|
|
||||||||
Articles |
1
Department of Chemical Pathology, University of Pretoria, P.O. Box 2034, Pretoria 0001, South Africa.
2
Severimed, Wiedaustrasse 202, 48163 Münster,
Germany.
3
Total homocyst(e)ine refers to the sum of the concentrations of free homocysteine, protein-bound homocysteine, the disulfide homocystine, and the mixed disulfide homocysteine-cysteine.
a Author for correspondence. Fax 27-12-3283600; e-mail jubbink{at}medic.up.ac.za
| Abstract |
|---|
|
|
|---|
Methods: Results from three different methods [HPLC with fluorescence detection, enzyme immunoassay (EIA), and fluorescence polarization immunoassay (FPIA)] to determine fasting (n = 163) and post-methionine load (n = 80) plasma tHcy concentrations were compared with those obtained by gas chromatographymass spectrometry (GC-MS). Difference plots on non-transformed and log-transformed data were used to assess the agreement between HPLC and GC-MS, EIA and GC-MS, and FPIA and GC-MS.
Results: The closest agreement between methods was observed between GC-MS and FPIA for fasting tHcy concentrations, with 95% of the FPIA values between 19% above and 24% below the corresponding GC-MS results. Post-methionine load tHcy concentrations measured by EIA showed the least agreement with GC-MS, with 95% of values measured by EIA ranging between 52% above and 16% below the GC-MS values. With respect to GC-MS, the above-mentioned methods showed a negative bias for fasting tHcy concentrations, but a positive bias for both immunoassays for post-methionine load tHcy concentrations.
Conclusions: The agreement among methods is insufficient to allow them to be used interchangeably. The intermethod differences emphasize the need for standardization of plasma tHcy assays.© 1999 American Association for Clinical Chemistry
| Introduction |
|---|
|
|
|---|
Chromatographic methods have been the methods of choice to determine plasma tHcy concentrations. Most laboratories have used methods based on the derivatization of homocysteine with thiol-specific reagents such as monobromobimane (7) or ammonium 7-fluorobenzo-2-oxa-1,3-diazole-4-sulfonate (SBDF) (8), or by o-phthaldialdehyde derivatization of the primary amine group (9). The fluorescent Hcy adduct is then separated from other thiol-containing compounds by HPLC and quantified by fluorescence detection. HPLC methods based on SBDF derivatization seem to be the most popular; data from the European Quality Assessment Scheme for Special Assays in Serum and Urine indicate that 38 of 90 participating laboratories used derivatization with SBDF to measure serum tHcy concentrations (10). Gas chromatographic methods to measure tHcy have also been described (11), and a few specialist centers measure tHcy by gas chromatographymass spectrometry (GC-MS) (12).
The chromatographic methods mentioned above require sophisticated and expensive equipment that generally is not available at routine clinical chemistry laboratories. Furthermore, few chromatographic methods are fully automated, and in all cases, the daily output is relatively low. In contrast, immunoassays usually lend themselves to full automation and also have the potential of a high daily throughput. Therefore, the recently described enzyme-linked immunoassay (EIA) (13) and fluorescence polarization immunoassay (FPIA) methods (14) for plasma tHcy may become popular with laboratories that offer this assay. Both EIA and FPIA rely on enzymatic conversion of homocysteine to S-adenosylhomocysteine, which is subsequently detected by a competitive immunoassay. Both methods have become commercially available.
In this study, we compared a GC-MS method for plasma tHcy with an HPLC method based on thiol derivatization with SBDF, an EIA method, and an FPIA method. This between-method comparison reveals that the above-mentioned methods cannot be used interchangeably and emphasizes the need for standardization of plasma tHcy assays.
| Materials and Methods |
|---|
|
|
|---|
The GC-MS and the HPLC assays were calibrated against calibrators prepared independently from crystalline L-homocystine and D,L-homocysteine, respectively. Both were obtained from Sigma Chemicals. Both the EIA and the FPIA were calibrated against calibrators supplied by the respective reagent kit manufacturers. Control specimens were analyzed in each batch, and results were only accepted when the control values were within the range specified by the manufacturer. Controls prepared in house were used for both the GC-MS and the HPLC methods.
statistical methods
Difference plots were used to assess the agreement between tHcy
results obtained with GC-MS vs HPLC, EIA, and FPIA, respectively
(16)(17). Fasting and post-methionine load tHcy
concentrations were assessed separately. Possible systematic biases
between GC-MS vs HPLC, GC-MS vs EIA, and GC-MS vs FPIA, respectively,
were assessed by computing the 95% confidence intervals for the mean
differences between GC-MS and each of the methods mentioned above.
In a subsequent analysis, the data from each assay were log transformed. Mean differences, as well as the limits of agreement (mean difference ± 2 SD), were calculated on the log-transformed data for GC-MS vs HPLC, GC-MS vs EIA, and GC-MS vs FPIA. Antilogs of the mean differences were calculated to assess the mean proportional bias of each method with respect to GC-MS. Antilogs of the limits of agreement were calculated to express these limits as ratios of GC-MS results vs HPLC, EIA and FPIA results, respectively (16)(17).
| Results |
|---|
|
|
|---|
|
|
Scatter plots of observed measurement differences (16) for
fasting tHcy determinations against the mean of GC-MS and the method
used in the assay are shown in Fig. 1
; Fig. 2
is similar to Fig. 1
, except that the scatter plots are derived
from plasma tHcy concentrations obtained after methionine loading. The
mean (SD) differences between GC-MS and HPLC, GC-MS and EIA, and GC-MS
and FPIA are reported in Table 3
. Using the standard errors of the mean differences, the 95%
confidence intervals were computed; these showed a negative bias for
HPLC, EIA, and FPIA with respect to GC-MS when fasting tHcy
concentrations were compared. This negative bias was the smallest for
FPIA. For post-methionine load tHcy concentrations, HPLC showed a
negative bias similar to the bias observed for fasting tHcy
concentrations; however, both immunoassays now demonstrated a positive
bias (Table 3
).
|
|
|
The central 0.95 interval (mean of the differences ± 2 SD) gives
an indication of the agreement between GC-MS and the other methods used
to measure plasma tHcy concentrations (16). Using this
approach, FPIA agreed the best with GC-MS for both fasting and
post-methionine load tHcy concentrations (Table 3
). Both HPLC and EIA
displayed a relatively wider scatter of difference data points.
However, this comparison may not be totally appropriate because there
appears to be a relationship between the difference and the mean for at
least certain scatter plots (HPLC vs GC-MS and EIA vs GC-MS). The data,
therefore, were log transformed, and the mean and SD values from the
log-transformed data set were used to calculate the limits of agreement
(with respect to GC-MS) as described by Bland and Altman
(16). The limits of agreement were then anti-logged and
expressed as intervals (ranges of percentages) by which 95% of the
tHcy determinations measured by HPLC, FPIA, and EIA, respectively, were
expected to differ from GC-MS (Table 3
). When GC-MS vs FPIA was
compared with GC-MS vs HPLC, it became apparent that the
above-mentioned ranges of percentages for the two comparisons were
virtually the same size for fasting tHcy concentrations, but that HPLC
showed a larger negative mean proportional bias (the antilog of the
mean difference calculated from log-transformed data) than FPIA. The
agreement between EIA and GC-MS was considerably less than that found
for the FPIA and HPLC.
For post-methionine load tHcy concentrations, the range of percentages
that contains 95% of the data points in the comparison with GC-MS was
the narrowest for HPLC, followed by FPIA and then EIA. The mean
proportional bias of HPLC vs GC-MS was very similar to that observed
for fasting tHcy concentrations. The mean proportional biases for FPIA
vs GC-MS and EIA vs GC-MS, respectively, became positive,
producing a considerable increase in the upper limits of agreement when
the immunological assays were compared with GC-MS (Table 3
).
| Discussion |
|---|
|
|
|---|
Compared with the other methods, the FPIA assay had the lowest analytical CV, at 4.5%. Considering that the intraindividual biological variation for tHcy has been reported as 9.4% (19), the FPIA tHcy assay was the only one that fulfilled the criterion of an analytical CV at least 50% lower than the intraindividual CV (20).
Universally accepted reference material for homocysteine is not yet
available, which implies that none of the four methods were calibrated
against appropriately certified reference materials. Both
chromatographic methods used calibrators prepared independently: The
GC-MS calibrator was prepared in Denver, CO; and the HPLC calibrator
was prepared in Pretoria, South Africa. The differences between HPLC
and GC-MS presumably are explained by differences in
calibrators. Furthermore, Dudman et al. (21) recently
showed that the SBD-Hcy adduct is light sensitive and that exposure of
this compound to fluorescent laboratory light may produce lower
estimates of tHcy concentrations. When we performed the HPLC assays in
1995, this was not yet known, and it is possible that destruction by
light could have contributed to the lower tHcy concentrations observed
with HPLC. This may be particularly true for those samples analyzed
just before recalibration. It should be noted that the analytical
imprecision induced by possible light destruction is
incorporated in the CV for the quality-control samples (Table 1
)
because the quality-control samples were always inserted randomly in
the daily sequence of samples awaiting HPLC analysis. The analytical CV
of the HPLC method compares well with that of GC-MS, indicating that
the possible effect of light destruction on the quality of the HPLC
results is probably only of minor importance in explaining the negative
bias of HPLC vs GC-MS.
The fasting plasma tHcy concentrations measured with both immunoassays
were lower than those measured with GC-MS. In contrast to HPLC, the
results for both immunoassays were higher than the GC-MS results for
post-methionine load tHcy concentrations. For GC-MS vs FPIA, a shift of
~11% in the 95% range of agreement was noted. For GC-MS vs EIA, the
scatter of data became wider (Table 3
). There may be several
explanations for the deviations in the limits of agreement between the
immunoassays and GC-MS after methionine loading. The independent
selection of calibrators for each method, differences in linearity
between methods, or the lack of standardization may explain the
observed deviations in the limits of agreement. It is also possible
that methionine loading increased the concentrations of a
cross-reactant in the circulation. Although the mean peak plasma
methionine concentrations increased 25-fold, to 0.6 mmol/L, after
methionine loading in this study (15), the reported
interference by methionine was negligible up to 5 mmol/L for both EIA
and FPIA, respectively (13)(14). It is therefore
unlikely that methionine was a cross-reactant in our study, but the
appearance of other, as yet unknown, cross-reactants after methionine
loading cannot be excluded.
In this study, each method used its own calibrator. It may be expected that intermethod agreement will improve when aliquots of the same set of calibrators are used. However, this does not happen in practice, and our study serves as an indication of currently existing interlaboratory and intermethod differences that should be taken into account when results from different centers are interpreted. Although 95% of the fasting tHcy concentrations measured by FPIA fell in the interval of 18.6% above or 24.2% below those measured by GC-MS, this interval may still be too large to allow these methods to be used interchangeably. The interval will probably become smaller if the two methods use the same set of calibrators. Our results indicate that GC-MS and EIA assays for plasma tHcy should not be used interchangeably, and thus support a similar conclusion made by Frantzen et al. (13) in their comparison of EIA and HPLC. We also consider the limits of agreement between GC-MS and HPLC too wide to allow these two methods to be used interchangeably.
We conclude that FPIA and HPLC show better agreement with GC-MS than the EIA method. With the possible exception of GC-MS and FPIA for fasting tHcy concentrations, none of the methods should be used interchangeably. Certified reference material is urgently required to improve intermethod and interlaboratory agreement.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
H. J. Huijgen, F. P.W. Tegelaers, C. H.H. Schoenmakers, C. J. Pronk-Admiraal, and S. Ekema Multicenter Analytical Evaluation of an Enzymatic Method for the Measurement of Plasma Homocysteine and Comparison with HPLC and Immunochemistry Clin. Chem., May 1, 2004; 50(5): 937 - 940. [Full Text] [PDF] |
||||
![]() |
Y. Tan, X. Sun, L. Tang, N. Zhang, Q. Han, M. Xu, X. Tan, X. Tan, and R. M. Hoffman Automated Enzymatic Assay for Homocysteine Clin. Chem., June 1, 2003; 49(6): 1029 - 1030. [Full Text] [PDF] |
||||
![]() |
N. Q. Hanson, J. H. Eckfeldt, K. Schwichtenberg, O. Aras, and M. Y. Tsai Interlaboratory Variation of Plasma Total Homocysteine Measurements: Results of Three Successive Homocysteine Proficiency Testing Surveys Clin. Chem., September 1, 2002; 48(9): 1539 - 1545. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Nexo, F. Engbaek, P. M. Ueland, C. Westby, P. O'Gorman, C. Johnston, B. F. Kase, A. B. Guttormsen, I. Alfheim, J. McPartlin, et al. Evaluation of Novel Assays in Clinical Chemistry: Quantification of Plasma Total Homocysteine Clin. Chem., August 1, 2000; 46(8): 1150 - 1156. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |