance. Ideally, an internal calibrator
should undergo the identical matrix suppression or enhancement as that
of the analyte element. However, matrix-induced analyte signal changes
are not uniform for all elements, but depend on the mass of the element
(21)(35)(36). As a consequence, a
close match of the mass number between the analyte and internal
calibrator is of prime importance to effectively correct for matrix
effects. Fig. 1
shows the effectiveness of the internal calibration in NaCl
solutions as a function of NaCl concentration. The Ge signal (line B)
is suppressed by NaCl. After normalization to the close-mass internal
calibrator 71Ga signal, the 72Ge signal (line
A) is then independent of NaCl concentration and comparable with the
signal in a solution without NaCl. The matrix effect can thus be
effectively corrected by using a suitable internal calibrator with a
mass close to that of the analyte element. However, it is obvious that
one internal calibrator cannot be used for all elements, since the
matrix-induced signal variation is mass dependent. Several internal
calibrators should be used over the entire mass range. Yet, using a
large number of internal calibrators does not necessarily guarantee
more accurate analytical performance. Too many internal calibrators
(e.g., 10) may result in a practical problem of selecting internal
calibrators and higher errors (37). In general, three or
four elements as internal calibrators are considered to be adequate for
the multielement analysis
(21)(34)(36)(38)(39).
However, one internal calibrator still covers a group of analyte
elements, even though four internal calibrators are used for the
multielement analysis. It is impossible to accurately correct for all
elements of interest in that group by one internal calibrator. In
addition, even an element with similar chemical and physical properties
and similar atomic mass may behave in a different way in certain
matrices (34)(40). Therefore, some procedure
must be used to improve the deficiency of internal calibration.
|
dilution of sample
Specimen dilution is necessary for ICP-MS analysis of biological
samples because large amounts of proteins and salts can cause an
irreversible reduction of the analyte signal intensity due to clogging
of the nebulizer, torch, sampling, and skimmer orifices. One solution
is to dilute biological samples with a solvent before injection. In
addition, dilution can also reduce matrix effects while the sample is
being measured, since matrix effects are dependent on the absolute
amount of matrix element rather than the relative concentration of
matrix to analyte (32)(33).
Besides preventing clogging and reducing the matrix effect, dilution is
also necessary to improve the accuracy of internal calibration. Fig. 2
illustrates the effectiveness of internal calibration as a
function of NaCl concentrations. The Se signal (line B) was largely
suppressed by NaCl. Normalizing to the 71Ga internal
calibrator, the 82Se signal (line A) is still influenced by
NaCl concentrations but the effectiveness of the internal calibration
is enhanced when NaCl concentrations are decreased. The effectiveness
of internal calibration can be improved by dilution.
|
calibrator addition
Calibrator addition is performed by adding increasing quantities
of the elements of interest to multiple aliquots of the sample to be
analyzed. The calibration set therefore consists of several
supplemented samples plus an unsupplemented original sample, all of
which have an identical matrix. Matrix effect is corrected for and
highly accurate and precise data can be produced. Internal calibration
is also included to correct for instrumental drift.
From a practical viewpoint, external calibration with internal
calibration and dilution is most attractive for routine application
since it is less time consuming and results in less introduction of
matrix materials into the instrument than the other calibration
methods. However, the use of calibrator addition provides a way to
evaluate the accuracy of external calibration and is sometimes
necessary to compensate for the insufficiency of external calibration.
External calibration is shown to be adequate to correct for encountered
biological matrix effects for most elements. (Data are
discussed later.) Yet, external calibration was inadequate for some
elements, such as Cs and Zn. For Cs the results obtained for reference
urine levels I and II from external calibration, with In as the
internal calibrator, are only 67.32% and 68.65% of those from
calibrator additions, probably due to the large mass difference between
133Cs and 115In. However, insufficient
correction for 68Zn via external calibration was also
observed, even with the close-mass internal calibrator
71Ga. The results for both urine and serum reference
materials (data shown in Tables 4
and 5
) from external calibration were
located in the very low end of acceptable ranges and the recoveries
were poorer than those from calibrator addition. On the other hand, a
better agreement between calibrator additions and target values was
found. The insufficient corrections for Zn are probably caused by the
discrepancy of the first ionization potential between the analyte and
the internal calibrator: Zn (9.394 eV) and Ga (5.99 eV). Even after
fivefold dilution, internal calibration was still ineffective for the
correction of the remaining signal suppression. Calibrator addition is
thus an improvement over external calibration for those elements.
|
|
spectral interferences
Spectral interferences can be a major limitation since most ICP-MS
units are equipped with a quadrupole mass analyzer that limits the
resolution to approximately unit mass. Therefore, ions with the same
nominal mass as the analyte, resulting from singly charged ions, doubly
charged ions, and polyatomic ions, cannot be resolved. Polyatomic
interferences, resulting from combination of precursors in the Ar
plasma, entrained atmospheric gases, reagents, or biological matrices,
are more problematic. Elements >82 amu are essentially free of
polyatomic interferences and suitable for ICP-MS analysis, but elements
between 40 and 82 amu can be compromised by these interferences.
Accurate analytical results require that spectral interferences be
identified and then avoided or corrected.
choice of the isotope
The polyatomic interferences can be avoided in different ways
(24)(33)(34). A simple way is to
carefully select an isotope for analysis free from significant
interferences. An accurate and precise result can thus be obtained. To
eliminate the interferences effectively, the encountered polyatomic
ions must be identified and the extent of these interferences must be
assessed. Since it is impossible to address all polyatomic
interferences for every element of interest, only Cu and Zn are
discussed to illustrate that the correct choice of the available
analyte isotope is of prime importance to obtain an accurate result and
minimize the interference.
determination of cu
Cu has two stable isotopes: 63Cu (69.09%) and
65Cu (30.91%). The determination of Cu is complicated
because of polyatomic ion interferences from Na at mass 63
(40Ar23Na) as well as from Ca and S at mass 65
(48Ca16OH, 48Ca17O,
33S16O16O,
32S16O17O,
32S33S), where Na, Ca, and S are abundant in
biological fluids. Table 2
shows the results of Cu analysis on urine and serum reference
materials with masses 63 and 65 under external calibration and
calibrator addition. For Bio-Rad Lyphochek urine levels I and II, the
results under both calibrations and historically established values by
the former method of ARUP with 63Cu were higher than the
target values. 63Cu signals were largely influenced by the
spectral overlap of 40Ar23Na since the Cu
concentration in this sample is low. On the other hand, the
determinations made at mass 65 showed excellent agreement. The
polyatomic ions at mass 65, from S or Ca, do not result in significant
interferences. For Kaulson Contox serum levels I and II, because the
serum Cu concentration range is high the signal at mass 63 is not as
susceptible to 40Ar23Na interference. However,
the results by 63Cu under calibrator additions are still
too high. Again the determination made at mass 65 agreed well with the
target values. Therefore, we can conclude that the ICP-MS Cu method
based on the isotope 65 is suitable for analysis of urine and serum and
avoids the major interference.
|
determination of zn
Zn has five stable isotopes: 64Zn (48.89%),
66Zn (27.81%), 67Zn (4.11%), 68Zn
(18.56%), and 70Zn (0.62%). Different Zn isotopes,
including 64Zn (36), 66Zn
(20)(22)(23)(24)(36)(41),
and 68Zn (21), have been attempted for Zn
determination in the biological fluids.
The determination of Zn in body fluids can be complicated by the
different degrees of S-containing interferences on the Zn isotopes
(Table 3
) (21). Fig. 3
gives the apparent concentration of Zn at masses 64, 66, 67,
and 68 as a function of sulfuric acid concentrations. The apparent Zn
concentrations at masses 64, 66, and 67 increase linearly with the
sulfuric acid concentration. The S-containing polyatomic ions interfere
at mass 64 and to a lesser extent at masses 66 and 67. Only mass 68
shows no interference due to S. S-containing polyatomic ions at mass 68
were not detected. This means determination of Zn with masses 64, 66,
and 67 may be compromised by S interferences, and mass 68 is more
suitable for analysis. The agreement between the target values and the
results for urine and serum reference materials with 68Zn
is illustrated in Table 4
and Table 5
. The results obtained by external calibration are in poorer
agreement with the target values than calibrator addition because of
the occurrences of matrix suppression. However, the results indicate
that 68Zn is generally suitable for body fluids. The use of
68Zn for routine clinical analysis is recommended to
completely eliminate the interferences from S, leading to falsely
higher results with 64Zn and 66Zn.
|
|
analysis of urine and serum reference materials
Table 4
and Table 5
summarize the results of multielement analysis
obtained on a reconstituted solution of the Bio-Rad Lyphochek urine and
Kaulson Contox serum reference materials, respectively. For all
reference materials, both external calibration and calibrator addition
were applied for analysis. Four internal calibrators were used for
multielement coverage: 9Be for low masses;
71Ga for first transition metals as well as for As, Se, Rb,
and Sr; 115In for medium masses; and 193Ir for
high masses. In consideration of balancing the requirement of high
sensitivity for ultratrace concentrations and reduction of matrix
effects, a fivefold dilution was applied. The isotopes measured and
internal calibrators used are given together with the analytical
results within 95% confidence limits. For urine, each result is based
on three sample aliquots, analyzed with three repeats, for a total of
nine measurements. The results are compared with the target values
verified by atomic absorption spectrometry (AAS). For serum, each
result is based on two sample aliquots, analyzed with three repeats,
for a total of six measurements. The results are compared with the
target values referenced by AAS.
In urine, for eight trace elements (As, Cd, Co, Cu, Mn, Ni, Pb, Sb), the results obtained from both calibration methods are in good agreement with target values. For Zn the results obtained from calibrator addition are in better agreement with the target values than external calibration. For Al the results are at the lower end of certified acceptable ranges. For Se the results are higher than the acceptable target ranges, except level II with calibrator addition. However, in comparison with the values, level I: 76.50 ± 5.51 µg/L and level II: 229.25 ± 16.56 µg/L obtained by ARUP, in which the methodology was verified by interlaboratory correlation with certified AAS methods (42), the agreement is excellent. For Tl the results are lower than the acceptable target ranges but in good agreement with ARUP's values (level I: 8.38 ± 0.52 µg/L; level II: 167.43 ± 6.95 µg/L). For seven other elements (Bi, Cs, Mo, Rb, Sn, Sr, W) no data are available for the comparison.
In serum, the results obtained for level II agreed with the target values, except for Mn under external calibration. The results obtained for level I agreed with the target values, except for Ni and Bi. For Ni the results were higher than the target value, which is probably because the signal at mass 60 is susceptible to the 44Ca16O interference at low concentration. For Bi the concentration was lower than the detection limit.
precision
Precision is evaluated by the CV of the repetitive analysis of
reference urine and serum under external calibration and calibrator
additions. Each result is the average of levels I and II and
corresponds to 95% confidence limits. For urine most of the elements
in calibrator addition have a comparable and higher precision than
external calibration. For either calibration method, elements with
ultratrace concentrations (<1 µg/L) have poorer precision. Elements
with concentrations >1 µg/L have better precision. The imprecision
in the collected data is significantly smaller than the uncertainties
of the target values. For serum, the precision for both calibration
methods is comparable. The use of calibrator addition does not
particularly enhance precision over external calibration. For elements
Sn and Pb, precision is poorer, especially in level I, owing to the
very low concentrations.
In conclusion, ICP-MS is a practical, versatile method for the determination of many trace and ultratrace elements in the clinical laboratory. However, nonspectral and spectral interferences need to be taken into consideration for the analysis of elements in biological matrices. With internal calibrators of close mass, internal calibration is able to effectively correct for encountered nonspectral interferences; however, specimen dilution is still necessary to improve the effectiveness of internal calibration. With four internal calibrators (Be, Ga, In, Ir) across the entire mass range and fivefold specimen dilution, external calibration is able to accurately measure a variety of trace and ultratrace elements in urine and serum, with the exception of some elements (e.g., Cs, Zn) that have a large discrepancy in atomic mass and ionization energy with that of the internal calibrator. Calibrator addition provides an alternative method to compensate for the insufficiency of external calibration and yields more accurate and precise results in certain cases. Spectral interferences can be eliminated by a simple approach of selecting the isotope with minimal interferences. With proper analytical processes, ICP-MS is generally suitable for the routine multielement analysis of body fluids.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
S. M. Smith, S. A. Mathews Oliver, S. R. Zwart, G. Kala, P. A. Kelly, J. S. Goodwin, and C. B. Dyer Nutritional Status Is Altered in the Self-Neglecting Elderly J. Nutr., October 1, 2006; 136(10): 2534 - 2541. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. Smith, J. E. Davis-Street, J. V. Fesperman, M. D. Smith, B. L. Rice, and S. R. Zwart Nutritional Status Changes in Humans during a 14-Day Saturation Dive: The NASA Extreme Environment Mission Operations V Project J. Nutr., July 1, 2004; 134(7): 1765 - 1771. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |