Clinical Chemistry 44: 1443-1451, 1998;
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (17)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nauck, M.
Right arrow Articles by Wieland, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nauck, M.
Right arrow Articles by Wieland, H.
Related Collections
Right arrow Lipids, Lipoproteins, and Cardiovascular Risk Factors
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 1998;44:1443-1451.)
© 1998 American Association for Clinical Chemistry, Inc.


Lipids and Lipoproteins

New immunoseparation-based homogeneous assay for HDL-cholesterol compared with three homogeneous and two heterogeneous methods for HDL-cholesterol

Matthias Naucka, Winfried März, and Heinrich Wieland

a Address correspondence to this author at: Department of Medicine, Division of Clinical Chemistry, Hugstetter Strasse 55, 79106 Freiburg i. Br., Germany. Fax 49-761-270 3444; e-mail manauck{at}mzl200.ukl.uni-freiburg.de.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
We evaluated four new commercial methods for HDL-cholesterol determination. The three completely homogeneous assays were an immunoseparation-based (IS) method from Wako, a polyethylene glycol-modified enzyme (PEG) method from Boehringer Mannheim, and a synthetic polymer-based (SP) method from Genzyme. The fourth method was a new heterogeneous method in which lipoproteins are removed using dextran sulfate-coated magnetic beads and Mg2+ (MB, Reference Diagnostics). We compared these methods with the conventional phosphotungstic acid/MgCl2 precipitation (PTA) procedure. The homogeneous assays had good intraassay imprecision with total CVs <2.3%, whereas the CVs of the MB assay were <5.9%. Adding HDL to serum to achieve HDL-cholesterol (HDL-C) concentrations up to 1000 mg/L revealed nearly complete recoveries in the IS, PEG, and MB assays, whereas the SP assay showed a lower recovery (~70%). The SP HDL-C apparently increased at increasing LDL-cholesterol and VLDL-triglyceride concentrations, whereas the IS, PEG, and MB methods were not influenced by LDL-cholesterol up to 6 000 mg/L (MB, 5 000 mg/L) and VLDL-triglycerides up to 9 000 mg/L. Free fatty acids above ~2 mmol/L produced falsely high HDL-C in the IS and SP assays, the error amounting to as much as 50% in some samples. An intermethod comparison in 291 fresh serum samples yielded correlation coefficients of at least r = 0.95 for all assays, when compared with the PTA procedure. The slopes and intercepts of the regression lines were 1.05 and 57 (IS), 1.12 and 9.9 (PEG), 1.00 and 39 (SP), and 1.0 and 38 mg/L (MB), respectively. The new assays are precise and simplify the determination of HDL-C, but in part they lack specificity or are susceptible to interferences, resulting in discrepancies when compared with the established PTA procedure.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
The positive association of coronary heart disease risk with total cholesterol and LDL-cholesterol (LDL-C)1 concentrations and its negative association with HDL-cholesterol (HDL-C) concentrations are well established (1)(2)(3)(4). HDL-C concentrations <350 mg/L are considered as a cardiovascular risk factor; HDL-C concentrations exceeding 600 mg/L may be protective (3). According to Friedewald et al. (5), together with total cholesterol and triglycerides, the determination of HDL-C allows the calculation of LDL-C. For these reasons, reliable and easy-to-perform methods are needed to quantify HDL-C.

Most frequently, HDL-C is measured in the supernatant after precipitation of apolipoprotein (apo) B-containing lipoproteins by phosphotungstic acid/MgCl2 (PTA) or dextran sulfate (6)(7). All of these methods involve both a precipitation and a centrifugation step, which prevent full automation. Previously, new procedures for the determination of HDL-C were published (8)(9)(10)(11)(12). They have in common that they avoid at least the tedious centrifugation step.

In 1996, we published the results of an evaluation of a homogeneous HDL-C assay that included polyethylene glycol (PEG) to complex the non-HDL lipoproteins (reagent 1); antibodies against apo B and apo C-III, which aggregate the non-HDL lipoproteins (reagent 2); enzymes for cholesterol determination (reagent 3); and guanidine hydrochloride to stop all reactions (reagent 4) (9). The performance characteristics of this method were good, but the disadvantage of the procedure was that it relied on four different reagents, which limited its application to a small number of automatic analyzers.

The aim of this study was to evaluate three homogeneous assays [an immunoseparation-based (IS) assay from Wako, a PEG-based assay from Boehringer Mannheim, and a synthetic polymer-based (SP) assay from Genzyme] and a dextran sulfate-coated magnetic bead-based (MB) procedure from Reference Diagnostics for the determination of HDL-C in parallel with the conventional PTA assay with respect to precision, linearity, interferences, and ease of handling.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
samples
Blood was drawn from 291 in- and outpatients of the University Hospital of Freiburg, including nonfasting samples. The blood was allowed to clot at room temperature, and serum was obtained by centrifugation at 1500g for 15 min. All analyses with the homogeneous and the precipitation assays were performed within 1 day of blood collection.

lipid measurements
Cholesterol and triglycerides were determined enzymatically with the CHOD-PAP and GPO-PAP methods, respectively. These reagents were purchased from Boehringer Mannheim. All measurements, including the HDL-C assays, were performed on a Boehringer Mannheim/Hitachi automatic analyzer type 911. Control sera Precinorm® L and Precipath® L (both from Boehringer Mannheim) were included in each analytical run. In addition, the nonesterified fatty acids were determined with the NEFA C assay, purchased from Wako, on a Wako 30R automatic analyzer in a subset of 130 samples. All tests, including the new homogeneous assays, were performed according to the manufacturers' recommendations and calibrated using the standards included in the test kits. Reagents of the homogeneous assays were lyophilized and solubilized before use.

is-based homogeneous hdl-c assay from wako
The reagents for the IS assay were obtained from Wako. In this assay, anti-human apo B antibodies react with non-HDL particles, i.e., chylomicrons, VLDL, and LDL (reagent 1). These complexes do not react with the enzymes involved in the subsequent enzymatic cholesterol determination, whereas the cholesterol content of the HDL particles is measured (reagent 2).

homogeneous hdl-c assay from boehringer mannheim based on peg-modified enzymes
The reagents for the PEG assay were obtained from Boehringer Mannheim. At a neutral pH value (pH 7.0) and in the presence of MgCl2, sulfated {alpha}-cyclodextrin and dextran sulfate form water-soluble complexes with LDL, VLDL, and chylomicrons (reagent 1), which are not accessible to PEG-modified enzymes (reagent 2). Because HDL is not complexed, the cholesterol moiety is readily available for enzymatic quantification using cholesterol esterase and cholesterol oxidase coupled to PEG (8).

sp-based homogeneous hdl-c assay from genzyme
The reagents for the SP assay were obtained from Greiner Biochemica, Flacht, Germany. Synthetic polymers and polyanions adsorb to the surface of LDL, VLDL, and chylomicrons and convert these lipoproteins into denaturation-resistant complexes (reagent 1). In a second step, a detergent destroys the HDL structure, and the cholesterol content of the HDL particles is determined enzymatically (reagent 2) (11).

mb-based hdl-c assay
All reagents, cups, and tubes for this assay were obtained from Reference Diagnostics. In a flat bottom cup, 500 µL of serum was incubated with 100 µL of reagent containing 250 mmol/L Mg2 and MBs. After the cups were mixed for a short period on a vortex-type mixer and incubated for 10 min, they were transferred to a magnet-containing tube, which was then directly placed in the Hitachi analyzer. The dextran sulfate binds all apo B-containing particles, which are pulled down in the magnetic field within 1 min. In severe hypertriglyceridemic samples, this separation can take up to 30 min. The cholesterol content of the HDL particles that remain in the supernate can easily be measured by the CHOD-PAP method. We applied a modified version of this assay, in which the accuracy at low cholesterol concentrations has been claimed to be improved by using a higher sample volume. The dilution of the sample was accounted for by multiplying the results by 1.2 (10).

hdl-c determination by a pta-based method
We used the PTA procedure as the comparison method. The reagents were purchased from Boehringer Mannheim. Two hundred microliters of serum were mixed with 500 µL of the PTA reagent and incubated for at least 5 min. The apo B-containing lipoproteins were sedimented by centrifugation (5 min), and the cholesterol content was measured in the supernate with the CHOD-PAP method.

hdl-c determination by a combined ultracentrifugation and precipitation method
In 130 samples, a combined ultracentrifugation-precipitation assay was used as the comparison method. Essentially, the protocol of the Lipid Research Clinics Program was followed, with modifications previously described (13)(14). In brief, VLDL were floated by ultracentrifugation (density = 1.006 kg/L), and the LDL was separated from the HDL in the infranatant by PTA.

isolation of vldl, ldl, and hdl, and preparation of lipoprotein-deficient serum
VLDL, LDL, and HDL were isolated by sequential ultracentrifugation using density <1.006 kg/L, 1.006 < density < 1.063 kg/L, and 1.063 < density < 1.21 kg/L as density limits (15). Lipoprotein-deficient serum was obtained by 2 x repeated ultracentrifugation at 150 000g for 48 h after adjusting the serum to density = 1.25 kg/L by adding solid KBr.

apo a-i
Apo A-I was measured turbidimetrically on the Wako-30R analyzer with reagents purchased from Greiner Biochemica. The assay was calibrated against the IFCC apo A-I standard.

linearity
Testing of linearity was performed by adding HDL isolated by ultracentrifugation to serum. To a constant volume of serum, increasing amounts of HDL were added. To modify the matrix as minimally as possible, the samples were brought to equal volumes with lipoprotein-deficient serum.

interferences
The influences of LDL and VLDL on HDL-C measurements were investigated in experiments in which they were added to samples. In addition, free fatty acids (FFAs) and bilirubin were measured in the samples of the intermethod comparisons. Interference from hemoglobin was analyzed according to Glick et al. (16).

statistical methods
Precision data were calculated according to recommendations of the NCCLS/EP5-T protocol (17). Regression analyses were performed using the method of Passing and Bablok (18). Total error was calculated as the sum of the systematic error plus random error (19)(20). Systematic error is calculated from the linear regression equation yc = bxc a, where b is the slope of the regression line and a is the y-axis intercept. At an HDL-C concentration of xc, systematic error is the absolute value of yc - xc. Random error is 1.96x SD from the run-to-run precision study.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
precision
Two commercial control sera with low and medium HDL-C concentrations and one human serum pool were used to assess the precision of the HDL-C assays (Table 1 ). The within-day imprecision values were determined as the average of the imprecision obtained each day. The between-day imprecision values were calculated from the imprecision of the means obtained each day, which were then adjusted for the within-day imprecision component. The total CVs ranged from 1.1% to 2.3% for the three homogeneous assays, from 3.9% to 5.9% for the MB assay, and from 3.1% to 3.7% for the PTA assay, respectively.


View this table:
[in this window]
[in a new window]
 
Table 1. Summary of the precision studies in control materials and a human serum pool.

linearity
The new HDL-C assays were linear up to at least 1000 mg/L, as revealed by increasing amounts of HDL prepared by sequential ultracentrifugation to sera (Fig. 1 ). For the SP assay, a slope of ~0.7 was obtained when the measured HDL-C was plotted against HDL-C added, suggesting that the recovery of this assay was <100%.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 1. Determination of HDL-C: linearity and analytical recovery.

Increasing amounts of HDL isolated by ultracentrifugation were added to a serum pool. The measured HDL increased linearly with the amounts of HDL added up to at least 1000 mg/L. The equations of the regression lines are: IS ({blacksquare}), y = 0.98x + 21.3; PEG ({blacksquare}), y = 0.99x + 19.3; SP ({bullet}), y = 70x + 14.4; and MB ({blacktriangleup}), y = 0.90x + 23.1.

specificity
HDL-C concentrations are plotted against added LDL-C and VLDL-triglycerides, respectively, in Fig. 2 .



View larger version (14K):
[in this window]
[in a new window]
 
Figure 2. Specificity of the HDL-C assays.

Cross-reactivity with LDL (A) and cross reactivity with VLDL (B) in the IS ({blacksquare}), PEG ({blacksquare}), SP ({bullet}), and MB ({blacktriangleup}) assays.

Cross-reactivity with LDL varied between assays. The IS assay was not influenced by LDL-C up to 6000 mg/L. The PEG assay showed a minimal increase over the whole concentration range, but deviations >5% occurred only in samples with LDL-C above 6000 mg/L. LDL-C increased the HDL-C values of the SP assay at any concentration. The MB assay was markedly influenced at LDL-C concentrations above 5000 mg/L.

Cross-reactivity with VLDL also varied between assays. Triglycerides did not substantially interfere with the IS and PEG assays up to 9000 mg/L VLDL-triglycerides. The HDL-C values of the SP assay increased over the whole concentration range, whereas VLDL-triglycerides interfered with the MB assay at concentrations of ~5000 mg/L. Both the IS and the MB assays showed an unexpected negative bias with increasing VLDL-triglycerides.

intermethod comparisons
As the comparison method, we used the PTA procedure either in whole serum (n = 161), or after removal of VLDL by ultracentrifugation (n = 130).

A total of 291 fresh sera were analyzed with each of the new methods and the comparison method. Mean total cholesterol and total triglycerides were 1690 and 1680 mg/L, respectively. Maximum cholesterol and triglyceride concentrations were 3670 and 6830 mg/L, respectively. HDL-C concentrations ranged between 35 and 906 mg/L, determined with the PTA method.

The correlation coefficients were between 0.949 and 0.987 (Table 2 ). When we estimated the parameters of the regression lines according to Passing and Bablok, slopes relating the homogeneous HDL-C IS and PEG assays to the PTA assay were significantly higher than 1.00 (P<0.05), whereas the SP and MB methods produced slopes close to 1.00 (18). All intercepts were significantly (P<0.05) different from 0.0 mg/L, which produced significantly (P<0.05) higher HDL-C values for all new methods, compared with the conventional precipitation procedure. The bias plots of the IS and PEG assays revealed small but systematic deviations, which increased slightly as HDL-C concentrations increased (Fig. 3 ). At HDL-C concentrations <350 mg/L, some of the IS results showed a deviation of approximately -200 mg/L, whereas the PEG assay produced only one outlier. The bias plot of the SP assay showed substantial scattering over the entire range. The results of the MB assay were negatively biased throughout the entire concentration range, with only a small scattering.


View this table:
[in this window]
[in a new window]
 
Table 2. Summary of intermethod comparisons.1



View larger version (23K):
[in this window]
[in a new window]
 
Figure 3. Bias plots of HDL-C (new assay minus PTA assay) in the IS (A), PEG (B), SP (C), and MB (D) assays.

The correlation coefficients between the apo A-I measurements and the HDL-C concentrations of the IS, PEG, SP, and MB assays were 0.92, 0.93, 0.89, and 0.93, respectively.

total error
The total errors were calculated at the decision points of 350 mg/L and 600 mg/L, respectively. As shown in Table 3 , the systematic deviations contributed much more to the total errors than did the random errors. The highest total errors were found for the IS and PEG assays.


View this table:
[in this window]
[in a new window]
 
Table 3. Total error for the new HDL-C assays.

free fatty acids
Conditions with increased FFAs occur mainly in inpatients receiving intravenous heparin therapy. At concentrations >2 mmol/L, there is substantial binding of FFAs to lipoproteins (21). This increases the electrophoretic mobilities of the lipoproteins and may change their interaction with precipitating reagents (22)(23)(24). The results obtained in those serum samples of the intermethod comparisons that had FFA concentrations >2.0 mmol/L (n = 22) indicated that the IS and SP assay were susceptible to interference from FFAs, whereas the assays from PEG and MB were not (Fig. 4 ). Experiments with samples enriched in FFA, according to the method of Spector and Hoak (25), completely confirmed these findings (data not shown).



View larger version (15K):
[in this window]
[in a new window]
 
Figure 4. Interferences of FFAs with the determination of HDL-C (ratio, new assay/PTA assay) in the IS (A), PEG (B), SP (C), and MB (D) assays.

interferences
Hemoglobin at concentrations up to 2 g/L did not interfere substantially with the new homogeneous HDL-C assays using IS and PEG (16). The SP assays yielded lower values, whereas in the MB and the PTA method HDL-C increased markedly when hemoglobin was added (Fig. 5 ).



View larger version (16K):
[in this window]
[in a new window]
 
Figure 5. Interferences of hemoglobin with the IS ({blacksquare}), PEG ({blacksquare}), SP ({bullet}), MB ({blacktriangleup}), and PTA ({circ}) HDL-C assays, tested according to Glick et al. (16).


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
As shown in many epidemiological and clinical studies, a low concentration of HDL-C is an important risk factor for coronary artery disease (1). Thus far, HDL-C has widely been determined by methods in which apo B-containing lipoproteins are precipitated with polyanions and bivalent cations. These methods are precise and agree well with ultracentrifugation methods but are tedious in respect to the centrifugation step involved (26)(27). Recently, several new HDL-C assays have been devised which have–with exceptions–satisfactory performance characteristics (8)(9)(10)(11)(12).

Here we present the results of a comparison study that included three principally different homogeneous HDL-C assays: the first (IS) uses antibodies for the separation of non-HDL lipoproteins coupled to a conventional enzymatic cholesterol assay; the second (PEG) uses MgCl2, sulfated {alpha}-cyclodextrin, dextran sulfate, and PEG-coupled enzymes; the third (SP) uses synthetic polymers, polyanions, and detergents coupled with a conventional enzymatic cholesterol assay for the quantification of HDL-C. The fourth method (MB) is based on the absorption of apo B-containing lipoproteins to dextran sulfate-coated magnetic beads. All of these assays were compared with the conventional PTA procedure.

General advantages of the homogeneous methods are that they obviate pretreatment of the samples and that only small sample volumes are needed. These features allow the fully automated determination of HDL-C even in the emergency laboratory, where HDL-C may be useful in the diagnosis of serious infections, such as malaria, or to monitor patients after liver transplantation (28)(29).

The National Cholesterol Education Program (NCEP) performance goals for 1998 demand that the CV of HDL-C determinations be <4% at concentrations >420 mg/L (30), a criterion that was met by all investigated HDL-C assays. The NCEP precision goal for HDL-C concentrations <420 mg/L is a SD <17 mg/L (30). The homogeneous assays and the PTA procedure fulfilled these criteria. The MB assay produced a SD >17 mg/L in one of three pools examined. Thus, the NCEP precision goals will obviously be met by all homogeneous assays considered here at both high and low HDL-C concentrations, whereas the MB assay may need further improvement. With regard to the precision, the differences between the three homogeneous HDL-C assays are small and appear not clinically relevant. These data agree with previous publications (9)(10)(11)(12).

The slopes of the experiments with added HDL indicated nearly complete recoveries with the IS and PEG assay. Our data confirm that the specificity of the PEG method is good (10)(12)(30)(31). Furthermore, they indicate that the antibody concentrations of the IS assay are sufficient to complex all non-HDL particles. The specificity of the IS assay is thus at least as good as that of the homogeneous HDL-C from International Reagents Corporation, Kobe, which also contains antibodies to form stable aggregates of the non-HDL lipoproteins. The latter method, however, has the disadvantage of including four different reagents (8).

On the other hand, only ~70% of the added HDL was recovered with the SP method. The addition of LDL and VLDL produced falsely high HDL-C concentrations, indicating limited specificity of this method, as described previously (11)(12).

The MB assay was specific, but the interferences from LDL-C and VLDL-triglycerides started at lower concentrations than in the IS and PEG assays.

Recently, Demacker et al. (27) showed that the PTA procedure agrees well with the CDC reference method. Therefore, and because the PTA procedure is very common in laboratories, it is well suited as a comparison method.

We obtained excellent correlation coefficients when we compared the homogeneous HDL-C assays with the PTA procedure.

The slope of the IS assay was >1.00, which caused the positive slope of the bias plot (Fig. 3Up ). This was due to a falsely high specification of the calibrator provided by the manufacturer, so that the results do not exactly agree with the PTA procedure.

In a previous premarketing evaluation of the PEG assay, a positive bias of 3% was observed (31). The optimistic data of this previous multicenter study were not reproduced in the current study. The PEG assay showed a slope significantly (P<0.05) exceeding unity. This deviation was due to a falsely high declaration of the calibrator, which may have occurred on large-scale production. Finally, a very recently performed evaluation of the PEG method performed in our laboratory showed that the positive bias had disappeared, indicating that this assay is now working well (y = 0.98x 11.9 mg/L; r = 0.997; n = 100).

The unsatisfactory results of the SP assay have been reported by other investigators as well (11)(12). The low specificity of this assay is hampering the use of this method in the clinical chemical laboratory, in spite of the good imprecision and a slope of 1.0 in the intermethod comparison. The scattering of the bias plots, caused by the nonspecificity of this method, is not acceptable.

The MB assay showed the best agreement with the comparison method, including an ideal slope of 1.0. It is worth noting that all results of the MB assay were slightly higher than the PTA procedure, caused by a small positive intercept of the regression line.

The NCEP goals for 1998 demand a total error of <13% (30). In the current study, the IS and PEG homogeneous assays have not met these criteria because of high systematic error caused by an incorrect calibration (slopes >1 plus a positive y-intercept in the intermethod comparison). The SP assay had the lowest total error because of a very small systematic error. Obviously, the limited specificity of the SP assay for HDL is balanced by a suitable calibrator, which accounts for the fact that LDL-C and VLDL-C are in part determined as HDL-C. The MB assay meets the NCEP criteria, the total error amounting to 13.2% at 350 mg/L.

A common denominator of the present study and of many earlier reports is that homogeneous assays tend to overestimate HDL-C in the range of ~10%. This raises the suspicion that homogeneous assays may yield an HDL fraction that differs in some respect from HDL found in the supernate of precipitation reactions. Such a systematic difference between homogeneous and conventional methods would not be entirely surprising. HDL represents a polydisperse system of particles markedly differing by physicochemical characteristic, apolipoprotein composition, and pathobiochemical importance. On the basis of their apo E content, two populations of HDL particles can be distinguished, those without apo E and those with apo E. The latter accounts for ~10% of the total HDL-C. In the PTA assay, the apo E-containing HDL particles are precipitated and therefore not included in the HDL-C (32). In the homogeneous PEG assay, in contrast, these particles contribute to HDL-C (33), a finding that may at least in part account for the bias found between the homogeneous and the precipitation methods. It is currently not known how different HDL subfractions behave in the other homogeneous methods.

What are the consequences of these differences in practice? The large epidemiological studies, which established the role of low HDL-C concentrations as a risk factor of coronary artery disease, were performed with precipitation methods adjusted to the CDC reference method. These studies will probably not be repeated in the future because they are expensive and the results would be biased through the widespread use of lipid-lowering therapies, even in primary prevention. It will probably not be feasible to change the cutoff point for HDL-C from 350 (without apo E-containing HDL-C) to 380 mg/L (including apo E-containing HDL-C). More pragmatic, however, will be adjusting the new methods to the CDC reference method (31)(34), thereby deliberately disregarding the fact that the homogeneous assays yield an HDL fraction that may physicochemically be distinct from HDL produced by precipitation reactions.

The reagents for the homogeneous HDL-C assay are approximately threefold more expensive than conventional precipitation reagents. With the exception of the SP assay, the homogeneous assays evaluated here are convenient to use in routine laboratories. The MB assay, which also avoids centrifugation, is substantially cheaper than the homogeneous assays and therefore represents an alternative for laboratories with a workload of <50 samples per day. However, in most laboratories, the higher reagent costs for the new HDL-C assays will probably be offset by economizing the laboratory work.

In summary, the homogeneous IS and PEG assays produce precise and accurate determinations of HDL-C. Even hypertriglyceridemic samples up to at least 9000 mg/L revealed unbiased results. The MB assay showed the best agreement with the PTA procedure, except in of hypertriglyceridemic and hemolytic samples. This assay is cheaper than the homogeneous assays but is not fully automated. All new HDL-C assays, with the exception of the SP assay, represent a major improvement in our methodology to quantify HDL-C and may facilitate the identification of individuals at increased risk of atherosclerosis. Although it was the only homogeneous assay to meet the NCEP performance goals, the SP assay proved nonspecific and is therefore not recommended for use in the clinical laboratory.


   Acknowledgments
 
We are grateful for the excellent technical assistance of Gabi Herr, Brigitte Kreisel, and Sibylle Rall. We also thank Wako, Boehringer Mannheim, Greiner, and Reference Diagnostics for providing HDL-C assay kits free of charge.


   Footnotes
 
University Hospital of Freiburg, 79106 Freiburg, Germany.

1 Nonstandard abbreviations: LDL-C, LDL-cholesterol, HDL-C, HDL-cholesterol; apo, apolipoprotein; PTA, phosphotungstic acid/MgCl2 precipitation; PEG, polyethylene glycol; IS, immunoseparation; SP, synthetic polymer; MB, dextran sulfate-coated magnetic bead; FFA, free fatty acid; and NCEP, National Cholesterol Education Program.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Law MR, Wald NJ, Thompson SG. By how much and how quickly does reduction in serum cholesterol concentration lower risk of ischaemic heart disease?. Br Med J 1994;308:367-372. [Abstract/Free Full Text]
  2. Wilson PW, Abbott RD, Castelli WP. High density lipoprotein cholesterol and mortality. The Framingham Heart Study. Arteriosclerosis 1988;8:737-741. [Abstract/Free Full Text]
  3. . National Cholesterol Education Program. Second report of the expert panel on detection, evaluation, and treatment of high blood cholesterol in adults (adult treatment panel II). Circulation 1994;89:1330-1445.
  4. . International Task Force for Prevention of Coronary Heart Disease. Prevention of coronary heart disease: scientific background and new clinical guidelines. Recommendations of the European Atherosclerosis Society. Nutr Metab Cardiovasc Dis 1992;2:113-156.
  5. Friedewald WT, Leva RJ, Fredrickson DS. Estimation of the concentration of low density lipoprotein cholesterol in plasma, without use of preparative ultracentrifuge. Clin Chem 1972;18:499-509. [Abstract]
  6. Warnick GR, Cheung MC, Albers JJ. Comparison of current methods for high density lipoprotein cholesterol quantification. Clin Chem 1979;25:596-604. [Abstract/Free Full Text]
  7. Demacker PNM, Vos-Janssen HE, Hijmans AGM, van't Laar A, Janssen AP. Measurement of high-density lipoprotein cholesterol in serum: comparison of six isolation methods with enzymatic cholesterol analysis. Clin Chem 1980;26:1780-1786. [Free Full Text]
  8. Sugiuchi H, Yoshinori U, Okabe H, Irie T, Uekama K, Kayahara N, Miyauchi K. Direct measurement of high-density lipoprotein cholesterol in serum with polyethylene glycol-modified enzymes and {alpha}-cyclodextrin. Clin Chem 1995;41:717-723. [Abstract/Free Full Text]
  9. Nauck M, März W, Haas B, Wieland H. Homogeneous assay for direct determination of high-density lipoprotein cholesterol evaluated. Clin Chem 1996;42:424-429. [Abstract/Free Full Text]
  10. Harris N, Galpachian V, Rifai N. Three routine methods for measuring high-density lipoprotein cholesterol compared with the reference method. Clin Chem 1996;42:738-743. [Abstract/Free Full Text]
  11. Harris N, Galpachian V, Thomas J, Iannotti E, Law T, Rifai N. Three generations of high-density lipoprotein cholesterol assays compared with ultracentrifugation/dextran sulfate-Mg2+ method. Clin Chem 1997;43:816-823. [Abstract/Free Full Text]
  12. Huang YC, Kao JT, Tsai KS. Evaluation of two homogeneous methods for measuring high-density lipoprotein cholesterol. Clin Chem 1997;43:1048-1055. [Abstract/Free Full Text]
  13. Manual of laboratory operation. Lipid Research Clinics Program. DHEW No. (NIH) 75–628. Bethesda, MD: National Heart and Lung Institute, 1974:1–74..
  14. Wanner C, Hörl WH, Luley C, Wieland H. Effects of HMG-CoA reductase inhibitors in hypercholesterolemic patients on hemodialysis. Kidney Int 1991;39:754-759. [ISI][Medline] [Order article via Infotrieve]
  15. Havel RJ, Eder HA, Bragdon JH. The distribution and chemical composition of ultracentrifugally separated lipoproteins in human plasma. J Clin Invest 1955;34:1345-1353.
  16. Glick MR, Ryder KW, Jackson SA. Graphical comparisons of interferences in clinical chemistry instrumentation. Clin Chem 1986;32:470-475. [Abstract/Free Full Text]
  17. . NCCLS. User evaluation of precision performance of clinical chemistry devices. Tentative Guideline EP5-T 1984 National Committee for Clinical Laboratory Standards Villanova, PA. .
  18. Passing H, Bablok W. Comparison of several regression procedures for method comparison studies and determination of sample size. Part II. J Clin Chem Biochem 1984;22:431-445.
  19. Carey RN, Garber CC. Evaluation of methods. Kaplan LA Pesce AJ eds. Clinical chemistry: theory, analysis and correlation 3rd ed. 1996:402-423 CV Mosby St. Louis. .
  20. Westgard JO, Carey RN, Wold S. Criteria for judging precision and accuracy in method development and evaluation. Clin Chem 1974;20:825-833. [Abstract]
  21. Nauck M, März W, Wieland H. Adding albumin normalizes electrophoretic mobility of lipoproteins in sera with high concentrations of free fatty acids. Clin Chem 1996;42:1283-1285. [Free Full Text]
  22. Armstrong V, Seidel D. Evaluation of a commercial kit for the determination of LDL cholesterol in serum based on precipitation of LDL with dextran sulfate. Ärztl Lab 1985;31:325-333.
  23. Nauck M, Friedrich I, Wieland H. Measurement of LDL and VLDL cholesterol with precipitation techniques. A comparison with the ultracentrifugation method. Klin Lab 1994;40:167-176.
  24. Siekmeier R, März W, Gross W. Insufficient accuracy and specificity of polyanion precipitation methods for quantifying low-density lipoproteins. Clin Chem 1990;36:2109-2113. [Abstract/Free Full Text]
  25. Spector AA, Hoak JC. An improved method for the addition of long-chain free fatty acid to protein solutions. Anal Biochem 1969;32:297-302. [ISI][Medline] [Order article via Infotrieve]
  26. März W, Gross W. Analysis of plasma lipoproteins by ultracentrifugation in a new fixed angle rotor: evaluation of a phosphotungstic acid/MgCl2 precipitation and a quantitative lipoprotein electrophoresis assay. Clin Chim Acta 1986;160:1-18. [ISI][Medline] [Order article via Infotrieve]
  27. Demacker PNM, Hessels M, Toenhake-Dijkstra H, Baadenhuijsen H. Precipitation methods for high-density lipoprotein cholesterol measurement compared, and final evaluation under routine operating conditions of a method with a low-to-reagent ratio. Clin Chem 1997;43:663-668. [Abstract/Free Full Text]
  28. Nilsson-Ehle I, Nilsson-Ehle P. Changes in plasma lipoproteins in acute malaria. J Intern Med 1990;227:151-155. [ISI][Medline] [Order article via Infotrieve]
  29. Nikkila K. Liver transplantation restores low serum levels of very low density and high density lipoproteins in end-stage primary cirrhosis. Ann Med 1992;24:129-136. [ISI][Medline] [Order article via Infotrieve]
  30. . for the National Cholesterol Education Program Working Group on Lipoprotein MeasurementWarnick GR, Wood PD. National Cholesterol Education Program recommendations for measurement of high-density lipoprotein cholesterol: executive summary. Clin Chem 1995;41:1427-1433. [Free Full Text]
  31. Nauck M, März W, Jarausch J, Cobbaert C, Sägers A, Bernard D, et al. Multicenter evaluation of a homogeneous assay for HDL cholesterol without sample pre-treatment. Clin Chem 1997;43:1622-1629. [Abstract/Free Full Text]
  32. Okamoto Y, Tsujii H, Haga Y, Tanaka S, Nakano H. Determination of apolipoprotein E in high-density lipoprotein fraction by immuno-fixation method and turbidimetric immunoassay after precipitation. J Atheroscler Thromb 1994;1:23-29. [Medline] [Order article via Infotrieve]
  33. Okamoto Y, Tanaka S, Nakano H. Direct measurement of HDL cholesterol preferable to precipitation method. Clin Chem 1995;41:1784.[ISI][Medline] [Order article via Infotrieve]
  34. Warnick GR, Wood PD. National Cholesterol Education Program recommendations for measurement of high-density lipoprotein cholesterol: executive summary. Clin Chem 1995;41:1427-1433.



The following articles in journals at HighWire Press have cited this article:


Home page
Eur Respir JHome page
J. Borgel, B. M. Sanner, A. Bittlinsky, F. Keskin, N. K. Bartels, N. Buechner, A. Huesing, L. C. Rump, and A. Mugge
Obstructive sleep apnoea and its therapy influence high-density lipoprotein cholesterol serum levels
Eur. Respir. J., January 1, 2006; 27(1): 121 - 127.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
G. R. Warnick, M. Nauck, and N. Rifai
Evolution of Methods for Measurement of HDL-Cholesterol: From Ultracentrifugation to Homogeneous Assays
Clin. Chem., September 1, 2001; 47(9): 1579 - 1596.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
M. L. Sampson, A. Aubry, G. Csako, and A. T. Remaley
Triple Lipid Screening Test: A Homogeneous Sequential Assay for HDL-Cholesterol, Total Cholesterol, and Triglycerides
Clin. Chem., March 1, 2001; 47(3): 532 - 539.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
A. M. Wagner, J. L. Sanchez-Quesada, A. Perez, M. Rigla, M. Cortes, F. Blanco-Vaca, and J. Ordonez-Llanos
Inaccuracy of Calculated LDL-Cholesterol in Type 2 Diabetes: Consequences for Patient Risk Classification and Therapeutic Decisions
Clin. Chem., November 1, 2000; 46(11): 1830 - 1832.
[Full Text] [PDF]


Home page
Clin. Chem.Home page
M. L. Sampson, G. Csako, and A. T. Remaley
Estimation of Serum Apolipoprotein B by a Modified Homogeneous Assay for HDL-Cholesterol,
Clin. Chem., June 1, 2000; 46(6): 869 - 871.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (17)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nauck, M.
Right arrow Articles by Wieland, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Nauck, M.
Right arrow Articles by Wieland, H.
Related Collections
Right arrow Lipids, Lipoproteins, and Cardiovascular Risk Factors
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS