Methods: We simultaneously investigated the stability of 24 analytes (a) after prolonged contact of plasma and serum with blood cells and (b) after immediate separation of plasma and serum (centrifuged twice at 2000g for 5 min). We verified biochemical mechanisms of observed analyte change by concomitant measurement of pH, PCO2, and PO2. Hemolysis was qualitatively and semiquantitatively assessed. All specimens were maintained at room temperature (25 °C) and analyzed in duplicate 0.5, 4, 8, 16, 24, 32, 40, 48, and 56 h after collection. Statistically significant changes from the 0.5 h mean were determined using repeated-measures ANOVA. The significant change limit was applied to determine clinically significant changes in measured analytes.
Results: Fifteen of 24 analytes in plasma and serum maintained in contact with cells showed clinically relevant changes, with the degree of change more pronounced in most plasma specimens. All analytes in plasma and serum immediately separated from cells after collection were stable.
Conclusion: Storage of uncentrifuged specimens beyond 24 h caused significant changes in most analytes investigated because of (a) glucose depletion and Na+,K+-ATPase pump failure; (b) the movement of water into cells, causing hemoconcentration; and (c) leakage of intracellular constituents and metabolites. Immediate separation of plasma or serum from cells provides optimal analyte stability at room temperature. When prolonged contact of plasma or serum with cells is unavoidable, use of serum is recommended because of the higher instability of plasma analytes.
| Introduction |
|---|
|
|
|---|
Published literature pertaining to chemical analyte stability has addressed many issues related to serum specimens but largely neglected plasma. The stability of 72 analytes after prolonged contact of serum with cells has been described (1)(2)(3)(4)(5)(6)(7). The effects of prolonged storage on the stability of 31 analytes in plasma and serum separated from cells by a gel barrier are also known (6)(8)(9)(10). Lastly, the stability of 30 analytes in serum immediately separated from cells has been described (11)(12)(13)(14)(15), but no similar studies are available on plasma.
Our study addressed (a) the simultaneous measurement and determination of stability for 24 analytes after prolonged contact of plasma and serum with cells; (b) the simultaneous measurement and determination of stability for 24 analytes after immediate separation of plasma and serum from cells with subsequent delays before analysis; (c) verification of certain biochemical processes within uncentrifuged specimens by measurement of pH, PCO2, and PO2; (d) qualitative and semiquantitative measurement of free hemoglobin to indirectly assess erythrocyte membrane integrity and rule out hemolytic interference with our testing methodologies; and (e) determination of clinically relevant changes using the significant change limit (SCL) 1 approach.
| Materials and Methods |
|---|
|
|
|---|
-glutamyltransferase (GGT), glucose, lactate, lactate dehydrogenase (LD), Mg2+, Pi, K+, Na+, total protein, triglycerides, uric acid, and urea were measured with the Dade Dimension RxL (Dade International Inc.). We used the Gem Premier 2.2 Blood Gas Analyzer (Mallinckrodt Sensor Systems, Inc.) to measure pH, PO2, and PCO2. Free hemoglobin was assessed by the Hitachi 747 (Boehringer Mannheim Corp.).
specimen collection and handling
The investigational proposal was approved by the Institutional Review Board. All volunteers were informed of the rationale for the study, and consent was obtained. Venous blood was collected from each of 10, nonfasting volunteers after antecubital venipuncture into 15 sterile (unopened) 5-mL Vacutainer® Serum Separator Tubes with clot activator (SST; Becton Dickinson) and 15 sterile (unopened) 4.5-mL Vacutainer Plasma Separator Tubes containing 1520 IU/mL lithium heparin (PST; Becton Dickinson); samples were collected sequentially, and tubes were filled completely. Blood was collected from a different volunteer each day over 10 contiguous days. Ambulatory volunteers were seated in an upright posture
1 min before venipuncture and throughout specimen collection. The tubes were randomized to negate the effect of draw order with the exception that all SSTs were collected before the PSTs to prevent anticoagulant contamination (16). The tourniquet used during collection was removed after 1 min to prevent hemoconcentration and the production of metabolic byproducts (17). Total collection time per volunteer was 1015 min.
Cell-free plasma and serum were obtained by allowing seven PSTs and seven SSTs from each volunteer to stand for 0.5 h at room temperature (25 °C), after which tubes were centrifuged (2000g for 5 min). The plasma and serum from six centrifuged PSTs and six centrifuged SSTs from each volunteer were pooled into separate 15-mL sterile red-top no-additive Vacutainer tubes, recapped, and centrifuged. From each 15-mL red-top tube, the double-spun plasma (DSP) and serum (DSS) were subsequently aliquoted into nine respective 1.5-mL samples cups, capped, and wrapped with Parafilm (American National Can). One double-spun aliquot of plasma and serum from each volunteer was immediately analyzed on the Dimension RxL for the above-mentioned analytes to serve as the double-spun, 0.5 h baseline value. The remaining eight double-spun aliquots of plasma and serum from each volunteer were stored upright at room temperature (25 °C) in the dark, where they remained unopened until analysis. The plasma and serum of the remaining centrifuged PST and SST from each volunteer were immediately analyzed on the Gem Premier for pH, PCO2, and PO2, then on the Dimension RxL for use as the 0.5 h baseline value for each analyte with prolonged contact with cells.
Plasma and serum specimens in prolonged contact with cells were obtained by allowing the remaining eight uncentrifuged PSTs and eight uncentrifuged SSTs from each volunteer to sit at room temperature (25 °C) in the dark, where they remained unopened until analysis. Of the remaining uncentrifuged specimens and double-spun aliquots, one of each was analyzed at 4, 8, 16, 24, 32, 40, 48, and 56 h. At each designated time an uncentrifuged PST and SST was centrifuged, and the plasma and serum were immediately analyzed on the Gem Premier, then on the Dade Dimension RxL. DSP and DSS aliquots were analyzed only on the Dimension RxL. Specimens analyzed on the Gem Premier were run singly, whereas all specimens analyzed on the Dade Dimension RxL were run in duplicate (back to back). Two levels of commercial reference sera were run in conjunction to monitor instrument variation.
Qualitative assessment of free hemoglobin was graded by visual inspection of each sample at all time intervals. Semiquantitative measurement of free hemoglobin was performed on the Hitachi 747 in a preliminary study, where 2 of the 10 volunteers in the current investigation were randomly selected to undergo the same specimen collection and handling procedures as explained previously. Their samples for all time intervals were analyzed in duplicate on the Hitachi 747.
statistical analysis
To determine time-dependent changes of analytes in plasma and serum with and without prolonged contact with cells, the mean from all 10 volunteers for each respective analyte was obtained for each time interval. Statistically significant changes were determined for each analyte by repeated-measures ANOVA. Clinically significant changes were determined by the SCL (18) approach, defined as:
![]() |
| Results |
|---|
|
|
|---|
|
All 24 analytes in DSP and DSS were stable (by SCL) over the 56-h period; their 0.5-h mean and greatest change were as follows: ALT (U/L), DSP (47 -2), DSS (47 -1); albumin (g/L), DSP (41 +2), DSS (42 +2); ALK (U/L), DSP and DSS (85 ±2); AST (U/L), DSP (21 +1), DSS (22 -1); direct bilirubin (mg/L), DSP and DSS (1.0 ±0); total bilirubin (mg/L), DSP and DSS (6.0 ±0); calcium (mg/L), DSP (90 +2), DSS (91 +3); tCO2 (mmol/L), DSP (25.1 -1.2), DSS (26.7 -1.7); chloride (mmol/L), DSP and DSS (103 ±0); total cholesterol (mg/L), DSP (1870 +20), DSS (1940 -20); creatinine (mg/L), DSP and DSS (10 ±0); CK (U/L), DSP (112 +3), DSS (115 -3); GGT (U/L), DSP (31 +5), DSS (40 ±0); glucose (mmol/L), DSP (6.7 ±0), DSS (6.4 ±0); lactate (mmol/L), DSP and DSS (1.9 ±0); LD (U/L), DSP (137 ±1), DSS (143 -2); Mg2+ (mmol/L), DSP and DSS (0.74 ±0); Pi (mmol/L), DSP and DSS (1.1 ±0); K+ (mmol/L), DSP (4.0 ±0), DSS (4.1 ±0); Na+ (mmol/L), DSP (140 -1), DSS (141 -1); total protein (g/L), DSP (76 ±0), DSS (75 -1); triglycerides (mg/L), DSP (1940 ±10), DSS (1990 ±20); uric acid (mg/L), DSP (61 -1), DSS (60 +1); urea (mg/L), DSP and DSS (150 ±0).
Qualitative assessment of free hemoglobin was negative for all specimens. Semiquantitative measurement of free hemoglobin in the double-spun specimens for plasma and sera at 0.5 h was 40 mg/L with variation of ±15 mg/L, and ±10 mg/L for plasma and serum, respectively. Semiquantitative analysis of specimens with prolonged contact with cells yielded free hemoglobin values at 0.5 h of 80 mg/L for plasma and 60 mg/L for serum with variations from 50 to 110 mg/L and from 40 to 90 mg/L, respectively, over the 56-h study.
All commercial reference materials for the 24 analytes and blood gas values were within ±2 SD of their respective target means during the entire investigation.
| Discussion |
|---|
|
|
|---|
5000 mg/L. Because the highest quantitative measurement of free hemoglobin in our study was 110 mg/L, we conclude that free hemoglobin was not at sufficient concentrations to interfere with any methodology for the 24 analytes included in our study.
dsp and dss with prolonged incubation at room temperature
All 24 analytes in plasma and sera immediately separated from cells were stable over the 56-h period. Our results for serum analytes were consistent with those obtained by previous authors (11)(12)(13)(14)(15) who investigated these serum analytes under similar conditions, with a few notable exceptions: Felding et al. (14) and Berg et al. (15) noted instability of total bilirubin, with the latter stating photodegradation as a likely etiology; Baer and Krause (11) reported calcium instability but cited possible bacterial contamination as an etiology. Our observations of plasma analyte stability are new and demonstrate that plasma is equal to serum when stored under similar conditions over 56 h.
prolonged contact of plasma and sera with cells at room temperature
Variation of metabolically active analytes.
Glucose concentrations decreased rapidly to 24 h, then decreased more slowly to 56 h. Although the decrease was more pronounced in plasma, a stoichiometric yield of lactate was observed in both plasma and serum via glycolysis (see Table 1
). The net loss of glucose was 5.6 mmol/L for plasma and 3.6 mmol/L for serum, and the net gain of lactate was 12.6 mmol/L for plasma and 8.9 mmol/L for serum. The ratio of net glucose loss to net lactate production was 1:2.25 for plasma and 1:2.47 for serum, somewhat higher than the theoretical molar ratio of 1:2. The deviation from theoretical was most likely attributable to the high concentration of initial, steady-state glycolytic intermediates (19) within the erythrocytes at the time of specimen collection, which subsequently underwent glycolysis to yield additional lactate.
Chloride and tCO2 showed a steady decrease over 56 h, the degree of change being more pronounced for plasma tCO2 (see Table 1
). Our observations were consistent with previous studies of these serum analytes under similar conditions (2)(4)(5)(6)(7). The decreases in chloride and tCO2 were likely attributable to the chloride-bicarbonate shift with subsequent buffering of H+ ion (from lactic acid) by bicarbonate with production of CO2 gas (see Eq. 1
)
![]() | (1) |
Pi showed a slight decrease from 8 to 16 h with only plasma Pi exceeding the SCL at 16 h; however, this 0.2 mmol/L decrease was not considered clinically significant. After 24 h, plasma and serum Pi concentrations increased continuously, with the degree of change more prominent in plasma (see Table 1
). Our observations for serum were consistent with other studies using serum in prolonged contact with cells (3)(4)(6); these studies attributed the increase to hydrolysis of intracellular phosphate esters with subsequent diffusion from the erythrocyte.
K+ was stable to 24 h, after which a rapid increase was noted; the degree of change was slightly more pronounced in plasma. Our observations for serum were consistent with previous studies using serum in prolonged contact with cells at room temperature (3)(4)(6)(20).
The increase in K+ after 24 h, the approximate nadir of extracellular glucose, was most likely attributable to Na+,K+-ATPase pump failure, with diffusion of K+ from the erythrocyte driven by the intracellular to extracellular concentration gradient of 25:1 (19).
Nonmetabolic analytes and hemoconcentration.
Analytes demonstrating clinically insignificant increases after 24 h (percentage increase) were total bilirubin (plasma and serum, +17%), Na+ (plasma and serum, +2%), and urea (plasma and serum, +6%). Analytes demonstrating clinically significant changes after 24 h (percentage increase) were albumin (plasma and serum, +7%), calcium (plasma, +6%), Mg2+ (plasma, +16%), and total protein (plasma, +5%). Our observation for serum Na+ was consistent with similar studies investigating serum with prolonged contact with cells at room temperature (3)(4)(7). Zhang et al. (7) also noted similar observations for albumin, total protein, calcium, Mg2+, and Na+ with serum specimens. These observations most likely represent hemoconcentration attributable to the movement of water into cells after 24 h. Uric acid showed a decreasing trend over the 56-h period, with the SCL surpassed at 40 h for plasma and 32 h for serum, which may be related to decreased uric acid solubility in the continuously increasingly acidic environment.
Lipids and enzymes.
Cholesterol trended upward, with the degree of change more pronounced in plasma than serum. A similar trend was noted previously in serum specimens, and the mechanism proposed was that cholesterol was being continuously removed from blood cells via lecithin:cholesterol acyltransferase (6). Triglycerides were stable over the 56-h period.
LD activity increased continuously over the investigational period, with the SCL surpassed at 16 h and the degree of change more prominent in plasma. Our serum observations were consistent those of authors of similar studies (2)(3)(4)(6)(7), who postulated that the increase was attributable to changes in cell membrane integrity. The activities of plasma and serum AST, ALK, and CK were stable over the 56-h period, and our serum observations were consistent with previous studies (2)(3)(4)(5)(6)(7).
Variation attributable to methodologic interference.
Plasma ALT was stable to 40 h, then lost 20% of activity at 48 and 56 h. This decrease was not observed in serum or the DSP and DSS aliquots and is therefore likely attributable to increased lactate concentrations in plasma specimens after 40 h, which may have interfered with the methodology [ALT-catalyzed conversion of L-alanine +
-ketoglutarate to L-glutamate + pyruvate, with subsequent LD-catalyzed conversion of pyruvate + NADH(H+) to lactate + NAD+].
The GGT activity in plasma specimens with and without prolonged contact with cells was 27% lower than in serum at 0.5 h; however, GGT activity steadily increased over the 56-h period, with only plasma in prolonged contact with cells exceeding the SCL at 48 h (see Table 1
). The discrepancy in plasma and serum activity at 0.5 h was most likely attributable to lithium heparin interference with the methodology (manufacturers package inserts state interference at or above 280 000 IU/L lithium heparin).
Creatinine increased by 110% in plasma and 60% in serum after 24 h, which was probably attributable to interference of pseudo-creatinines (6) with the kinetic Jaffe reaction. The dramatic increase could not be explained simply by hemoconcentration, because most other nonmetabolic analytes (i.e., albumin, Na+, total protein, and urea) increased by only 26%.
| Acknowledgments |
|---|
| Footnotes |
|---|
-glutamyltransferase; LD, lactate dehydrogenase; SST, serum separator tube; PST, plasma separator tube; USD, usual SD; DSP, double-spun plasma; and DSS, double-spun serum. | References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
M. Tanner, N. Kent, B. Smith, S. Fletcher, and M. Lewer Stability of common biochemical analytes in serum gel tubes subjected to various storage temperatures and times pre-centrifugation Ann Clin Biochem, July 1, 2008; 45(4): 375 - 379. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E Cust, R. Kaaks, C. Friedenreich, F. Bonnet, M. Laville, A. Tjonneland, A. Olsen, K. Overvad, M. U. Jakobsen, V. Chajes, et al. Metabolic syndrome, plasma lipid, lipoprotein and glucose levels, and endometrial cancer risk in the European Prospective Investigation into Cancer and Nutrition (EPIC) Endocr. Relat. Cancer, September 1, 2007; 14(3): 755 - 767. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A.R. Bowen, Y. Chan, J. Cohen, N. N. Rehak, G. L. Hortin, G. Csako, and A. T. Remaley Effect of Blood Collection Tubes on Total Triiodothyronine and Other Laboratory Assays Clin. Chem., February 1, 2005; 51(2): 424 - 433. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. R. Miles, R. F. Roberts, A. R. Putnam, and W. L. Roberts Comparison of Serum and Heparinized Plasma Samples for Measurement of Chemistry Analytes Clin. Chem., September 1, 2004; 50(9): 1704 - 1706. [Full Text] [PDF] |
||||
![]() |
M. Rosa-Fraile, A. Sampedro, J. Rodriguez-Granger, E. Camacho, and E. Manrique Suitability of Frozen Serum Stored in Gel Separator Primary Sampling Tubes for Serological Testing Clin. Vaccine Immunol., January 1, 2004; 11(1): 219 - 221. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |