|
|
||||||||
Articles |
1
Department of Clinical Biochemistry, St. Bartholomew's and the Royal London School of Medicine and Dentistry, Turner Street, London E1 2AD, UK.
2
Dade Behring Incorporated, Glasgow Research Laboratory,
Bldg. 700, P.O. Box 6101, Newark, DE 19714-6101.
a Author for correspondence. Fax 44 171 377 1544; e-mail c.p.price{at}mds.qmw.ac.uk
| Abstract |
|---|
|
|
|---|
Methods: We studied the Stratus® II assay, the Opus® II assay, the Access® assay, and a research-only cTnI heterogeneous immunoassay that uses the Dade Behring aca® plus immunoassay system equipped with two new noncommercial monoclonal antibodies. Because the aca plus cTnI assay is for research only, we first evaluated and analytically validated it for serum and citrated plasma. Initially, each method was calibrated using the method-specific calibrator supplied by each manufacturer; however, the aca plus cTnI assay was calibrated using patient serum pools containing cTnI and selected on the basis of increased creatine kinase MB isoenzyme and with values assigned by use of the Stratus cTnI assay. For method comparisons, individual patient sample cTnI values were determined and compared with the Stratus II assay.
Results: Passing and Bablock regression analysis yielded slopes of 1.44 (r = 0.96; n = 72) for the Opus II vs Stratus II assays; 0.07 (r = 0.91; n = 72) for the Access vs Stratus II assays; and 0.90 (r = 0.91, n = 72) for the aca plus vs Stratus II assays. The recalibration of each method with a Stratus II-assigned serum pool improved, but did not entirely eliminate, the slope differences between the different assays (range, 1.001.16). The observed scatter in the correlation curves remained.
Conclusion: There is a need to further explore the specificities of these assays with respect to the different circulating forms of cTnI.© 1999 American Association for Clinical Chemistry
| Introduction |
|---|
|
|
|---|
Several studies have demonstrated the specificity and usefulness of cTnI as a biochemical marker of myocardial damage(1)(2)(3)(4)(10)(11)(12)(13)(14)(15)(16). Increased cTnI has not been found in marathon runners, patients with skeletal muscle damage(17), or patients undergoing non-cardiac surgery. cTnI is significantly increased in the serum of patients with acute myocardial infarction (10)(11)(12)(13) and exhibits characteristic release kinetics, being increased ~4.5 h post infarct (vs 6.3 h for the creatine kinase MB mass) and remaining increased for 59 days(18).
After myocardial injury, there is a progressive rise in circulating immunoreactive cTnI. Empirically, the in vivo t1/2 of purified cTnI in dogs was found to be ~67 min (10), which agrees with the renal clearance of a protein possessing this molecular weight. Still unclear, however, is the t1/2 of the immunoreactive form released from damaged myocardial tissue. Approximate calculations, generated from cardiopulmonary bypass studies in which reproducible injuries are made, suggested a much longer t1/2 of several hours (e.g., 7 h), which supported the hypothesis that the major circulating form is complexed. Recent characterization studies have verified that circulating cTnI, released following myocardial damage, represents a combination of intact complex and its separate component parts(19)(20)(21)(22). Other potentially important circulating forms of cTnI may also exist (in particular, phosphorylated, dephosphorylated, reduced, oxidized, and/or proteolytically degraded). For example, cTnI can be phosphorylated on two adjacent N-terminal serine residues (Ser 22 and 23), which have important regulatory roles in muscle contraction(23). These phosphorylated forms (apo-, mono-, and di-), which can be distinguished immunologically (24), influence the stability of the heterotrimer and its resistance to proteolytic degradation by the calcium-dependent protease µ-calpain(25). Similarly, the two cTnI cysteine residues have been shown to exist in either the reduced or oxidized forms, the latter producing an intrachain disulfide bond (21)(22). Although the importance of these factors with respect to the immunoreactivity detected by different cTnI immunoassays has not been fully established, understanding these differences could be important when further characterizing current methods or developing new methods.
To compare and understand cTnI values obtained from four different cTnI methods, we first report the evaluation and analytical validation of immunoreagents developed for a noncommercial application that quantifies cTnI using the Dade Behring aca® plus immunoassay system. The assay is based on chromium dioxide particle technology (26) and uses two mAbs recognizing different epitopes on cTnI than the other commercial cTnI methods. After analytical validation of these immunoreagents, individual patient sample values were obtained and compared in four cTnI immunoassays: the Stratus® II assay, the Opus® II assay, the Access® assay, and the noncommercial Dade Behring aca plus cTnI assay. Finally, the extent of agreement between patient sample values obtained with these methods were compared after each method was recalibrated using human serum pools with cTnI values assigned using the Stratus II cTnI assay.
| Materials and Methods |
|---|
|
|
|---|
aca plus CTNI ASSAY IMMUNOREAGENTS
Two mAbs (designated 144B3.63.5 and 144B5.2.1) were generated for,
and used in, the prototype aca plus assay described in this study.
Initially, to confirm the appropriate cTnI specificity of these
reagents, immunoblot detection with purified mAbs 144B3.63.5,
144B5.2.1, and 3I-35 [known to recognize both cTnI and skeletal TnI
(skTnI); Spectral Diagnostics] as the primary antibodies was
performed. Briefly, either purified human cTnI or human skTnI
(Biodesign International) was loaded into the large well of individual
two-lane 420% gradient sodium dodecyl sulfate-polyacrylamide gel
electrophoresis reducing gels (Novex Electrophoresis). Prestained
protein markers (ISS) were electrophoresed in the second smaller well.
After electrophoresis at a constant 125V for ~2 h, the proteins were
transferred to nitrocellulose (Bio-Rad) at 150 mA for 2 h, using a
TE22 Transphor Electrophoresis Unit (Hoeffer Scientific Instruments).
Immunoblotting was performed with an Immunetics Miniblotter 25 such
that 1 µg of antigen was present in each lane. Retained primary mAbs
were detected with alkaline phosphatase-conjugated goat anti-mouse
antibodies followed by the addition of
5-bromo-4-chloro-3-indolylphosphate/nitro blue tetrazolium substrate.
calibration and quality control
Two types of material were used. The first (calibrator I) was
obtained from the manufacturer and was specific to the test
methodology. The second (calibrator II) was prepared from human serum
samples that were selected on the basis of the creatine kinase MB
isoenzyme concentration. Individual samples were pooled to five
different target concentrations, and each pool was assigned a cTnI
value, using the Stratus II cTnI method. Three quality-control (QC)
pools were prepared using a partially purified cTnI fraction from human
heart extracts (SCIPAC, Sittingbourne, UK). This material was added
into a human serum pool (prescreened for low background cTnI
concentrations using the Stratus II method) at ~1, 35, and 60 mg/L.
Materials were stored aliquoted at -20 °C.
statistical analyses
Passing and Bablock analyses (27) were performed using
the Astute Statistical Package (DDU Software, University of Leeds),
with other statistical analyses performed using Statview II for the
Macintosh (Abacus Concepts).
analytical validation
Assay range, hook effect, and detection limit.
The assay
working range and hook effect were determined by examination of the
assay signal using increasing amounts of partially purified human cTnI,
up to 1000 µg/L. The assay detection limit was defined as two times
the SD of 20 replicates of the zero calibrator.
Imprecision.
The three human serum QC pools
containing 1, 35, and 60 µg/L cTnI that were prepared and frozen in
aliquots were used to assess imprecision. The within-day imprecision
(CV) was determined by analysis of 20 aliquots of each pool randomized
in a single analytical run. The between-day CV was determined by ANOVA
using the NCCLS protocol EP5-T with duplicate analysis of each pool
performed on 20 nonconsecutive days.
Analytical recovery and parallelism.
For recovery experiments,
a human sample with high cTnI was added to 16 low-concentration cTnI
patient samples to increase the baseline concentration by ~5 µg/L,
using one volume of high-concentration sample to nine volumes of
low-concentration sample. All samples were analyzed by both the Stratus
and aca plus methods. To determine parallelism, a series of 10 sera
with increased cTnI concentrations, obtained from the routine
laboratory, were diluted with the calibrator II zero calibrator to give
75%, 50%, 25%, and 12.5% of the original. Each dilution was
analyzed in duplicate with the aca plus application. Linearity was
assessed by linear regression analysis.
Method comparison.
A total of 72 samples were analyzed in
duplicate for cTnI using the aca plus, Stratus, Access, and Opus
methods, each performed according to the manufacturer's instructions.
Interference.
Two approaches were taken to study the potential
interference from skTnI. In the first approach, purified skTnI
(Spectral Diagnostics) was added to a human serum calibrator base pool
at 1000 and 10 000 µg/L. In the second approach, purified skTnI at
10 and 100 µg/L was added to five patient samples, each with high
endogenous cTnI. The cTnI concentrations were measured in all samples
by both the aca plus and Stratus cTnI assays.
Interference from anticoagulants was studied by analysis of matched serum and citrated plasma from 12 patients with increased cTnI.
| Results |
|---|
|
|
|---|
|
No significant cross-reaction was found in the aca plus method when up
to 10 000 µg/L skTnI added into nondiseased human serum was run as a
sample (Fig. 1
). In addition, Western blot analysis of the two antibodies
144B3.63.5 and 144B5.2.1 (both IgG1) is shown in
Fig. 2
. Neither antibody recognized purified skTnI, but 144B3.63.5
recognized several low-molecular mass bands at 22, 16, 14 and 8 kDa,
and a higher molecular mass band at ~40 kDa in the purified cTnI blot
(Fig. 2A
). Western blotting was also performed with both antibodies and
troponin T together with an anti-troponin mAb as a control; neither of
the anti-cTnI mAbs recognized troponin T (data not shown).
|
|
Finally, no interference from citrate as an anticoagulant was demonstrated. No significant difference between 12 matched serum and citrate plasma samples was observed, P >0.1 using a paired t-test (data not shown).
The experimental protocol used in this study yielded a mean analytical recovery of 96.2% ± 17.3% on the aca plus analyzer and 93.9% ± 13.6% on the Stratus II system. Finally, no lack of parallelism was demonstrated, with the slope of the regression line not significantly different from 1 (P >0.05).
Method comparison data are shown in Fig. 3
along with regression analyses using the Passing and Bablock
approach (27). Correlation coefficients were calculated
separately, using standard linear regression (Table 2
). Fig. 4
shows the difference plots for the same comparisons using
BlandAltman scatter analysis (28). To demonstrate the
possible clinical implications of the variability between the
experimental aca plus and Stratus data, we ranked the results as above
or below the cutoff value of 1.5 µg/L: for the 9 samples for which
the aca plus results were <1.5 µg/L, the Stratus assay results were
<1.5 µg/L for 6 samples and >1.5 µg/L for 3 samples; for the 63
samples for which the aca plus results were >1.5 µg/L, the Stratus
assay results were <1.5 µg/L for 5 samples and >1.5 µg/L for 58
samples. The method comparison data were also analyzed using the same
patient samples after each method was calibrated using calibrator II,
as shown in Table 2
. Within-assay imprecision profiles were
calculated for all methods from the 72 duplicate measurements
used in the method comparisons, as shown in Fig. 5
.
|
|
|
|
| Discussion |
|---|
|
|
|---|
As part of the analytical validation of the aca plus cTnI assay, we performed a series of recovery experiments using a human serum pool containing a high concentration of endogenous cTnI, which was added to separate human serum samples. Because there was little published information concerning analytical recovery, we measured the same series of samples using the Stratus assay as the reference method. There was no significant difference in the mean recovery between the two methods: 96.2% ± 17.2% for the aca plus application and 93.9% ± 13.6% for the Stratus method. The SDs for both sets of data, however, indicated significant sample-to-sample variability in recovery. This variability in recovery might be attributed to differences (a) in the stability of the added cTnI in the different samples, as reported by Waskiewicz et al. (29) and Morjana (30); (b) in the relative concentrations of the cTnI forms in the individual samples (31); or (c) in the epitope specificity of the antibodies (32)(33). Katrukha et al. (32), in a careful study using three different combinations of paired mAbs shown to recognize different fragments of cTnI, demonstrated that the protein in serum is susceptible to proteolysis. Similar experiments on the protein extracted from tissue indicated that the central fragment of the protein (residues 30110) was the most stable. The data also demonstrated that the degree of instability observed was method dependent. Katrukha et al.(19) used 15 mAbs against cTnI in a sandwich assay (a total of 196 combinations) to study the TnI released into the bloodstream after a myocardial infarction and found that the majority was released as a complex with troponin C. The authors showed that combination of TnI with cardiac troponin C led to a reduction in the interaction of the antibody pairs, indicating that intermethod variability of results could be attributed to the region of the cTnI recognized by the antibodies and the proportion of free and complexed protein present in the serum. Nevertheless, this observation was consistent with the recovery data, yet would still enable the samples to be diluted out with acceptable linearity.
Method comparison between the aca plus and the Stratus assays showed
excellent agreement, although there was significant scatter around the
regression line (Fig. 3A
). This scatter, when viewed in the context of
a cutoff value of 10 µg/L for the Stratus assay as the reference
procedure, indicates a sensitivity and a specificity of 79% and 86%,
respectively, for the aca plus experimental method. Similar scatter was
also evident when the Stratus method was compared with either the Opus
or the Access methods (Fig. 3
, B
and C
). When each assay was calibrated
with the method-specific calibrator (calibrator I), the aca plus and
Opus methods each showed good correlation with the Stratus assay. The
aca plus assay showed good agreement (slope, 0.90) compared with the
Stratus assay, whereas the Opus assay demonstrated a slope of 1.44
compared with the Stratus assay. A significant correlation slope bias,
however, was observed between the Access and Stratus methods (slope,
0.07). Similar slope differences between the Stratus and Opus and the
Stratus and Access methods were observed in a recently published
abstract (34). Potential explanations for these slope
differences are discussed below.
Unlike Bhayana et al. (34), however, we extended our study
to compare the patient sample results from the same samples, obtained
after calibrating each method with calibrator II. As summarized in
Table 2
, the correlation slopes between the Stratus II and Opus II cTnI
methods and Stratus II and aca plus methods improved to ~1.0. More
importantly, the correlation slope between the Stratus II and Access
analyzers improved significantly, to 1.16. Although the correlation
slopes improved after each method was recalibrated with a Stratus cTnI
assay-assigned patient serum pool, the scatter around the correlation
lines was unchanged and still significant.
BlandAltman difference plots enabled a direct comparison of the bias
and exploration of the concentration-dependent nature of that bias(28). These plots enable differences between methods to be
evaluated, with respect to bias and scatter, in a more robust manner
than regression analysis, even Passing and Bablock. In this study, in
which the measure of agreement between several methods was sought, it
was the only analysis that could facilitate this. If there is no
systematic bias between methods, then the mean difference will be zero.
If there is a fixed systematic bias between methods, then the mean
difference will be more or less than zero. With a
concentration-dependent (slope) bias between methods, the data scatter
will show a nonrandom distribution with a slope significantly different
to zero. Fig. 4A
clearly shows a large degree of scatter around zero,
which in most cases is random. However, there is a high bias for the
Access assay relative to the others. Calibration of each method with
calibrator II largely eliminated this bias, but the scatter essentially
remained. In agreement with recent studies, these method comparison
results suggest that a common reference preparation would improve the
agreement between the available cTnI methods(34)(35) but have little effect on the scatter.
The data around the regression lines (Fig. 4B
) showed significant
scatter, implying that the immunoreaction in each assay is different.
Each assay has inherent precautions against nonspecific sample
interferences and has been shown to be unaffected by such. In addition,
these assays have been reported to be free of interference from skTnI
isoforms. The most likely cause of method-related differences,
therefore, is that each assay (i.e., antibody pair) recognizes the
sample immunoreactive cTnI differently. The antibody pairs used in each
method do not react with the same epitopes on cTnI. According to
published results (8)(9)(36), the
Opus, Access, and Stratus assays recognize distinct epitopes N-terminal
to the TnC binding site. Epitope mapping has determined that the aca
plus antibodies recognize epitopes between residues 80 and 153
(unpublished data). Unique epitope recognition by each antibody pair,
therefore, could lead to different cTnI values being reported because
of variable detection of the many circulating forms of cTnI. For an
example from the present study, Western blotting showed that mAb
144B3.63.5 recognized several lower molecular mass cTnI bands, in
addition to intact cTnI, that were not recognized by mAb 144B5.2.1.
When together in the immunoassay reaction cuvette, these mAbs will
detect only cTnI fragments or complexes recognized by both. Other
examples from the literature have shown that circulating immunoreactive
cTnI is, in fact, a heterogeneous mixture of fragments and complexes(19)(20)(21)(22)(30). The concentrations and
identification of the circulating cTnI forms following myocardial
damage are not yet clearly resolved. Because the three commercial
assays investigated here appear to recognize basically the same
fraction of circulating immunoreactive cTnI, however, the clinical
significance seemingly is minimal. Further studies must be performed
before this is verified.
In conclusion, there was an overall improvement in individual patient cTnI values between the four cTnI methods evaluated in this study, three commercial and one research, when a patient-derived serum pool calibrator was used. Scatter in the correlation plots between the different methods was still observed, presumably because each assay uses immunoreagent pairs that recognize different troponin epitopes. These individual patient sample differences were likely related to the different circulating forms of cTnI and the relative abilities of the different antibodies to recognize them. Despite the fact that different immunoreagent pairs were used, however, the commercial cTnI immunoassays still recognize similar, but not identical, circulating immunoreactive cTnI fraction(s). Further studies to fully characterize the many circulating forms of cTnI are warranted.
| Acknowledgments |
|---|
| Footnotes |
|---|
2 Nonstandard abbreviations: TnI, troponin I; cTnI, cardiac TnI; mAb, monoclonal antibody; skTnI, skeletal TnI; and QC, quality control. ![]()
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
S. Eriksson, M. Junikka, P. Laitinen, K. Majamaa-Voltti, H. Alfthan, and K. Pettersson Negative Interference in Cardiac Troponin I Immunoassays from a Frequently Occurring Serum and Plasma Component Clin. Chem., July 1, 2003; 49(7): 1095 - 1104. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Uettwiller-Geiger, A. H.B. Wu, F. S. Apple, A. W. Jevans, P. Venge, M. D. Olson, C. Darte, D. L. Woodrum, S. Roberts, and S. Chan Multicenter Evaluation of an Automated Assay for Troponin I Clin. Chem., June 1, 2002; 48(6): 869 - 876. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Welsh, G. D. Kukes, and L. M. Sandweiss Differences of Creatine Kinase MB and Cardiac Troponin I Concentrations in Normal and Diseased Human Myocardium Ann. Clin. Lab. Sci., January 1, 2002; 32(1): 44 - 49. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Labugger, L. Organ, C. Collier, D. Atar, and J. E. Van Eyk Extensive Troponin I and T Modification Detected in Serum From Patients With Acute Myocardial Infarction Circulation, September 12, 2000; 102(11): 1221 - 1226. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Datta, K. Foster, and A. Dasgupta Comparison of Immunoreactivity of Five Human Cardiac Troponin I Assays toward Free and Complexed Forms of the Antigen: Implications for Assay Discordance Clin. Chem., December 1, 1999; 45(12): 2266 - 2269. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |