| Abstract |
|---|
|
|
|---|
Methods: A quantitative point-of-care test system (Stratus CS®; Dade-Behring) for cTnI with an incorporated centrifuge was evaluated in 412 patients with chest pain less than 12 h.
Results: Results were available within 15 min. CVs were 4.5% at
0.1 µg/L, 4.2% at 0.25 µg/L, and 6.5% at 0.82 µg/L. The
detection limit was 0.01 µg/L. The 97.5% percentile in a
healthy population was 0.08 µg/L. Based on ROC curve analysis,
a threshold of 0.15 µg/L was calculated for the detection of acute
myocardial infarction (AMI). With it, sensitivity for the detection of
patients with AMI (n = 62) was 63% at arrival and 98% after
4 h (Stratus II®, 48% and 85%, respectively;
P <0.01). In 42% of patients with unstable angina
(n = 121), cTnI was
0.08 µg/L (Stratus II, 28%;
P <0.01). During 30 days, death or AMI occurred in
25.5% of these cTnI-positive vs 2.9% of cTnI-negative patients
(Stratus II, 29.4% vs 5.8%).
Conclusion: The Stratus CS provided better analytical performance and comparable or better prognostic information than the Stratus II.
| Introduction |
|---|
|
|
|---|
The first commercially available ELISA test system for determination of cTnI was presented in 1995 (Stratus II®). The Stratus II has been shown to provide sufficient analytical sensitivity for the detection of minor myocardial injury in patients with acute coronary syndrome (10)(24). In this prospective study, we evaluated the new Stratus CS® test system assumed to combine the advantage of point-of-care testing with the analytical performance of second-generation quantitative ELISA technology. We investigated this new test system in a routine setting of patients presenting with acute chest pain in the emergency room.
| Materials and Methods |
|---|
|
|
|---|
The Stratus CS is a fluorometric enzyme immunoassay analyzer for quantitative determination of the cardiac marker CK-MB mass, myoglobin, and troponin I only. The test system is designed to analyze closed routine sample tubes containing anticoagulated whole blood [lithium-heparin (Li-heparin)]. Alternatively, this test system can analyze preprocessed plasma specimens placed into sample cups. For separation of the plasma from whole blood samples, the centrifugation step is incorporated into the test system. Tests to be performed are selected by introducing unitized, bar-coded test packs. Up to four test packs can be introduced for each sample. All required reagents are enclosed within the test packs, which are transferred by disposable pipette tips. Dilutions can be performed automatically by adding test-specific dilution packs. The test system utilizes radial partition immunoassay technology, which has been enhanced through the use of monoclonal capture antibody coupled to Starburst® dendrimers. These dendrimers are well-defined polymers composed of polyamidoamine groups, distinguished by uniformity in molecular size and shape (25). The cTnI-specific antibodies used are identical to those used on the Stratus II system (26) and likewise are capable of detecting both free and complexed cTnI. The dendrimer technology provides for better presentation and functionality of the capture antibody on the glass fiber solid phase surface used in the assay. This in turn leads to more efficient capture of the target antigen. The assay process is initiated by applying the dendrimer-antibody reagent onto the glass fiber matrix to form a reaction zone, which serves to capture the analyte of interest. Centrifuged plasma is then added, followed by the first incubation period. Thereafter, the alkaline phosphatase-labeled second antibody is applied to the matrix, followed by a second incubation period. The unbound, labeled antibody fraction is removed from the reaction zone by radial elution using the substrate-wash reagent. Captured phosphatase-labeled antibodies convert the included enzyme-substrate into a fluorescent product; this permits quantification of the cardiac marker by front surface fluorescence measurement. Analysis on the Stratus CS was performed by six specially trained medical and paramedical collaborators in the emergency room. For the evaluation of analytical performance, centrifuged plasma samples were used; in the clinical study, whole blood samples were analyzed at the point of care.
The Stratus II, a two-step ELISA with a detection limit of 0.35 µg/L for cTnI, has been described elsewhere (10)(24). According to previous studies, values above 0.7 µg/L were considered positive, and a diagnostic threshold of 1.5 µg/L for the diagnosis of myocardial infarction was used (27)(28).
In parallel CK-MB mass and myoglobin were quantified by means of the Stratus II test system. Their upper limits of normal are 5.0 µg/L and 110 µg/L, respectively. All biochemical analyses were conducted on the Stratus II. In addition, CK and CK-MB enzyme activity were measured routinely in the emergency laboratory (Hitachi 717; Boehringer, Mannheim; diagnostic threshold value for CK, 80 U/L in males and 70 U/L in females).
evaluation of analytical performance
The Stratus CS was evaluated relative to the Stratus II regarding:
20%) by a stepwise dilution series of
plasma samples from patients with minor myocardial injury;
clinical evaluation in the emergency room
From August to December 1998, 412 consecutive patients (143
females and 269 males; ages, 59.4 ± 9.1 year) were enrolled at
the emergency room of the University Hospital Hamburg, Germany. The
study inclusion criterion was chest pain lasting <12 h before arrival,
as possibly associated with an acute coronary syndrome. The mean
duration of symptoms at arrival was 4.8 ± 3.5 h; chest pain
continued in 48% of the patients when enrolled in the study.
The study exclusion criterion was evidence of characteristic ST-segment elevations on the baseline electrocardiogram. Patients having myocardial infarction or coronary revascularization during the preceding 14 days were also excluded because of the possible continuous release of troponins after myocardial infarction or potential myocardial injury during revascularization.
Plasma samples anticoagulated with Li-heparin, Na-heparin, and EDTA, respectively, were collected within 15 min after the patient's arrival in the emergency room and additionally 4 h later as a supplement to the routine blood collection. For patients arriving within 2 h after onset of symptoms, the second blood sample was shifted to 6 h after arrival. Whole blood samples were analyzed imme-diately for cTnI with the Stratus CS system at the point of care. Additional plasma samples were centrifuged within 15 min after collection and immediately analyzed for cTnI, CK-MB mass, and myoglobin with the Stratus II system.
definition of clinical diagnoses
AMI at the time of arrival was defined according to the WHO
criteria (29). We included patients with a final diagnosis
of AMI, defined by typical increase of CK enzyme activity greater than
twice the upper limit of normal associated with CK-MB activity
6% of
CK activity during the first 24 h after onset of symptoms.
Unstable angina was defined according to the Braunwald classification
(30). Of 121 patients with a clinical diagnosis of unstable
angina, 69 patients had a history of CHD. For 62 of these patients,
coronary angiography was performed during the index hospital stay. All
remaining 52 patients without history of CHD underwent coronary
angiography; functional CHD was documented in 98% of this subgroup.
Control patients were those without evidence of CHD. These patients had no ST-segment changes during serial electrocardiographic recordings in the emergency room and a negative treadmill test, stress echocardiography, or angiogram (42%). Additionally, during a 30-day follow-up period, the control patients had no occurrences of major cardiac events (death, nonfatal AMI, urgent revascularization). None of these control patients returned to the hospital because of a recurrence of chest pain. Thus, these control patients represent a subgroup with very low cardiac risk (15).
statistical analysis
All results for continuous variables are expressed as means
± SD; comparison between two subgroups was performed by the
MannWhitney U-test (two-sided). A comparison of
categorical variables was generated by the Fisher exact test. Linear
regression analysis was performed on a single test from each sample
over the dynamic range of the assays. ROC analysis was used for
clinical test evaluation; results are expressed as areas under the
individual ROC curves, including 95% confidence intervals.
Significance was set at P values <0.05. Measured
concentrations of cardiac markers were classified as positive in the
case of results above or equal to the chosen threshold value and as
negative for all samples values below threshold.
| Results |
|---|
|
|
|---|
The lowest cTnI concentration as measured with acceptable precision (CV
20%), was 0.8 µg/L for the Stratus II and 0.03 µg/L for the
Stratus CS.
Detection limit.
The detection limit (defined as the lowest
cTnI concentration creating a test signal higher than the mean + 2 SD
of 20 replicates of the zero calibrator) for the Stratus II was 0.5
µg/L (one-third of the diagnostic threshold value) and for the
Stratus CS was 0.01 µg/L (one-twentieth of the diagnostic threshold
value).
analytical effect of delayed cTnI DETERMINATION
Serial blood samples were analyzed for cTnI with delays measuring
up to 12 h after blood collection. For each of five patients
providing different concentrations of cTnI (Stratus CS, 0.050.5
µg/L), a total of seven plasma samples anticoagulated with Li-heparin
plasma were collected. Determination of cTnI in duplicate was performed
at each time point (1, 2, 3, 4, 6, 8, and 12 h),
with samples maintained at room temperature. No differences were
observed for delays of up to 2 h, whereas significantly lower
values were obtained after 8 h. At 12 h, the measured cTnI
values were ~50% of the original cTnI values for samples with
0.1
µg/L cTnI (Fig. 1
). For samples with cTnI values <0.1 µg/L, after
6 h, the cTnI concentration was below the detection limit.
|
stratus cscorrelation with different analytical materials
As the clinically most important range, increasing concentrations
of cTnI between 0.05 and 5 µg/L (Stratus CS) in plasma samples with
different anticoagulants were compared. Linear regression analysis
(including 95% confidence intervals for slope and intercept) indicated
no differences between plasma samples (n = 236) for all
tested anticoagulants as analyzed by the Stratus CS (Fig. 2
and
Fig. 3
):
|
|
comparison of results with stratus ii
The Stratus CS and Stratus II were compared in a total of 786
samples. Values below the detection limit of each test system were
encoded as zero. For samples with cTnI concentrations <4 µg/L, the
relationship (Fig. 4
, top) was:
|
Stratus CS = 0.34 (0.320.36) Stratus II + 0.06 (0.030.09); r2 = 0.62
In contrast, the relationship for cTnI values >4.0 µg/L was:
Stratus CS = 0.99 (0.961.01) Stratus II - 1.72 (-3.39 to -0.02); r2 = 0.98
For all analyzed samples the relationship (Fig. 4
, bottom) was:
Stratus CS = 0.97 (0.950.99) Stratus II + 0.04 (-0.29 to -0.09); r2 = 0.98
Accordingly, the major differences between both test systems were detected for cTnI values below 0.5 µg/L, as analyzed by the Stratus CS. In only one sample with a cTnI value above 0.7 µg/L as analyzed by the Stratus II were the test results with the Stratus CS below 0.08 µg/L. However, in 6.5% of the samples with cTnI values above 0.08 µg/L as analyzed by the Stratus CS, cTnI concentrations obtained with the Stratus II showed no increase above the threshold value of 0.7 µg/L.
non-cardiac chest pain patients
To calculate a diagnostic threshold value that would distinguish
those patients with myocardial injury from those without, we drew fresh
plasma samples from 161 patients with non-cardiac chest pain.
Additionally, thawed plasma samples collected from 218 patients meeting
an identical inclusion criterion were also analyzed. The cTnI
concentration obtained with the Stratus CS (0.024 ± 0.029 µg/L)
was significantly lower than the value obtained with the Stratus II
(0.51 ± 0.34 µg/L; P = 0.002; Fig. 5
). Accordingly, for clinical decision-making regarding patients
with acute chest pain, cTnI values
0.08 µg/L (mean + 2 SD) as
analyzed by the Stratus CS were set as positive.
|
clinical evaluation in patients with acute chest pain
Of 412 consecutive patients presenting with acute chest pain,
15.1% had AMI (as based on a typical increase of CK enzyme activity
within 24 h after arrival), 29.3% had unstable angina, and 9.0%
had stable angina. In 1.0% of the patients (n = 4), pulmonary
embolism was diagnosed, in 2.2% acute heart failure (n = 9) was
diagnosed, and in 0.7% myocarditis (n = 3) was diagnosed. No
evidence for CHD was found in 39.1% of the patients.
In 30.6% of these patients with chest pain, at least one blood sample
was positive with the Stratus CS (
0.08 µg/L). In 24.8% of the
patients, cTnI was above 0.7 µg/L for the Stratus II. Initial testing
on admission identified only 65% and 60%, respectively, of the
positive cTnI test results. CK-MB mass (as analyzed by the Stratus II)
was increased in 13.6% of the patients on arrival and in 19.2% of the
patients 4 h later. In 14.6% of the patients on arrival and in
19.4% of the patients after 4 h, myoglobin was above the upper
limit of normal.
diagnostic sensitivity for the detection of ami
The ROC curves for cTnI, as analyzed by both the Stratus CS and
the Stratus II, are shown in Fig. 6
. Analysis was based on all of the 412 patients with chest pain
for <12 h. For cTnI test results at the time of arrival as analyzed by
the Stratus CS, the area under the curve (0.859) was significantly
greater than that of cTnI on the Stratus II (0.813; P
<0.001), CK-MB mass (0.726; P <0.001), and myoglobin
(0.771; P <0.001). At a diagnostic threshold value of 0.15
µg/L, 63% of the patients with myocardial infarction (39 of 62) were
identified at the time of arrival (3.2 ± 2.9 h after onset
of symptoms regarding AMI patients). Taking into account those patients
with unstable angina and "false positive" findings of increased
cTnI, the corresponding clinical specificity was 92.6%. With the
sample collected 4 h after a patient's arrival, cTnI was above
this diagnostic threshold in 98% of the patients (and specificity was
89.5%). Only one patient had a cTnI concentration below 0.15 µg/L.
The measured cTnI concentration of 0.10 µg/L, however, was already
above the calculated threshold value of 0.08 µg/L for the detection
of minor myocardial injury. For the Stratus II, 48% of patients on
arrival and 85% of patients within 4 h after arrival had cTnI
concentrations above the established diagnostic threshold value of 1.5
µg/L, with corresponding specificities of 95.2% and 93.3%,
respectively. For a threshold value of
0.7 µg/L, the diagnostic
sensitivity increased to 52% and 90%, respectively, (not significant
for both) and specificity decreased to 93.6% and 91.5%, respectively
(not significant for both).
|
CK-MB mass and myoglobin were positive on arrival in 48% and 65% of the patients, respectively. For the second sample, CK-MB mass and myoglobin were increased in 90% and 97% of the patients, respectively. However, two patients with a negative test for myoglobin in the second blood collection had been positive for myoglobin in the first blood sample, yielding a 100% cumulative sensitivity.
minor myocardial injury
Among 121 patients with unstable angina, the measured cTnI
concentrations were 1.18 ± 1.34 µg/L (range, 0.0114.18
µg/L) for the Stratus CS, 1.4 ± 3.32 µg/L (0.416.5 µg/L)
for the Stratus II, 62.2 ± 49.5 µg/L (0.011.03 µg/L) for
myoglobin, and 2.21 ± 3.75 µg/L (0.39.12 µg/L) for CK-MB
mass. We used the mean cTnI value + 2 SD of a healthy population as the
threshold value for minor myocardial injury. Accordingly, cTnI was
0.08 µg/L for 45% of the patients with the Stratus CS; cTnI was
0.7 µg/L for 28% with the Stratus II. CK-MB mass was increased in
only 8.3% of patients; myoglobin was increased in 5.0% of the
patients with unstable angina. Furthermore, serial blood samples during
an extended period of 24 h were collected for two particular
patients suffering from angina pectoris at rest and arriving within
2 h after onset of symptoms. The measured concentrations of cTnI
and CK-MB mass for these two patients are depicted in Fig. 7
. Serial measurements of CK-MB enzyme activity remained within
the reference interval, ruling out non-Q-wave myocardial
infarction in these patients.
|
During the 30-day follow-up period, death or nonfatal AMI was documented in 25.9% of the patients with unstable angina and a positive cTnI test result within 4 h after arrival (as analyzed by Stratus CS); with the Stratus II, cardiac events were observed in 29.4% of the patients. In contrast, for patients with all negative cTnI test results, only 1.5% (n = 1) sustained a major cardiac event during the 30-day follow-up period, whereas for 5.8% of the cTnI-negative patients (n = 5) as analyzed by the Stratus II suffered death or myocardial infarction. The one patient with a negative cTnI test (as analyzed by the Stratus CS) had an AMI on the 27th day after presentation with chest pain in the emergency room.
Other patients with a positive cTnI test result had pulmonary embolism (Stratus CS, n = 1; Stratus II, n = 1), congestive heart failure (n = 2; n = 0), suspected myocarditis (n = 2; n = 1), or unexplained increases in cTnI (n = 1; n = 1).
In 11 patients, CK-MB mass was increased without cTnI release (as analyzed by both cTnI test systems); in 14 patients, myoglobin was increased without detectable cTnI. There was no evidence in any of these patients of CHD, nonischemic myocardial injury, or cardiac event (death, AMI, revascularization) during the 30-day follow-up period.
| Discussion |
|---|
|
|
|---|
4.0 µg/L).
Furthermore, this study documents that in terms of lower
cTnI concentrations (<4.0 µg/L), the Stratus CS presents major
advantages in analytical performance. The CV for the lowest internal
control (0.1 µg/L) was 4.5% as analyzed by the Stratus CS but >10%
as analyzed by the Stratus II. The detection limit for the Stratus II
was 0.5 µg/L compared with 0.01 µg/L for the Stratus CS. These
striking differences might be related to the improved ELISA design of
the Stratus CS, which uses dendrimer technology. In spite of using
antibodies identical to those used by the Stratus II, the Stratus CS
produces an enhanced test signal and a consistently lower background
signal (25). Accordingly, for lower cTnI
concentrations (<4.0 µg/L), the correlation between the Stratus II
and the Stratus CS was poor (r2 =
0.62). It should be emphasized that the observed differences in
analytical sensitivity between the Stratus CS and the Stratus II are
presumably related less to the instruments themselves than to improved
ELISA technology. Other assays for cTnI capable of such improved
sensitivity might provide comparable results. The clinical evaluation of the Stratus CS in patients presenting with acute chest pain at the emergency room confirmed its higher sensitivity for the detection of myocardial infarction. A diagnostic threshold of 0.15 µg/L for the Stratus CS was calculated on the basis of 412 patients with chest pain. With the threshold set on this value, diagnostic sensitivity for the detection of AMI was superior to that of cTnI determination and CK-MB mass measurements, both analyzed by the Stratus II. In our study population, the diagnostic sensitivity was comparable to that of myoglobin. However, it should be noted that patients with AMI arrived in the emergency room 3.2 ± 2.9 h after onset of chest pain. Only 18% of the patients arrived within 3 h after onset of symptoms. A higher rate of early arriving patients with AMI might have produced significant differences between the two markers, myoglobin and cTnI. Four hours after arrival, for 98% of the patients with a final diagnosis of myocardial infarction, cTnI values were above the threshold value of 0.15 µg/L as analyzed by the Stratus CS. For myoglobin, at least one positive test result was obtained in these patients during this observation period of 4 h.
Only 8% of patients with unstable angina had increased CK-MB
concentrations according to the predefined threshold value of 5 µg/L;
in 5% of these patients, myoglobin was increased. By using a threshold
value of 0.08 µg/L for cTnI (as analyzed by the Stratus CS) as based
on the mean cTnI value + 2 SD of a healthy population, we found
increased cTnT in 45% of patients with unstable angina; for the
Stratus II, in 28% of the patients cTnI was
0.7 µg/L. This higher
rate of positive tests for the Stratus CS was connected with a higher
negative predictive value during the 30-day follow-up period (98.5% vs
94.2%; P <0.01). No cardiac events occurred during the
initial 3 weeks after presentation in these patients with <0.08 µg/L
cTnI release. It is important to emphasize that the evaluation of
different threshold values was performed in patients with non-cardiac
chest pain. Several previous studies have shown that the troponin
values in these patients are higher than those in healthy volunteers.
With respect to the excellent sensitivity of the Stratus CS, it should
be noted that all analyses were performed with fresh blood samples
immediately after sample collection, which might have increased
sensitivity for the detection of minor increases of troponin I.
point-of-care testing
Troponins can be determined both by qualitative rapid bedside
tests and by quantitative ELISA test systems. Off-site quantitative
determination of troponins has the disadvantages of delayed
availability of test results and higher fixed costs. Therefore, rapid
test systems have been developed based on monoclonal antibodies and
chromatographic immunologic solid-phase technology
(11)(26). These bedside tests can be performed
at the point of care by medical or paramedical staff without
additional equipment. The tests provide qualitative test results
(positive/negative) within 1520 min without time-consuming
preparations such as clotting and centrifugation. Rapid testing for
troponins has been shown to provide high prognostic value
(15).
However, recent bedside tests pose limitations for cardiac marker determination: qualitative rapid testing for troponins remains subject to reading errors, with substantial interobserver variability. The utilization of rapid tests requires considerable familiarity with training samples. The rapid tests must be read within an adequate time interval and under adequate illumination (11)(26). To overcome these disadvantages, quantitative whole blood test systems for point-of-care testing have been introduced. Recently, a quantitative whole blood system for cTnI, CK-MB mass, and myoglobin was shown to provide sensitivity and specificity for the diagnosis of AMI comparable to established quantitative test systems (31). Determination of cTnI provided highest diagnostic efficacy, and neither serial nor parallel analysis of the three markers increased sensitivity or specificity for the detection of myocardial infarction. In the current study, we were able to show similarly that the new Stratus CS (including a centrifugation step for whole blood samples) is easy to handle and provides reliable cTnI test results while preserving the advantages of a rapid test system. For both qualitative rapid tests and the Stratus CS, test results are available within 15 min after blood collection. With respect to reliability and maintenance of a test system, it might be preferable on a long-term basis to perform cTnI measurements in the laboratory with a large-scale analyzer that provides cost-effective analysis of multiple analytes. Although point-of-care testing with the Stratus CS was uncomplicated during the study period of 5 months, the long-term usage of point-of-care test systems remains controversial.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
K. M. Eggers, T. Kempf, T. Allhoff, B. Lindahl, L. Wallentin, and K. C. Wollert Growth-differentiation factor-15 for early risk stratification in patients with acute chest pain Eur. Heart J., October 1, 2008; 29(19): 2327 - 2335. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Heeschen, S. Dimmeler, C. W. Hamm, E. Boersma, A. M. Zeiher, M. L. Simoons, and on Behalf of the CAPTURE (c7E3 Anti-Platelet Thera Prognostic Significance of Angiogenic Growth Factor Serum Levels in Patients With Acute Coronary Syndromes Circulation, February 4, 2003; 107(4): 524 - 530. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. T. Sandri, D. Cardinale, L. Zorzino, R. Passerini, P. Lentati, A. Martinoni, G. Martinelli, and C. M. Cipolla Minor Increases in Plasma Troponin I Predict Decreased Left Ventricular Ejection Fraction after High-Dose Chemotherapy Clin. Chem., February 1, 2003; 49(2): 248 - 252. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. Kim, O. F. Laterza, K. G. Hock, J. F. Pierson-Perry, D. M. Kaminski, M. Mesguich, F. Braconnier, R. Zimmermann, M. Zaninotto, M. Plebani, et al. Performance of a Revised Cardiac Troponin Method That Minimizes Interferences from Heterophilic Antibodies Clin. Chem., July 1, 2002; 48(7): 1028 - 1034. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Venge, B. Lindahl, and L. Wallentin New Generation Cardiac Troponin I Assay for the Access Immunoassay System Clin. Chem., May 1, 2001; 47(5): 959 - 961. [Full Text] [PDF] |
||||
![]() |
J.-P. Chapelle, M.-C. Aldenhoff, L. Pierard, and J. Gielen Comparison of Cardiac Troponin I Measurements on Whole Blood and Plasma on the Stratus CS Analyzer and Comparison with AxSYM Clin. Chem., November 1, 2000; 46(11): 1864 - 1866. [Full Text] [PDF] |
||||
![]() |
F. S. Apple, F. P. Anderson, P. Collinson, R. L. Jesse, M. C. Kontos, M. A. Levitt, E. A. Miller, and M. M. Murakami Clinical Evaluation of the First Medical Whole Blood, Point-of-Care Testing Device for Detection of Myocardial Infarction Clin. Chem., October 1, 2000; 46(10): 1604 - 1609. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Altinier, M. Mion, A. Cappelletti, M. Zaninotto, and M. Plebani Rapid Measurement of Cardiac Markers on Stratus CS Clin. Chem., July 1, 2000; 46(7): 991 - 993. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |