|
|
||||||||
Articles |
1
Veterans Administration Medical Center San Diego and
2
University of California San Diego, 3350 La Jolla Village Dr., San Diego, CA 92161.
a Address correspondence to this author at: VAMC-113, 3350 La Jolla Village Dr., San Diego, CA 92161. Fax 619-552-7479; e-mail rlfitzgerald{at}vapop.ucsd.edu
| Abstract |
|---|
|
|
|---|
Methods: We modified a commercially available column-switching instrument, the REMEDi HS from Bio-Rad Diagnostics, to make it compatible with atmospheric pressure ionization. The system we developed was based on electrospray ionization and used three LC columns to extract, purify, and separate drugs directly from urine specimens. Drugs and metabolites were tentatively identified on the basis of retention times and (M+H)+ ions. Tandem mass spectrometry (MS/MS) was used to confirm the qualitative identification of suspected drugs, using data-dependent acquisition. For quantitative analysis, the cocaine metabolite benzoylecgonine was analyzed using isotope dilution and selected reaction monitoring.
Results: Seventeen basic drugs from a variety of classes of compounds were identified directly from urine without the need for prior sample extraction, using LC and MS/MS. Quantitative analysis was demonstrated for benzoylecgonine. When benzoylecgonine-d3 was used as the internal standard, the method was linear from 30 to 10 000 µg/L (range tested). At these concentrations, the within-run accuracy was ± 10% of the target concentration, with CVs <10%. Analytical results by LC/MS/MS compared favorably with GC/MS values for 50 benzoylecgonine-containing specimens and for 25 negative specimens.
Conclusions: The ability to directly analyze urine for a wide variety of drug classes, combined with the sensitivity and specificity of LC/MS/MS makes this technique attractive for many clinical, forensic, and biotechnology applications.© 1999 American Association for Clinical Chemistry
| Introduction |
|---|
|
|
|---|
Automated solid phase extraction (SPE) is one potential solution to minimizing the time spent by technical staff preparing samples for GC/MS analysis. Robotic systems for the analysis of a variety of drugs are commercially available, and several publications have demonstrated their utility for the analysis of drugs of abuse (1)(2)(3)(4). These studies demonstrated that robotic systems can prepare samples in a consistent manner with accuracy and precision equal or superior to that of manual extractions. However, a major drawback with SPE is that many procedures are generally class specific. For example, an extraction/derivatization method designed for GC/MS analysis of amphetamines will not necessarily work for opiates. Additionally, once the robotic system has prepared the samples for analysis, many SPE methods include manual steps such as transferring the extract to an autosampler vial. These inherent sample preparation requirements make GC/MS analysis of many biologically important compounds a time-consuming process that cannot be bypassed with automation.
Liquid chromatography/mass spectrometry (LC/MS) is an alternative approach that can reduce the off-line sample preparation required with GC/MS because relatively nonvolatile compounds can be analyzed (5)(6)(7)(8)(9). LC/MS applications for forensic and clinical toxicology have been reviewed recently (10). With the proper choice of stationary and mobile phases, acceptable LC separations can be achieved for very polar analytes such as amines, alcohols, and carboxylic acids. This capability of analyzing a variety of compounds in a liquid phase makes LC attractive for the direct analysis of biological samples. Several studies have reported the analysis of compounds directly from biological matrices by LC (11)(12)(13)(14)(15)(16)(17)(18)(19)(20)(21), but only a few have reported direct analyses of biofluids by LC/MS (22)(23)(24). Generally, the approach for the direct LC analysis of biological specimens involves restricted access media (RAM) columns and/or column switching.
RAM columns are multifunctional stationary phases that are inert to large biomolecules such as proteins, allowing them to pass through the column while preferentially binding the smaller hydrophobic molecules of interest. RAM stationary phases have a porous hydrophilic exterior and a hydrophobic interior. The pore size of the exterior coating allows small molecules access to the reversed-phase interior coating, which binds the molecules until the mobile phase is changed to selectively elute them. This combination of size exclusion and reversed phase has been used for a variety of chromatographic separations, primarily involving serum (17)(18)(19)(20)(21).
A second approach to the direct analysis of compounds in a biological matrix uses multiple LC columns to isolate the compounds of interest on-line before analysis (5)(6)(7)(8)(9)(10)(11)(12)(25), a method that was first developed in the early 1970s (26)(27). The REMEDi HS from Bio-Rad Laboratories is a commercially available column-switching LC instrument designed to detect >700 different drugs directly from urine or serum. The REMEDi HS bases identification on retention times relative to an internal standard and on ultraviolet (UV) spectra that are searched against a library. Several reports have demonstrated the utility of the REMEDi HS in both clinical and forensic settings (28)(29)(30)(31)(32)(33)(34)(35). One limitation of the REMEDi HS is its reliance on the UV spectra for identification purposes. Compared with MS, the use of UV spectra for identification is a relatively insensitive and nonspecific technique that requires good chromatographic peak resolution when searching library spectra. Many compounds have similar UV spectra, and when they coelute, positive identification is difficult. Because UV spectra are not as definitive as mass spectra, we were interested in developing LC/MS for the direct analysis of biological specimens.
The purpose of the present work was to demonstrate that a column-switching LC/MS system could identify a broad range of analytes directly from a urine specimen on the basis of retention times, full-scan spectra, and product-ion spectra. Building on established chromatographic techniques and taking advantage of recent improvements in MS, we designed an online LC extraction, purification, and separation procedure compatible with atmospheric pressure ionization (API). This system was used to show, for the first time, that a wide variety of basic drugs could be qualitatively identified from a urine specimen in a single analysis. Using selected reaction monitoring (SRM) and isotope dilution, we validated the use of this system for the quantitative analysis of benzoylecgonine.
| Materials and Methods |
|---|
|
|
|---|
For qualitative identifications, the LCQ was operated in the data-dependent mode. The mass spectrometer was scanned from 50 to 500 atomic mass units (amu), with a full-scan target of 5 x 107 relative units and a tandem mass spectrometry (MS/MS) target of 2 x 107 relative units. The maximum ionization time was set to 400 ms, and three microscans were collected for each data point.
For SRM of benzoylecgonine, the same basic MS settings were used with the exception that the maximum ionization time was 200 ms. The precursor ions of benzoylecgonine and benzoylecgonine-d3 were 290.2 and 293.2 m/z with an isolation width of 2 amu. Product-ion scans were from 165.5 to 170.5 m/z and from 168.5 to 173.5 m/z for benzoylecgonine and benzoylecgonine-d3, respectively. Quantitative analysis was based on peak area ratios of the 168.1 ± 0.5 amu ion relative to the internal standard ion at 171.1 ± 0.5 amu. Each day of analysis, a two-point calibration curve at 150 and 1500 µg/L was run in duplicate. Benzoylecgonine controls were analyzed at 112 and 246 µg/L (n = 5) each day of analysis for 5 different days (total n = 25).
chemicals and reagents
The acetonitrile, isopropanol, and methanol were all HPLC grade or
better. The "application buffer" was 1 g/L potassium borate, pH 8.
The "exchange buffer" was 16 mmol/L ammonium acetate, pH 6.3,
containing 20 g/L isopropyl alcohol. The "mobile phase" was 16
mmol/L ammonium acetate, pH 7.05, containing 330 mL/L acetonitrile. The
"transfer buffer" contained 2 g/L pentanesulfonic acid, 200 mL/L
ethanol, and 20 g/L acetic acid. The "wash solution" was HPLC-grade
methanol.
lc columns
Column 1 was a PRP-1 [16 µm poly(styrene-divinylbenzene)
copolymer; 3.2 mm (i.d.) x 20 mm], column 2 was an Aminex, an
anion-exchange column [11 µm; 4.6 mm (i.d.) x 20 mm], and
column 3 was unmodified silica [45 µm; 6 mm (i.d.) x 160 mm].
All LC columns were obtained from Bio-Rad Diagnostics.
sample preparation
Sample preparation consisted of adding 200 µL of the internal
standard solution to 1 mL of urine. The internal standard was a 6 mol/L
ammonium acetate buffer, pH 8.0, containing 20 mL/L
isopropanol and
7-chloro-1,3,-dihydro-1-ethyl-5-phenyl-2H-1,4-benzodiazepine-2-one
(N-ethylnordiazepam) as internal standard 1 and
chlorpheniramine as internal standard 2. When added
to 1 mL of urine, the concentrations of internal standards 1 and 2 were
2000 and 3000 µg/L, respectively. After centrifugation (2 min at
9500g), 1 mL of the sample solution was injected by an
autosampler. For quantitative analysis of benzoylecgonine, the same
procedure was used with the addition of 10 µL of 15.8 mg/L
benzoylecgonine-d3.
Specimens containing >10 000 or 100 000 µg/L benzoylecgonine as determined by GC/MS were diluted 1:10 or 1:20, respectively, with drug-free urine before LC/MS/MS analysis. Comparisons of GC/MS and LC/MS/MS were made on specimens collected from routine drug screens that had been stored frozen (less than -20 °C) for ~1 year.
time event table
The time event table for column switching shown in Table 1
was developed by varying the settings and monitoring the
recovery of drugs of interest as a function of interfering substances
and stability of ion current. It begins with two wash steps immediately
after the urine sample is loaded onto column 1, including a
forward-flow wash and a reversed-flow wash using the application
buffer. Drugs and metabolites with different polarities are absorbed
onto the resin of column 1, whereas proteins, minerals, salts, and very
hydrophilic components are washed off to the waste by the application
buffer. Upon completion of the washing steps, a small amount of
exchange buffer is applied to column 1 from the reverse direction to
rinse off the application buffer and to prepare the cartridge for the
release of drugs and metabolites. The mobile phase then continues the
transfer from column 1 to column 2. Up to this point, all solvents are
delivered by pump A. Immediately after step 5, pump B is used to
complete the transfer of analytes to column 2. Basic drugs and
metabolites pass through column 2 with minimal retention, whereas most
neutral, acidic, and very polar components are retained by the
anion-exchange resin. Both columns 1 and 2 are switched out of the loop
immediately after the basic drugs are transferred to column 3 and are
washed with the wash solution. From this point on, pump B continues to
deliver mobile phase for the separation of basic drugs and metabolites
to column 3 at a rate of ~1.7 mL/min. The total process takes ~12
min per sample: the analytical separation time is 9 min, and the sample
loading and online purification take 3 min.
|
REMEDi LC/MS INTERFACE
The effluent from the UV cell of the REMEDi HS was connected to an
adjustable PEEK splitter (Upchurch) with ~6 feet of 0.01-inch
(i.d.) PEEK tubing. Effluent from the splitter was sent to waste or to
the LCQ through 0.005-inch (i.d.) PEEK tubing. The split ratio was
adjusted so that 95% of the mobile phase was sent to waste with ~85
µL/min going to the mass spectrometer. The LCQ analysis list was
started by a contact closure initiated from the REMEDi HS.
| Results |
|---|
|
|
|---|
|
One disadvantage of using ESI to identify unknowns is that it often
produces very little fragmentation. Compounds, such as morphine and
hydromorphone, that have the same nominal molecular weight will have
similar ESI spectra as shown in Fig. 2
, A and B. Although the chromatographic system used separates
these two compounds by >2 min, it is necessary to perform
collisionally induced dissociation (CID) MS/MS experiments to clearly
distinguish these compounds. The LCQ enables MS/MS data to be collected
in a data-dependent mode. Fig. 1C
shows the data-dependent scanning
chromatogram for the urine sample shown in Fig. 1A
. The jagged nature
of the peaks in Fig. 1C
are the result of the instrument switching from
a full-scan mode to an MS/MS mode. When an ion exceeds a preset
threshold in the full-scan mode, the instrument switches to collect a
CID spectrum of the most intense ion. Because the relative ion current
drops when MS/MS is performed, the total signal drops, causing a sharp
decline in the chromatographic peak. Following the MS/MS scan, the mass
spectrometer reverts back to full-scan mode, collects a spectrum, and
checks to see if the data-dependent threshold is exceeded. If the
threshold is still exceeded, it then collects another MS/MS spectrum of
the most intense peak. This process is repeated until the end of the
chromatographic run. Data-dependent scanning is important because CID
spectra of drugs generally contain more characteristic ions than
full-scan single MS experiments. Although CID spectra can be obtained
in several different modes (e.g., source CID and separate
injections), the ability to perform data-dependent scanning can
eliminate the need for a second injection. The CID spectra of morphine
and hydromorphone are shown in Fig. 2
, C and D. Although the full-scan
spectra of these drugs are similar, the CID spectra of morphine and
hydromorphone are clearly distinct.
|
It is possible to generate CID spectra using a single-stage MS instrument within the ion source by changing the electrospray voltages, a technique sometimes referred to as "source CID" or "up-front CID". However, a primary limitation of doing such experiments is that the lineage of the product ions is lost. This is an important point when analyzing biological extracts and for identifying coeluting compounds. For example, if source CID was done for benzoylecgonine, which coelutes with several unidentified endogenous compounds, the resulting spectrum would be a composite of both the benzoylecgonine CID spectrum and the endogenous peaks CID spectrum. By performing MS/MS in the ion trap, we can first isolate the benzoylecgonine protonated molecular ion (290 m/z) and then perform CID in the trap, with the knowledge that any resulting product ions originated from the 290 m/z ion.
We tested 17 drugs by adding them to urine specimens and analyzing the
specimens, using the data-dependent mode to collect CID spectra. The
retention times, observed protonated molecular ions, and relative
intensities of the product ions in the CID spectra are shown in Table 2
. Weinmann and Svoboda (36) recently published
several CID spectra for drugs of abuse obtained with a triple
quadrupole mass spectrometer. Although there is some general agreement
between our spectra and those obtained by Weinmann and Svoboda
(36), there are some notable differences. For example, in
our experiments, the CID spectrum of amphetamine shows almost
exclusively a single product ion at 119 m/z, with 91
m/z having a relative abundance of <1%. The amphetamine
CID spectrum obtained with the triple quadrupole is nearly the reverse
of what we observed with the ion trap: 91 m/z was the most
intense product ion, with 119 m/z having a relative
abundance of a few percent (36). There are some
obvious differences in the way in which the triple quadrupole CID data
were collected compared with our ion trap data (e.g., collision energy,
collision gas, and ion transmission) that could explain the observed
differences. However, the important point is that for drugs of abuse,
CID spectra collected on different types of instruments will not be
identical. At present, there are no widely available libraries of CID
spectra for drugs of abuse, but it appears that the best CID libraries
will be instrument specific.
|
The commercially available REMEDi HS UV uses chlorpheniramine as an internal standard, making it difficult to identify the use of this drug when it is present at low concentrations. However, previous experience with REMEDi showed that the chlorpheniramine metabolite norchlorpheniramine is often detected in urine samples from patients taking this drug (32). We did not analyze any samples from patients taking chlorpheniramine to determine whether its metabolites were detected with the mass spectrometer, but we anticipate eventually substituting a deuterium-labeled internal standard for chlorpheniramine to help ensure that exposure to this antihistamine will be detected.
Many drugs are excreted as glucuronide metabolites, including the opiates and benzodiazepines, and it is important to know how the system we developed works for conjugated metabolites. Although we did not fully evaluate the system for the analysis of glucuronide metabolites, two patient samples that were immunoassay positive for opiates did not show any glucuronide metabolites when injected into the REMEDi/MS. GC/MS analysis of these samples, after glucuronidase hydrolysis, SPE, and derivatization, showed that sample 1 contained morphine at 611 µg/L and sample 2 contained both morphine and codeine at 435 and 827 µg/L, respectively. ß-Glucuronidase pretreatment before REMEDi/MS analysis was successful in detecting morphine in sample 1 and both morphine and codeine in sample 2 without a noticeable increase in sample background. These preliminary REMEDi/MS experiments showed that off-line hydrolysis was compatible with the system we developed, but further studies are required to validate this finding.
quantitative analysis
Benzoylecgonine was chosen as the target compound for initial
quantitative studies because it is widely measured as an indicator of
cocaine abuse. With the MS/MS settings used, benzoylecgonine
predominately formed a product ion at 168 m/z; this ion
probably is produced by cleavage of the benzoyloxy group as shown in
Fig. 3
. This product ion is consistent with CID fragmentation of other
cocaine metabolites and a previous report (37). Because the
deuterium labels are on the methyl group attached to the nitrogen, the
internal standard forms an analogous product ion at 171 m/z.
|
To eliminate space charging and ion molecule reactions, it is necessary
to regulate the number of ions in the trap (38). The LCQ
uses a short prescan to determine the number of ions entering the trap
and then adjusts the ionization time to ensure that the trap is storing
ions in an optimal manner. Consequently, with the ion trap, the number
of data points collected per unit of time depends on the intensity of
the ion signal. This is in contrast to linear quadrupole mass
spectrometers, which collect a certain number of data points per unit
time interval. The three important settings that determine sampling
rate with the ion trap are target setting, maximum ionization time, and
number of microscans. We left the MS/MS target, which reflects the
desired number of ions in the trap, at the default setting of 2 x
107 relative units and varied the maximum
ionization time and the number of microscans to optimize precision of
SRM for benzoylecgonine. Table 3
shows that the combination of three microscans and a 200-ms
maximum ionization time provided the best precision for quantifying ion
ratios of benzoylecgonine at 150 µg/L. The settings were optimized at
this concentration because it reflects the current cutoff for a
positive confirmation of a presumptive positive immunoassay result. At
150 µg/L, the within-run CV was 3.4% (n = 6).
|
The REMEDi UV chromatograms of a blank urine specimen and a urine
specimen containing 150 µg/L of benzoylecgonine are shown in Fig. 4
, A and B. These chromatograms are indistinguishable, clearly
demonstrating the that REMEDi UV would not work for quantitative
analysis of benzoylecgonine at this concentration. Representative SRM
chromatograms of a blank urine specimen and a 150 µg/L urine
calibrator are shown in Fig. 4
, C and D. As can be seen from these
chromatograms, there are no peaks identifiable as benzoylecgonine in
the blank, whereas the 150 µg/L calibrator clearly shows a peak
corresponding to benzoylecgonine at 5.17 min. The benzoylecgonine
retention time difference of ~0.35 min between the UV chromatogram
and the SRM plots is produced by the interface between the UV
detector and the mass spectrometer.
|
The accuracy and imprecision of the SRM method for measuring
benzoylecgonine added to urine specimens is shown in Table 4
. The within-run CVs were 2.88.8%. The relative errors were
0.29.3% at concentrations between 75 and 10 000 µg/L. The lowest
concentration tested, 30 µg/L, had a relative error of 19.6%, which
meets the acceptability criteria used in most forensic urine drug
testing laboratories. We did not evaluate the lowest concentration
detectable because 30 to 10 000 µg/L met the needs of a drug testing
laboratory. At 30 µg/L, the signal-to-noise ratio was >100:1.
Although more experiments are required, better limits of detection
could probably be achieved by increasing the chromatographic efficiency
of the system. Narrow bore columns could be developed that would
introduce more concentrated peaks to the mass spectrometer, and the
current practice of sending 95% of the sample to waste before MS
analysis could be minimized.
|
The linearity of the method was evaluated by adding 30 to 10 000
µg/L benzoylecgonine to drug-free urine samples (n = 4 for each
concentration). As shown in Fig. 5
, the method was linear in this concentration range
[y = (0.97 ± 0.01)x - (6.1
± 47); r2 = 0.999]. This dynamic
range is better than most GC/MS procedures based on trimethylsilyl
derivatization, primarily because the inherent limitation of the
isotope cascade caused by the A+2 effect of 30Si
is bypassed in the LC/MS procedure. The isotope cascade effect of silyl
derivatives is particularly problematic when internal standards with
three or less deuteriums are used, as is commonly done in many
toxicology laboratories (e.g.,
benzoylecgonine-d3,
codeine-d3, and
morphine-d3).
|
Carryover was evaluated by injecting a blank urine specimen containing the benzoylecgonine-d3 internal standard immediately after a sample that contained 10 000 µg/L benzoylecgonine. There was no identifiable benzoylecgonine peak in the blank sample that followed the 10 000 µg/L calibrator.
To establish that our system provided satisfactory results in a
realistic setting, we evaluated 50 urine specimens collected from
random drug screens that had previously confirmed positive for
benzoylecgonine by GC/MS. The GC/MS procedure was based on isotope
dilution, SPE, and trimethylsilyl derivatization. Fig. 6
shows the correlation between GC/MS and LC/MS/MS over a wide
range of concentrations [y = (0.97 ±
0.02)x - (263 ± 839);
r2 = 0.987]. All LC/MS/MS results
agreed within ± 20% of the original concentrations assigned by
GC/MS. The accuracy of the GC/MS method over this concentration range
was ± 20%, and there were no significant differences between the
LC/MS/MS and GC/MS quantitative values, using a paired
t-test (P = 0.05). To ensure that no
false-positive results would occur, 25 specimens that were negative for
benzoylecgonine were analyzed by LC/MS/MS. As expected, none of the
negative specimens had peaks corresponding to benzoylecgonine.
|
The precision of the SRM method for quantifying benzoylecgonine was
evaluated over a period of several weeks by analyzing a
low-concentration control (112 µg/L benzoylecgonine) and a
high-concentration control (246 µg/L benzoylecgonine) with each batch
of specimens. A typical batch size was 38 samples and consisted of
donor specimens mixed with calibrators and controls. The concentration
of controls was selected to demonstrate that we could reliably
distinguish a positive sample from one that contained benzoylecgonine
at a concentration below the cutoff of 150 µg/L. As shown in Table 5
, the imprecision was 13% for the low control and 8.7% for the
high (above cutoff) control. All of the controls were properly
classified as to being above or below the cutoff, demonstrating that
the method has good discriminating power around the 150 µg/L cutoff.
The precision of the method using the LC/MS/MS ion trap for analyzing
controls was not as good as we usually get with our routine linear
quadrupole GC/MS method, which has CVs of ~4%. This is
consistent with our previous experiments that showed an ion trap
similar to the one used in the present studies did not measure ion
ratios as precisely as a linear quadrupole (39).
|
| Discussion |
|---|
|
|
|---|
A primary limitation of direct analysis of biofluids by MS/MS is that
the resolving power of a chromatographic system is eliminated. The
importance of a chromatographic separation should not be overlooked.
Even with the moderate resolution of the LC system we used, compounds
with the same nominal molecular weight, such as morphine and
hydromorphone, are separated by >2 min. This has important
implications for quantitative analysis when both these compound are
present in the same sample because there is some overlap of virtually
all of the ions in the CID spectra of these two isomers (Fig. 2
). Fig. 2D
shows three ions in the hydromorphone spectrum that could
potentially be used to distinguish it from morphine [m/z
(relative abundance), 185 (100%), 227 (34%), and 243 (23%)].
However, a careful look at the morphine spectrum (Fig. 2C
) shows
that these ions are also present [m/z 185 (10%), 227
(4%), and 243 (0.5%)]. When performing quantitative analysis of a
sample containing a mixture of these two compounds, it would be
difficult to get accurate results, especially if hydromorphone was
present at low concentrations in a sample with high amounts of
morphine. In addition to eliminating spectral overlap, the retention
time is also characteristic of the molecule of interest and provides
additional support for conclusive identifications.
The choice of mobile phases and columns will determine what type of compounds can be analyzed with LC/MS/MS. Various combinations of column switching, RAM, and immunoaffinity extraction have been used successfully for the direct analysis of compounds from biological specimens. Rule and Henion (42) demonstrated that an immunoaffinity column before a RAM column dramatically improved sample cleanup for the analysis of propranolol directly from urine specimens. These authors also showed that large volumes (20 mL) of dilute urine could be analyzed with immunoaffinity chromatography to detect low concentrations of lysergic acid diethylamide present in urine. Another particularly attractive feature of the procedure of Rule and Henion is the general applicability of the protein G column used for immunoaffinity purification. Because protein G has a high affinity for the Fc region of IgG, it can be potentially be used for any IgG antibody. The limitation of this procedure is that an antibody is required for each compound of interest and consequently would be less useful as a broad-spectrum screen.
Our choice of LC columns was based on previous publications demonstrating that various combinations of a PRP-1 column, a reversed-phase column, an anion-exchange column, and a bare silica column provided retention and separation of the majority of drugs of abuse directly from urine specimens. De Jong et al. (43) were the first to show that a PRP-1 column (column 1) in series with an Aminex (column 2) and a C18 column could be used for the direct analysis of barbiturates from urine specimens with UV detection. Slais et al. (44) used a similar configuration for the direct analysis of amphetamines from urine specimens. Binder et al. (11) improved the chromatographic separation by adding an unmodified silica column and showed that by using four columns, they could identify a wide variety of compounds when a scanning UV detector was used. In our preliminary experiments, we also used the four columns described by Binder et al., but because each column adds an additional expense, we were interested in developing a three-column system. We eliminated the C18 column and were able to resolve most of the drugs of interest with the exception of the benzodiazepines, which coelute at the beginning of the chromatographic run. Although we did not do conclusive studies, the major impact of eliminating the C18 column was on the separation of benzodiazepines, which are not retained on the bare silica column. However, because the clinically important benzodiazepines all have different molecular weights, they are easily differentiated by the mass spectrometer, obviating the need for chromatographic separation. Other than the effect on benzodiazepine separation, eliminating the C18 column appeared to have little effect on the overall performance of the system. Our final combination of a PRP-1 column, an anion-exchange column, and a bare silica column is the first description of this combination for broad spectrum drug isolation and separation.
Previous work demonstrated that sodium borate was optimal for binding drugs to the first column during the application step and that borate was useful as the exchange buffer when initiating the transfer of drugs from column 1 to column 2 (11). Sodium borate is not volatile and, consequently, not compatible with API. As the electrospray droplets evaporate, any sodium borate present will crystallize on the ESI needle and capillary inlet, preventing ions from reaching the mass analyzer. We retained the sodium borate in the application step but eliminated it from the exchange buffer. As expected, by using the previously described application buffer, we were able to bind drugs of interest to the first column. Developing a new exchange buffer by substituting ammonium acetate for sodium borate allowed us to initiate drug transfer from column 1 to column 2 and prevented sodium borate from reaching the electrospray interface. Our new exchange buffer retains the desired properties of the sodium borate buffer used previously yet is amenable to API. The final mobile phase used in previous studies also contained the nonvolatile buffer potassium phosphate as well as tetramethylammonium hydroxide and dimethyloctylamine, which clearly are not compatible with our ESI interface (11). We designed a new mobile phase that uses ammonium acetate and acetonitrile. Various concentrations of ammonium acetate (from 4 to 30 mmol/L) in 670 mL/L water330 mL/L acetonitrile were evaluated as the final mobile phase. The best compromise between peak shape and resolution was 16 mmol/L ammonium acetate; consequently, this was used for all of our studies.
The ability to bypass the need for manual extractions while maintaining the capabilities of a broad-spectrum analyzer makes this LC/MS method attractive in a variety of settings. Clinically, it is critical that any screening technique be sufficiently sensitive to detect a range of potential intoxicants, whereas specificity is required to help identify treatment strategies. The sensitivity of electrospray combined with the specificity of MS/MS enhances the diagnostic value of a urine drug screen and should aid in the treatment of suspected overdosed patients. In a postmortem forensic setting, this LC/MS/MS system could provide vital information as to cause of death, especially for compounds not detected by other routine techniques. Finally, from a more global perspective, this methodology could allow analytical chemists, through the proper selection of columns, mobile phases, and timed events, to design online sample preparations for virtually any compound amenable to API.
In conclusion, there have been several reports that used LC/MS to analyze single components or single classes of drugs directly from biological specimens. For the first time we have demonstrated that a wide range of basic drugs, from a variety of classes of compounds, can be identified directly from urine specimens using LC/MS/MS. We were successful in coupling recent advances in API and ion trap MS with evolving online chromatographic sample preparation to identify drugs based on retention times and full-scan and product-ion spectra. Using SRM and isotope dilution, we were able to validate the method for the quantitative analysis of the cocaine metabolite benzoylecgonine at concentrations encountered in routine forensic urine drug testing. The sensitivity can be improved by reducing the inner diameter of the LC columns to eliminate the need for effluent splitting just before MS analysis. Future experiments will be devoted to optimizing the LC system for more efficient separations, applications for other biofluids, and to software development for automated drug identification.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
F.-L. Sauvage, F. Saint-marcoux, B. Duretz, D. Deporte, G. Lachatre, and P. Marquet Screening of Drugs and Toxic Compounds with Liquid Chromatography-Linear Ion Trap Tandem Mass Spectrometry Clin. Chem., September 1, 2006; 52(9): 1735 - 1742. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Halket, D. Waterman, A. M. Przyborowska, R. K. P. Patel, P. D. Fraser, and P. M. Bramley Chemical derivatization and mass spectral libraries in metabolic profiling by GC/MS and LC/MS/MS J. Exp. Bot., January 1, 2005; 56(410): 219 - 243. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Combourieu, P. Besse, M. Sancelme, J.-P. Godin, A. Monteil, H. Veschambre, and A.-M. Delort Common Degradative Pathways of Morpholine, Thiomorpholine, and Piperidine by Mycobacterium aurum MO1: Evidence from 1H-Nuclear Magnetic Resonance and Ionspray Mass Spectrometry Performed Directly on the Incubation Medium Appl. Envir. Microbiol., August 1, 2000; 66(8): 3187 - 3193. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||