Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 48: 473-483, 2002;
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (17)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Qin, Q.-P.
Right arrow Articles by Pettersson, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Qin, Q.-P.
Right arrow Articles by Pettersson, K.
(Clinical Chemistry. 2002;48:473-483.)
© 2002 American Association for Clinical Chemistry, Inc.

Point-of-Care Time-resolved Immunofluorometric Assay for Human Pregnancy-associated Plasma Protein A: Use in First-Trimester Screening for Down Syndrome

Qiu-Ping Qin1a, Michael Christiansen2 and Kim Pettersson1

1 Department of Biotechnology, University of Turku, Tykistökatu 6, 20520 Turku, Finland.

2 Department of Clinical Biochemistry, Statens Seruminstitut, Artillerivej 5, DK-2300 S Copenhagen, Denmark.

aAuthor for correspondence. Fax 358-2-333-8050; e-mail qiqin{at}utu.fi.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Background: Screening for Down syndrome in the first trimester by a combination of fetal nuchal translucency thickness and maternal serum pregnancy-associated plasma protein A (PAPP-A) and free ß-human chorionic gonadotropin has been shown to be effective and efficient. We aimed to develop a fast point-of-care assay that could be placed in one-stop clinics for the measurement of PAPP-A.

Methods: We developed a two-site, one-step assay that uses two monoclonal antibodies (mAbs) to PAPP-A, based on a dry-reagent, all-in-one immunoassay concept with a stable fluorescent lanthanide chelate and time-resolved fluorometry. One antibody (mAb 10E1) was biotinylated, and the other (mAb 234-5) was europium-labeled, both via the {epsilon}-amino groups of surface lysine residues. The assay was performed on an AIO immunoanalyzer at 36 °C in single, streptavidin-coated microtitration wells that contained the dry reagents. PAPP-A, either in free or complexed form, was detected by the antibodies used.

Results: The assay procedure required 20 min and used 10 µL of sample. The calibration curve was linear from 5 to 10 000 mIU/L. The detection limit was 0.5 mIU/L. Intra- and interassay imprecision (CV) was <=4.3% and 8.3%, respectively, for whole blood, plasma, or serum samples. Recovery was 93–96% for serum, 95–108% for heparin-derived whole blood, and 98–103% for heparin-derived plasma. Parallelism was observed in all three matrices. Results correlated [slope = 0.85 (confidence interval, 0.82–0.87); intercept = -33 (confidence interval, -58 to -9); Sy|x = 85 mIU/L; r = 0.991; n = 100] with those obtained by a Delfia assay. Heparin did not affect the assay, but EDTA markedly reduced PAPP-A values. PAPP-A was stable at 4 °C for at least 18 days in serum and for 8 days in heparin-derived whole blood or plasma.

Conclusions: The present assay appears suited for use in one-stop clinics for screening for Down syndrome in the first trimester, with results available within 1 h.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Pregnancy-associated plasma protein A (PAPP-A), 1 an {alpha}2-mobile glycoprotein, was first identified in the early 1970s as a high-molecular weight constituent found in human late pregnancy serum (1). PAPP-A secreted from the trophoblastic tissue of the placenta is considered the main source of circulating PAPP-A in pregnancy. The circulating form of PAPP-A is a heterotetrameric complex composed of two 200- to 250-kDa PAPP-A subunits disulfide-bridged to two 50- to 90-kDa molecules of the proform of eosinophil major basic protein (proMBP) (2). However, pregnancy serum or plasma is also reported to contain traces (<1%) of uncomplexed PAPP-A (3).

Recently, PAPP-A has been found to specifically cleave insulin-like growth factor-binding protein-4 (IGFBP-4) in an IGF-dependent manner (4)(5) and IGFBP-5 in an IGF-independent manner (6). IGFs (IGF-I and IGF-II) play an important role in promoting cell differentiation and proliferation in various biological systems. The biological activity of IGFs is strictly regulated by six homologous high-affinity IGFBPs. IGF bound to IGFBP is unable to interact with its receptor, but bioactive IGF can be released by cleavage of the binding protein (4)(5)(6). In addition to the syncytiotrophoblasts and the trophoblast-derived septal X cells from placenta (7), PAPP-A can also be secreted by osteoblasts (4), marrow stromal cells (5), fibroblasts (4), granulosa cells (8), and vascular smooth muscle cells (9). PAPP-A secreted from all these sources possesses IGFBP proteinase activity.

In uncomplicated pregnancies, PAPP-A concentrations in maternal serum increase with gestational age until term. As a marker, PAPP-A was once suggested to be a risk marker for threatened abortion (10). However, monitoring circulating concentrations of PAPP-A was then shown to be of no clinical value in predicting the outcome of pregnancies threatened by abortion (11). Later, decreased PAPP-A in the first trimester was found to be associated with fetal Down syndrome (DS), and it is now one of the best biochemical markers in early pregnancy, commonly used to screen for DS during the first trimester (12)(13)(14)(15)(16).

DS is the most prevalent autosomal chromosome abnormality in live births and the most common chromosomal abnormality causing mental retardation (17). No method is currently available for its treatment, but avoidance of DS birth through prenatal screening followed by invasive prenatal diagnosis and selective therapeutic termination of the affected pregnancies has become an option for pregnant women. Second-trimester serum screening (between 15 and 20 weeks of pregnancy) for DS by maternal age and measurements of {alpha}-fetoprotein, human chorionic gonadotropin, and unconjugated estriol, with or without inhibin A can identify 65–75% of affected pregnancies with a false-positive rate of ~5% (18). First-trimester screening (between 10 and 13 weeks of pregnancy) for DS, based on maternal age, PAPP-A and free ß-human chorionic gonadotropin concentrations, and measurement of an ultrasound marker, i.e., nuchal translucency, can detect 85–90% of affected pregnancies with a false-positive rate of ~ 5% (19)(20)(21)(22). In addition to the higher detection rate, screening in the first trimester would allow more time to counsel the patients, and more importantly, pregnancy termination would become much easier to perform, leading to less complications occurring thereafter (23). In consequence, there is now increasing evidence to support performing DS screening in the first trimester.

The future trend for first-trimester screening for DS could combine the measurements of PAPP-A, free ß-human chorionic gonadotropin, and nuchal translucency in a near-patient environment. The measurements must be performed quickly with adequate analytical quality to provide rapid results to allow doctors to estimate the risks for a pregnant woman at the time of her clinic visit. Thus, considerable savings both in time and economy and better quality of care can be achieved. A recent prospective study has shown promise in this direction (22).

We have previously reported the development of monoclonal-based single-label or dual-label time-resolved immunofluorometric assays for PAPP-A in serum (24)(25). The analytical accuracy is well validated by the excellent clinical results obtained from the use of the assays (19)(26)(27)(28). However, because of the nature of the antibodies used and the reaction conditions, these assays are time-consuming (3–6 h), and some cannot be used for plasma or whole blood samples.

We describe a rapid point-of-care, time-resolved immunofluorometric assay for PAPP-A developed for one-stop clinics to allow screening for DS in the first trimester. The assay is based on a system that embodies a dry-reagent all-in-one immunoassay concept and utilizes a stable fluorescent lanthanide chelate as well as the time-resolved fluorometry (29). The entire procedure is fully automated and is equally applicable to serum, heparin-derived plasma, and heparin-containing whole blood samples.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
instrumentation
The 1420 multilabel counter (VictorTM) for time-resolved surface measurement of solid-phase fluorescence, the 1230 Arcus fluorometer for time-resolved measurement of DELFIA enhanced fluorescence in tube format, the 1234 DELFIA fluorometer for time-resolved measurement of enhanced fluorescence in plate format, and DELFIA platewash were obtained from Perkin-Elmer Life Sciences, Wallac Oy. The AIO immunoanalyzer to perform automated all-in-one assays was from Innotrac Diagnostics Oy. The iEMS Incubator/Shaker from Labsystems Oy was used for plate incubation at 36 °C. The fast-performance liquid chromatography system with a SuperdexTM 200 HR 10/30 (a prepacked column) was from Pharmacia Biotech.

reagents
ITC-TEKES Eu3+ fluorescent chelate of 4-[2-(4-isothiocyanatophenyl)ethynyl]-2,6,-bis{[N,N-bis(carboxymethyl)-amino]methyl}pyridine and biotin isothiocyanate (BITC) were obtained from Perkin-Elmer Life Science, Wallac Oy. Assay buffer and wash solution were prepared as described previously (30). Low-fluorescence 12-well Maxisorp microtitration strips (ultraviolet-quenched) were purchased from NUNC. Streptavidin-coated single wells and strips were obtained from Innotrac Diagnostics Oy. Bovine serum albumin (BSA) was purchased from Intergen. NAP-5TM and NAP-10TM columns were from Pharmacia Biotech. Heparin was purchased from CalBiochem. All other chemicals used were of analytical grade.

monoclonal antibodies
Fourteen monoclonal antibodies (mAbs) against PAPP-A were initially included in this study. Antibodies Hyb234-2, Hyb234-3, Hyb234-4, and Hyb234-5 were obtained from Statens Serum Institut; mAbs 1B2, 1F12, 3B6, 1C3, 4D2, 5H9, 8H1, 10E, 10E2, and 16D4 were obtained from HyTest Oy.

calibrators
Calibrators were made from a 26-week-pregnancy serum diluted in 50 mmol/L Tris-HCl buffer (pH 7.8) containing 150 mmol/L NaCl and 60 g/L BSA (6% BSA-TSA buffer). The calibrators were calibrated against WHO IRP 78/610 for pregnancy-associated proteins (WHO International Laboratory for Biological Standards, Statens Serum Institut).

antibody biotinylation
Biotinylation of the antibodies, using the BITC reagent, has been described previously (25). Briefly, to 1 mL of antibody solution (1 g/L) in 50 mmol/L sodium carbonate buffer (pH 9.6) we added a 50-fold molar excess of BITC dissolved in dimethylformamide. The mixture was left at room temperature for 3 h. After removal of excess free reagents by gel filtration with a NAP-5 and a NAP-10 column, the biotinylated antibody was stored in 50 mmol/L Tris-HCl (pH 7.8) containing 150 mmol/L NaCl, 0.5 g/L NaN3, and 1 g/L BSA.

labeling of antibodies with a fluorescent Eu3+ chelate
The antibodies were labeled overnight at room temperature in 50 mmol/L sodium carbonate buffer (pH 9.6) containing a 100-fold molar excess of ITC-TEKES Eu3+ chelate. The labeled antibody was separated from free chelate and aggregated proteins by gel filtration on a Superdex 200 HR 10/30 column equilibrated and eluted with 50 mmol/L Tris-HCl (pH 7.8) containing 0.15 mol/L NaCl and 0.5 g/L NaN3. BSA and trehalose were added to the labeled antibody solution to final concentrations of 5 and 125 g/L, respectively. Under these conditions, the labeled mAbs contained 7–10 Eu3+ molecules per IgG molecule.

antibody immobilization
Biotinylated antibodies were immobilized on the surface of streptavidin-coated microtitration wells at room temperature. To each well we added 50 µL of biotinylated antibody solution. Wells were incubated for 2 h with continuous slow shaking or overnight without shaking. After a washing step, these antibody-immobilized wells were ready for use.

antibody selection
Fourteen antibodies against PAPP-A were tested in pairs with each used as a capture or a detection antibody. A one-step sandwich assay format was used together with a 5000 mIU/L PAPP-A calibrator and a blank solution. The procedure were as follows: 10 µL of PAPP-A calibrator or blank solution and 400 ng of Eu3+-labeled antibody in 20 µL of assay buffer were added, in triplicate, to wells containing 400 ng of immobilized biotinylated antibody. Subsequent incubation was performed at 36 °C for 15 min with shaking. After that, the wells were washed six times with wash solution and dried with a stream of hot dry air for 5 min; the fluorescence was then measured with a 1420 multilabel counter (Victor).

preparation of aio dry-reagent wells
The procedure used for preparation of dry-reagent wells for PAPP-A was basically the same as that described previously (31). Briefly, to the microtitration wells containing 400 ng of immobilized biotinylated antibody, we added 40 µL of an insulating solution; the wells were then dried overnight. We then dispensed 1 µL of Eu3+-labeled antibody solution on top of the dried insulating layer and immediately dried the wells. Dried wells were either packed on top of each other in plastic reagent "pens" or kept within microwell frames, and were thereafter stored in sealed airtight bags with desiccant until use. Dried calibrator wells were prepared similarly to the sample wells. However, in contrast to the sample wells, dried calibrator wells contained a known concentration of PAPP-A in the insulating layer.

assay procedure
To each AIO dry-reagent well, we added 10 µL of sample (serum, plasma, or whole blood) and 20 µL of assay solution. Subsequent incubation was performed at 36 °C for 20 min with continuous shaking. After six washes, the wells were dried for 35 s under a stream of hot dry air and immediately brought to a time-resolved fluorometer for fluorescence measurement. The concentrations of unknown samples were obtained by calibrating their fluorescence signals against a calibration curve constructed using the calibrator wells. The whole assay procedure was automatically performed by the AIO immunoanalyzer.

sample materials
One hundred first-trimester serum samples were collected from a prenatal screening program for severe malformations, infections, and DS at Statens Seruminstitut. The samples were stored at -20 °C before the measurement. Twelve whole blood samples (5 mL/sample) were obtained from healthy volunteers with heparin as the anticoagulant in the blood collection tubes. Each of these 12 whole blood samples was divided into two aliquots. To each aliquot we added a different concentration of PAPP-A, generating a total of 24 whole blood samples containing different amounts of PAPP-A. Subsequently, 24 corresponding plasma samples were prepared by centrifugation at 3000g for 5 min. These samples were stored at 4 °C before the measurement. The hematocrits (Hcts) of these 12 blood samples were determined in Innotrac Diagnostics Oy by centrifugation at 14 926g for 3 min in a desk centrifuge (Biofuge Hemeo) from Heraeus Instruments. Hct values for the PAPP-A-enriched samples were derived from those of respective original whole blood samples by volume correction.

effects of anticoagulants
The effect of EDTA was investigated by adding a known amount of PAPP-A to tubes containing 500 µL of the 6% BSA-TSA buffer and different concentrations of EDTA (1–100 mmol/L). These samples were mixed, and their PAPP-A concentrations were measured with two assays, e.g., the present point-of-care assay and a liquid reagent-based, one-step assay configured with a biotinylated mAb (10E1; 400 ng/well), and a Eu3+-labeled mAb (234-2; 200 ng/well). For the latter assay, the incubation was conducted at room temperature for 3 h. After the wells were washed six times, 200 µL of enhancement solution was added to each well. After 5 min of slow shaking, the fluorescence in each well was measured.

The effect of heparin was similarly studied. Briefly, a known amount of PAPP-A was added to tubes containing 500 µL of the 6% BSA-TSA buffer and different concentrations of heparin (1–1000 kIU/L). These samples were well mixed, and their PAPP-A concentrations were measured with two assays, e.g., the present point-of-care assay and a liquid reagent-based, one-step assay configured with a biotinylated mAb (234-4; 400 ng/well) and a Eu3+-labeled mAb (234-2; 100 ng/well). For the latter assay, the incubation was conducted at room temperature for 3 h. After the wells were washed six times, 200 µL of enhancement solution was added to each well. After 5 min of slow shaking, the fluorescence in each well was measured.

stability
We studied the stability of PAPP-A in different sample matrices, including 6% BSA-TSA buffer, serum, EDTA-based plasma, EDTA-based whole blood, heparin-based plasma, and heparin-based whole blood. These samples were stored at 4 °C over certain periods of time during which the PAPP-A concentrations were measured with the point-of-care assay.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
best antibody pair
Antibody selection was first based on the results of antibody combination studies. Those antibody pairs that gave lower detection limits and linear dose-response signals were preliminarily chosen. Further selection was based on data on assay kinetics and specificity. Only those antibody combinations that enabled rapid measurement of PAPP-A were selected for dry-reagent assay optimization and final characterization. An antibody pair in which 10E1 served as capture and 234-5 as detection antibody fulfilled these requirements and was therefore selected as the best antibody pair.

kinetics
In the present point-of-care assay, biotinylated 10E1 immobilized on the well surface was kept in close contact with PAPP-A and Eu3+-labeled mAb 234-5 by the use of a small reaction volume, higher incubation temperature, and continuous mechanical shaking. Binding of both the low- and high-concentration buffer-based calibrators reached almost 100% in 20 min. Serum samples had virtually the same kinetics as the buffer-based calibrators, whereas plasma and whole blood samples obtained with heparin as the anticoagulant exhibited slightly slower kinetics (91–92% binding). A similar kinetic reaction was observed at concentrations of 5–10 000 mIU/L.

calibration curve and precision profile
The calibration curve for buffer-based calibrators was linear at concentrations of 5–10 000 mIU/L; the CVs at these concentrations were <10%, as shown in Fig. 1 . No hook effect was observed up to at least 50 000 mIU/L, a fivefold higher concentration than the highest calibrator used, because the signal achieved was still increasing.



View larger version (19K):
[in this window]
[in a new window]
 
Figure 1. Calibration curve (•) and corresponding within-assay imprecision profile ({square}) of the developed point-of-care assay for PAPP-A.

Each data point is the mean of eight determinations. The background fluorescence has been subtracted from all measurements.

methodologic evaluation
Detection limit.
The detection limit, estimated as the concentration of PAPP-A giving a fluorescence signal equivalent to that of the mean of 20 replicates of the zero standard + 3 SD, was 0.5 mIU/L.

Intra- and interassay imprecision.
Intra- and interassay imprecision (as CVs) was determined with serum samples, heparin-derived plasma samples, and heparin-derived whole blood samples covering low and high PAPP-A concentrations. As shown in Table 1 , both intra- and interassay CVs were <=8.3% irrespective of the sample matrix.


View this table:
[in this window]
[in a new window]
 
Table 1. Intra- and interassay imprecision with regard to three sample matrices.

Recovery and parallelism.
Analytical recovery of 1000 mIU/L PAPP-A was 93–95% in two serum samples, 98–103% in two heparin plasma samples, and 95–108% in two heparin whole blood samples with low (400–800 mIU/L) and high (3000–5000 mIU/L) PAPP-A.

Parallelism was examined by diluting three samples in different matrices with the 6% BSA-TSA buffer, and these diluted samples were measured again as the new samples. Agreement between the expected values and the measured values was good, as shown in Fig. 2 .



View larger version (19K):
[in this window]
[in a new window]
 
Figure 2. Linearity of results after serial dilution of three samples with different matrices.

Each data point is the mean of two determinations.

Carryover.
The carryover of the AIO instrument was examined by measuring a serum sample with a high PAPP-A concentration (50 000 mIU/L) and the zero standard in sequence. The carryover was 0.0078% in this case, which is below the allowable limit of 0.01% set for third-generation thyroid-stimulating hormone assays (32).

Effects of anticoagulants.
Heparin at concentrations of 1–1000 kIU/L had no effect on the point-of-care assay, but it did have an effect on the assay that used biotinylated mAb 234-4 and Eu3+-labeled mAb 234-2, producing substantially lower signals and, hence, lower measured concentrations of PAPP-A. Noticeably, this effect was dose dependent: the higher the heparin concentration, the lower the fluorescence signal obtained, as shown in Fig. 3 .



View larger version (17K):
[in this window]
[in a new window]
 
Figure 3. Effect of increasing concentrations of heparin on PAPP-A concentrations measured by two immunoassays.

After the addition of heparin, the samples were left at room temperature for 3 h and then assayed. After background subtraction, each measured binding signal was compared with the binding signal obtained for a sample that did not contain heparin. In this way, the relative binding signal was derived. Each data point is the mean of two determinations.

In contrast to heparin, EDTA had an effect on the point-of-care assay. Assay signals were markedly decreased in the presence of EDTA, as seen in Fig. 4 . This reduction was concentration dependent, with more pronounced reductions being observed in the presence of higher concentrations of EDTA. Furthermore, the signal decreased progressively with increasing time, revealing that the decrease was time dependent. However, EDTA had nearly no effect on the assay configured with biotinylated mAb 10E1 and Eu3+-labeled mAb 234-2, as shown in Fig. 4 , indicating that EDTA affected only the binding of mAb 234-5 to PAPP-A.



View larger version (25K):
[in this window]
[in a new window]
 
Figure 4. Effect of increasing EDTA concentration on PAPP-A, as measured by two immunoassays.

After the addition of EDTA, the samples were left at 4 °C, and the relevant PAPP-A measurements were performed at different time intervals. Each data point is the mean of two determinations.

stability
The concentration of PAPP-A in the 6% BSA-TSA buffer did not change during 3 months of storage at 4 °C (data not shown). In addition, the PAPP-A concentration in serum did not change when the sample was stored at 4 °C for 18 days. The concentration of PAPP-A in heparin-based plasma was stable over a period of 8 days at 4 °C, as was also the case for heparin-based whole blood. However, the PAPP-A concentration in EDTA-based plasma decreased with time at 4 °C; this observation also applied to EDTA-based whole blood, as shown in Fig. 5 .



View larger version (17K):
[in this window]
[in a new window]
 
Figure 5. Relationship of PAPP-A stability with sample matrix.

PAPP-A concentrations were determined by the point-of-care assay and are plotted as concentrations relative to the first time measurements. Each data point is the mean of two determinations.

correlation between whole blood samples and plasma samples
The present assay was applied to 24 whole blood samples as well as the corresponding plasma samples. The results from the whole blood samples correlated very well with those from the plasma samples, as shown in Fig. 6A . However, the slope of the curve was ~0.62, indicating that the mean difference between the plasma samples and the whole blood samples was 0.38-fold of the values obtained from the plasma samples. This difference coincided with the mean Hct of the whole blood samples. As shown in Fig. 6B , the slope of the curve reached 0.993 after the whole blood measurements were corrected by the corresponding Hct values.



View larger version (13K):
[in this window]
[in a new window]
 
Figure 6. Correlation between results from heparin-derived whole blood and corresponding plasma samples (n = 24).

In A, the whole blood measurements have not been corrected with the respective Hct values, whereas they have been corrected in B. Each data point is the mean of two determinations. Equations for the lines: (A), y = 0.62x - 40; r = 0.991; n = 24; (B), y = 0.99x + 22; r = 0.996; n = 24.

comparison of the assay with a delfia assay
The test results obtained by the point-of-care assay were compared with those obtained by a Delfia assay configured with biotinylated mAb 234-4 and Eu3+-labeled mAb 234-2 as described previously (24) for serum samples (n = 100). The correlation was good (r = 0.991) between the results obtained with two methods. The regression analysis yielded the equation: y = 0.85x - 33, as shown in Fig. 7A . As can be seen in the difference plot (Fig. 7B ), however, the mean percentage of difference between the two methods was ~25% (24.99% ± 17.65%) and was statistically significant (P <0.001, two-tailed t-test). The bias was not concentration dependent over the entire concentration range.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 7. Comparison of PAPP-A results obtained by the point-of-care and Delfia assays.

(A), correlation between the PAPP-A results obtained with the point-of-care assay and the Delfia assay for 100 serum samples from the first trimester. The slope was 0.85 ± 0.01, the y-intercept was -33 ± 13 mIU/L, and Sy|x was 85 mIU/L. (B), difference plot comparing PAPP-A values in 100 sera measured with the point-of-care assay and the Delfia assay. The x axis represents (PAPP-Apoint-of-care assay + PAPP-ADelfia assay)/2, and the y axis represents [(PAPP-Apoint-of-care assay - PAPP-ADelfia assay) x 100]/[(PAPP-Apoint-of-care assay + PAPP-ADelfia assay)/2]. On the difference plot, the thick solid lines indicate the 95% confidence interval of the differences and the thin solid line indicates the mean difference between the two methods.


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
Screening for DS in the second trimester of pregnancy by measurement of maternal serum biochemical markers is currently widely performed in many countries. However, the trend to move DS screening from the second to the first trimester is strong. This is not only attributable to the fact that first-trimester screening for DS based on fetal nuchal translucency and maternal serum biochemical markers would provide a higher detection rate than biochemical screening in the second trimester of pregnancy, but also to the fact that the implementation of a combined first-trimester screening program and its related diagnostic tests, such as amniocentesis and chorionic villus sampling, is becoming clinically more widespread. It is now generally believed that screening for DS in the first trimester would offer more advantages than screening in the second trimester (33), but this requires the availability of robust and well-characterized assays for the serologic markers, particularly because small increases in assay variation may considerably reduce performance (34).

Among the most important issues is to design a highly efficient, client-friendly prenatal screening program for DS in the first trimester. In most protocols of published first-trimester studies, the biochemical markers and the ultrasound result were examined in different units and the test results were subsequently collected for risk calculation. This procedure is technically very time-consuming. It is also psychologically stressful because a woman must first visit a doctor for pretest counseling and then go to the phlebotomy room where her blood sample is taken and transported to the clinical laboratory for the measurement of biochemical markers. After that, she visits the ultrasound unit where fetal nuchal translucency thickness and crown-rump length are measured. Finally, all the test results are delivered to the doctor for risk assessment. With this procedure, the woman cannot get her risk report until several days or weeks later, when she can schedule posttest counseling with the doctor.

Recently, Spencer et al. (22) proposed a new program in which measurements of biochemical markers and ultrasound markers are quickly performed in a single multidisciplinary clinic. Thus, the risk assessment and related posttest counseling can be completed within 1 h with just a single visit. Women with increased risk are then referred for diagnostic tests for fetal karyotyping. Spencer et al. implemented this program in a fairly large clinical trial involving a total of 4190 singleton pregnancies screened between 10 and 13 weeks of gestation. As a result, they obtained a detection rate of 86% for DS and 95% for all aneuploidies, demonstrating that the new program is feasible and effective. They further showed that integration of ultrasonographic, biochemical, and counseling services into a single clinic allowed more efficient use of total resources. Noticeably, this new procedure was also well accepted by the pregnant women; the percentage of women accepting the screening was high (98%). Furthermore, first-trimester screening is not influenced by mode of conception (35).

The assay developed here is intended for use in such a new screening program. Because it is based on the all-in-one reagent concept, the assay can be automated to yield a precise and sensitive result. Full automation reduces assay imprecision and simplifies the measurement. Consequently, the assay can be conducted by non-laboratory-trained individuals and still give results of adequate analytical quality. Because of the fast assay kinetics, the whole assay procedure is completed in 20 min. Thus, a quantitative test result should be available within 30 min after a blood sample is obtained.

Because >99% of PAPP-A exists as the PAPP-A/proMBP complex in the blood of pregnant women, polyclonal antisera raised against PAPP-A isolated from serum will invariably recognize the proMBP part of the complex as well as the mature eosinophil MBP (2). Unfortunately, serum concentrations of proMBP exceed those of PAPP-A by 4- to 10-fold on a molar basis throughout pregnancy (36). In addition, polyclonal antisera of different preparations have been shown to react with pregnancy-specific glycoprotein 1 and haptoglobin, which are also present in excess of PAPP-A in pregnancy serum (37)(38). Therefore, the PAPP-A values measured with the use of polyclonal antibodies may be severely biased unless an extensive adsorption procedure is applied to remove antibody factions reactive with these cross-reactants (7)(37). It has been demonstrated that, for DS screening, the performance of polyclonal antibody-based PAPP-A assays in which interfering substances are removed by adsorption methods was similar to that of assays based on the use of mAbs (16)(24)(37). However, such treatment is usually time-consuming and may cause considerable damage to the antibodies.

The use of mAbs has made it easier to develop a very specific assay that detects only PAPP-A rather than other blood components. The two mAbs used in the point-of-care assay are reactive only with the PAPP-A subunit of the complex. mAb 234-5, which serves as the detection antibody in the assay, in particular has been extensively characterized (39). No reaction has been observed between this antibody and a large number of human proteins, including proMBP, MBP, hepatoglobin, and pregnancy-specific glycoprotein 1. In addition, mAb 234-5 has been shown to stain the villous trophoblast in immunohistochemical studies of placental tissue. The assay constructed here with these two antibodies is thus very specific for the detection of PAPP-A in either free or complexed form.

PAPP-A is known to bind to heparin (40), and this property has often been used for the extraction and purification of PAPP-A. Although the amino acid sequence of PAPP-A is known, little is known about how and where the heparin molecules are bound to the PAPP-A complex. It was previously reported that heparin-derived plasma samples gave significantly higher PAPP-A values in an electroimmunoassay than did the corresponding serum samples (41)(42). The reason was attributed to a change in the electrophoretic mobility of PAPP-A caused by the binding of the highly charged heparin (41). In the present study, heparin was not found to have any effect on the point-of-care assay. PAPP-A concentrations determined in the presence of various amounts of heparin were the same as those determined in the absence of heparin. However, we did find that the assay configured with biotinylated mAb 234-4 and Eu3+-labeled mAb 234-2 was affected by heparin, leading to a substantially decreased measured PAPP-A value compared with that measured in the absence of heparin. Furthermore, this effect on the measurement of PAPP-A is dependent on the dose of heparin. The reason that heparin has no effect on the point-of-care assay is believed to be that the epitopes recognized by mAbs 10E1 and 234-5 are located at sites distant from the area to which heparin molecules bind on the PAPP-A molecule. Thus, binding of heparin does not affect the two antibodies that react with the respective epitopes. Whereas in the case of the assay constructed with biotinylated mAb 234-4 and Eu3+-labeled mAb 234-2, the epitopes recognized by either one or both antibodies are situated in or in close proximity to the heparin-binding area on the PAPP-A molecule. The binding of heparin thus masks the related epitopes, inhibiting binding of the antibodies to the PAPP-A molecule. Furthermore, the number of heparin-bound PAPP-A molecules relative to non-heparin-bound PAPP-A molecules is associated with the heparin concentrations in the samples. The reason that this heparin effect is concentration dependent is therefore understandable.

In studies using rocket immunoelectrophoresis (electroimmunoassays) or RIAs, it was reported that EDTA-derived plasma would yield significantly increased PAPP-A values compared with serum (41)(43). Until now there has been no plausible explanation available for the mechanism of this EDTA effect. In the present study, EDTA was found to affect the point-of-care assay, producing markedly reduced measured PAPP-A values. Furthermore, this effect was EDTA dose dependent. However, EDTA was not found to affect the assay configured with biotinylated mAb 10E1 and Eu3+-labeled mAb 234-2. Unlike the heparin effect, the mechanism of the EDTA effect is here thought to involve conformational changes in the PAPP-A molecular structure. As a strong chelator, EDTA can tightly chelate zinc ions. It is known that PAPP-A contains an elongated zinc-binding motif at position 482–492 of its 1547-residue sequence (44). The binding of zinc to this motif is considered important for the structural conformation as well as the proteolytic activity of PAPP-A. This idea is supported by the facts that PAPP-A proteolytic activity can be blocked by EDTA (3) and that substitution of Glu at position 483 with Ala can cause a complete loss of proteolytic activity (45). When EDTA chelates zinc ions from PAPP-A, conformational changes in the PAPP-A molecular structure could occur, which in turn could lead to the disappearance or modification of certain epitopes of the molecule. Consequently, some antibodies would be unable to react with these changed epitopes.

In RIAs or electroimmunoassays, reduced amounts of antigen-antibody immunocomplex are formed because some fractions of the antibodies are incapable of reacting with PAPP-A in the presence of EDTA, which leads to overestimated values derived from the calibration wells in which EDTA is not present. In the present study, the assay that uses biotinylated 10E1 and Eu3+-labeled 234-2 was insensitive to EDTA, meaning that the epitopes recognized by mAbs 10E1 and 234-2 were not affected by EDTA. The point-of-care assay based on the same capture antibody (biotinylated 10E1) but a different detection antibody (234-5) was sensitive to the EDTA effect, clearly indicating that the epitope recognized by mAb 234-5 is affected by EDTA. In addition, the fact that the EDTA effect was both dose and time dependent suggests that the interaction of EDTA with PAPP-A is rather slow.

Serum PAPP-A concentrations determined by a polyclonal antibody-based ELISA were previously shown to be stable after storage at 4 or 25 °C for 3 days (46). In the present study, PAPP-A in 6% BSA-TSA buffer was found to be stable at least for 3 months. Furthermore, heparin-derived plasma and heparin-derived whole blood samples could be left at 4 °C for 8 days with no change in the measured PAPP-A values. This was also true for serum samples but not for EDTA-derived plasma and EDTA-derived whole blood samples. In the latter case, PAPP-A values gradually decreased with time. Although this reduction in PAPP-A concentrations could not be reversed by the addition of calcium (data not shown), it does not necessarily indicate that PAPP-A molecules are broken down by the action of EDTA. On the basis of these findings, together with those observed in the investigation of anticoagulant effects, it is obvious that serum samples, heparin-derived whole blood samples, and heparin-derived plasma samples are the most appropriate sample materials for the present point-of-care assay for PAPP-A.

Because it uses a lanthanide chelate and time-resolved fluorometry, the point-of-care assay has all the characteristics typical of the time-resolved assays, such as a broad working range and high sensitivity. Compared with conventional time-resolved assays, the point-of-care assay can be performed much faster because of the favorable assay environment for kinetics and the right antibodies. In the point-of-care assay, the lanthanide chelate is fluorescent; its fluorescence is measured in situ on solid well surfaces with a surface measurement mode. This measurement mode eliminates the need for an additional liquid-handling step as is used in dissociation-enhanced lanthanide fluoroimmunoassays (Delfia) (47); at the same time, it avoids potential contamination from external lanthanide ions. Furthermore, this fluorescent chelate is also chemically highly stable; it does not interact with blood cells and any other blood constituents. This allows immunologic reactions to be performed even in the presence of blood cells. We have shown that heparin-derived whole blood is the correct sample material for the point-of-care assay because heparin has no effect on the assay. Importantly, the use of whole blood samples will further shorten the time required for obtaining a test result (i.e., decreased turnaround time) because the need for separation of blood cells from plasma or serum can be obviated. In addition, we have shown that there is an excellent correlation between whole blood samples and the relevant plasma samples; we believe that this is also the case for the relationship between whole blood samples and the corresponding serum samples, as has been demonstrated for other analytes (31)(48). It is noteworthy that after a Hct correction, the results obtained from the whole blood samples become fully comparable to the results obtained for the corresponding plasma samples. Thus, there may be no need to establish a new reference interval for whole blood samples because a conventionally used reference interval might also be valid for the Hct-corrected whole blood PAPP-A values.

Despite excellent correlation with the Delfia assay configured with mAbs 234-4 and 234-2, the point-of-care assay gives concentrations ~25% higher than those obtained with the Delfia assay. This fact is of no consequence for first-trimester screening because it is the multiple of the normal median, rather than the absolute concentration, that is used in screening programs to calculate the individual risk of having a DS-affected pregnancy. When unaffected and affected samples are measured with the point-of-care assay, the multiple of the normal median obtained for each sample would be very close to that obtained with the Delfia assay. Furthermore, we have previously shown that differences in PAPP-A concentrations determined by different mAb combinations did not have a significant effect on discrimination between DS and non-DS pregnancies (24).

The point-of-care assay described here, with a detection limit of 0.5 mIU/L, is among the most sensitive assays developed for PAPP-A to date. Its broad working range obviates the need for preassay dilution for first-trimester samples obtained at different gestational weeks. Because it is automated, the assay has rather low intra- and interassay CVs. These features, together with other performance characteristics, such as good recovery and parallelism, are very comparable to those of state-of-the-art laboratory immunoassays.

It is our belief that the present assay is particularly suited for use in one-stop clinics for screening for DS in the first trimester. The rapid and accurate test results obtained directly from whole blood samples combined with the results from other markers produce a risk assessment within a 1-h clinic visit. The decision that the pregnant woman with a positive screening result should have a further diagnostic test can be made before the patient has left the clinic. On the other hand, a screening-negative result will provide the woman with immediate relief from anxiety and stress. Thus, more efficient use of resources and better management of DS screening can be achieved. It is now necessary to test the assay in clinical conditions.

It was recently reported that circulating PAPP-A concentrations were significantly increased in patients with acute coronary syndromes (49). At a cutoff of 10 mIU/L, PAPP-A was shown to detect patients with acute coronary syndromes with a sensitivity of 89.2% and a specificity of 81.3%, suggesting that PAPP-A is a valuable marker for the identification of patients with acute coronary syndromes. The present assay completely covers the concentration range of interest and should thus be well suited for this new application.

In conclusion, we have developed a point-of-care time-resolved fluorometric assay for PAPP-A for use in one-stop clinics for first-trimester DS screening. The assay is sensitive and reproducible. The whole procedure is fully automated and is equally applicable to serum and heparin-derived plasma as well as to heparin-derived whole blood samples. An accurate and quantitative test result should be available within 30 min after a blood sample is obtained. This, together with other marker values, makes it possible to produce a risk assessment within 1 h in a one-stop clinic for screening for DS in the first trimester.


   Acknowledgments
 
This study was supported financially by a grant (Project 43025) from the Academy of Finland. We express our gratitude to Dr. Maria Severina (HyTest Oy, Turku, Finland), who kindly supported the study with the donation of antibodies.


   Footnotes
 
1 Nonstandard abbreviations: PAPP-A, pregnancy-associated plasma protein A; proMBP, proform of eosinophil major basic protein; IGFBP, insulin-like growth factor-binding protein; DS, Down syndrome; BITC, biotin isothiocyanate; BSA, bovine serum albumin; mAb, monoclonal antibody; TSA, Tris-saline-azide; and Hct, hematocrit.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Lin TM, Galbert SP, Kiefer D, Spellacy WN, Gall S. Characterization of four human pregnancy-associated plasma proteins. Am J Obstet Gynecol 1974;118:223-236.[ISI][Medline] [Order article via Infotrieve]
  2. Oxvig C, Sand O, Kristensen T, Gleich GJ, Sottrup-Jensen L. Circulating human pregnancy-associated plasma protein-A is disulfide-bridged to the proform of eosinophil major basic protein. J Biol Chem 1993;268:12243-12246.[Abstract/Free Full Text]
  3. Overgaard MT, Haaning J, Boldt HB, Olsen IM, Laursen LS, Christiansen M, et al. Expression of recombinant human pregnancy-associated plasma protein-A and identification of the proform of eosinophil major basic protein as its physiological inhibitor. J Biol Chem 2000;275:31128-31133.[Abstract/Free Full Text]
  4. Lawrence JB, Oxvig C, Overgaard MT, Sottrup-Jensen L, Gleich GJ, Hays LG, et al. The insulin-like growth factor (IGF)-dependent IGF binding protein-4 protease secreted by human fibroblasts is pregnancy-associated plasma protein-A. Proc Natl Acad Sci U S A 1999;96:3149-3153.[Abstract/Free Full Text]
  5. Qin X, Byun D, Lau KW, Baylink DJ, Mohan S. Evidence that the interaction between insulin-like growth factor (IGF)-II and IGF binding protein (IGFBP)-4 is essential for the action of the IGF-II-dependent IGFBP-4 protease. Arch Biochem Biophys 2000;379:209-216.[ISI][Medline] [Order article via Infotrieve]
  6. Laursen LS, Overgaard MT, Soe R, Boldt HB, Sottrup-Jensen L, Giudice LC, et al. Pregnancy-associated plasma protein-A (PAPP-A) cleaves insulin-like growth factor binding protein (IGFBP)-5 independent of IGF: implications for the mechanism of IGFBP-4 proteolysis by PAPP-A. FEBS Lett 2001;504:36-40.[ISI][Medline] [Order article via Infotrieve]
  7. Bonno M, Oxvig C, Kephart GM, Wagner JM, Kristensen T, Sottrup-Jensen L, et al. Localization of pregnancy-associated plasma protein-A and colocalization of pregnancy-associated plasma protein-A messenger ribonucleic acid and eosinophil granule major basic protein messenger ribonucleic acid in placenta. Lab Invest 1994;71:560-566.[ISI][Medline] [Order article via Infotrieve]
  8. Conover CA, Oxvig C, Overgaard MT, Christiansen M, Giudice LC. Evidence that the insulin-like growth factor binding protein-4 protease in human ovarian follicular fluid is pregnancy associated plasma protein-A. J Clin Endocrinol Metab 1999;84:4742-4745.[Abstract/Free Full Text]
  9. Bayes-Genis A, Schwartz RS, Lewis DA, Overgaard MT, Christiansen M, Oxvig C, et al. Insulin-like growth factor binding protein-4 protease produced by smooth muscle cells increases in the coronary artery after angioplasty. Arterioscler Thromb Vasc Biol 2001;21:335-341.[Abstract/Free Full Text]
  10. Westergaard JG, Sinosich MJ, Bugge M, Madsen LT, Teisner B, Grudzinskas JG. Pregnancy-associated plasma protein A in the prediction of early pregnancy failure. Am J Obstet Gynecol 1983;145:67-69.[ISI][Medline] [Order article via Infotrieve]
  11. Ruge S, Pedersen JF, Sorensen S, Lange AP. Can pregnancy-associated plasma protein A (PAPP-A) predict the outcome of pregnancy in women with threatened abortion and confirmed fetal viability?. Acta Obstet Gynecol Scand 1990;69:589-595.[Medline] [Order article via Infotrieve]
  12. Brambati B, Tului L, Bonacchi I, Shrimanker K, Suzuki Y, Grudzinskas JG. Serum PAPP-A and free ß-hCG are first-trimester screening markers for Down syndrome. Prenat Diagn 1994;14:1043-1047.[ISI][Medline] [Order article via Infotrieve]
  13. Spencer K, Aitken DA, Crossley JA, McCaw G, Berry E, Anderson R, et al. First trimester biochemical screening for trisomy 21: the role of free ß hCG, {alpha} fetoprotein and pregnancy associated plasma protein A. Ann Clin Biochem 1994;31:447-454.
  14. Wald NJ, George L, Smith D, Densem JW, Petterson K. Serum screening for Down’s syndrome between 8 and 14 weeks of pregnancy. International Prenatal Screening Research Group. Br J Obstet Gynaecol 1996;103:407-412.[ISI][Medline] [Order article via Infotrieve]
  15. Qin QP, Nguyen TH, Christiansen M, Larsen SO, Norgaard-Pedersen B. Time-resolved immunofluorometric assay of pregnancy-associated plasma protein A in maternal serum screening for Down’s syndrome in first trimester of pregnancy. Clin Chim Acta 1996;254:113-129.[ISI][Medline] [Order article via Infotrieve]
  16. Haddow JE, Palomaki GE, Knight GJ, Williams J, Miller WA, Johnson A. Screening of maternal serum for fetal Down’s syndrome in the first trimester. N Engl J Med 1998;338:955-961.[Abstract/Free Full Text]
  17. Tolmie J. Down syndrome and other autosomal trisomies. Rimoin DL Conner JM Pyeritz RE eds. Principles and practice of medical genetics, Vol. 1 3rd ed. 1996:925-971 Churchill Livingstone New York. .
  18. Wald NJ, Densem JW, Smith D, Klee GG. Four-marker serum screening for Down’s syndrome. Prenat Diagn 1994;14:707-716.[ISI][Medline] [Order article via Infotrieve]
  19. De Graaf IM, Pajkrt E, Bilardo CM, Leschot NJ, Cuckle HS, van Lith JM. Early pregnancy screening for fetal aneuploidy with serum markers and nuchal translucency. Prenat Diagn 1999;19:458-462.[ISI][Medline] [Order article via Infotrieve]
  20. De Biasio P, Siccardi M, Volpe G, Famularo L, Santi F, Canini S. First-trimester screening for Down syndrome using nuchal translucency measurement with free ß-hCG and PAPP-A between 10 and 13 weeks of pregnancy—the combined test. Prenat Diagn 1999;19:360-363.[ISI][Medline] [Order article via Infotrieve]
  21. Krantz DA, Hallahan TW, Orlandi F, Buchanan P, Larsen JW, Jr, Macri JN. First-trimester Down syndrome screening using dried blood biochemistry and nuchal translucency. Obstet Gynecol 2000;96:207-213.[Abstract/Free Full Text]
  22. Spencer K, Spencer CE, Power M, Moakes A, Nicolaides KH. One stop clinic for assessment of risk for fetal anomalies: a report of the first year of prospective screening for chromosomal anomalies in the first trimester. BJOG 2000;107:1271-1275.[Medline] [Order article via Infotrieve]
  23. Wadhera S, Millar WJ. Second trimester abortions: trends and medical complications. Health Rep 1994;6:441-454.[Medline] [Order article via Infotrieve]
  24. Qin QP, Christiansen M, Oxvig C, Pettersson K, Sottrup-Jensen L, Koch C, et al. Double-monoclonal immunofluorometric assays for pregnancy-associated plasma protein A/proeosinophil major basic protein (PAPP-A/proMBP) complex in first-trimester maternal serum screening for Down syndrome. Clin Chem 1997;43:2323-2332.[Abstract/Free Full Text]
  25. Qin QP, Christiansen M, Lövgren T, Norgaard-Pedersen B, Pettersson K. Dual-label time-resolved immunofluorometric assay for simultaneous determination of pregnancy-associated plasma protein A and free ß-subunit of human chorionic gonadotrophin. J Immunol Methods 1997;205:169-175.[ISI][Medline] [Order article via Infotrieve]
  26. Tsukerman GL, Gusina NB, Cuckle HS. Maternal serum screening for Down syndrome in the first trimester: experience from Belarus. Prenat Diagn 1999;19:499-504.[ISI][Medline] [Order article via Infotrieve]
  27. Casals E, Aibar C, Martinez JM, Borrell A, Soler A, Ojuel J, et al. First-trimester biochemical markers for Down syndrome. Prenat Diagn 1999;19:8-11.[ISI][Medline] [Order article via Infotrieve]
  28. Niemimaa M, Suonpaa M, Perheentupa A, Seppala M, Heinonen S, Laitinen P, et al. Evaluation of first trimester maternal serum and ultrasound screening for Down’s syndrome in Eastern and Northern Finland. Eur J Hum Genet 2001;9:404-408.[ISI][Medline] [Order article via Infotrieve]
  29. Lövgren T, Merio L, Mitrunen K, Makinen ML, Makela M, Blomberg K, et al. One-step all-in-one dry reagent immunoassays with fluorescent europium chelate label and time-resolved fluorometry. Clin Chem 1996;42:1196-1201.[Abstract/Free Full Text]
  30. Pettersson K, Soderholm JR. Ultrasensitive two-site immunometric assay of human lutropin by time-resolved fluorometry. Clin Chem 1990;36:1928-1933.[Abstract]
  31. Pettersson K, Katajamaki T, Irjala K, Leppanen V, Majamaa-Voltti K, Laitinen P. Time-resolved fluorometry (TRF)-based immunoassay concept for rapid and quantitative determination of biochemical myocardial infarction markers from whole blood, serum and plasma. Luminescence 2000;15:399-407.[ISI][Medline] [Order article via Infotrieve]
  32. Sadler WA, Murray LM, Turner JG. Influence of specimen carryover on sensitive thyrotropin (TSH) assays: is there a problem?. Clin Chem 1996;42:593-597.[Abstract/Free Full Text]
  33. . American College of Obstetricians and Gynecologists.. First-trimester screening for fetal anomalies with nuchal translucency. Committee on genetics [Opinion no. 223]. 1999 American College of Obstetricians and Gynecologists Washington. .
  34. Christiansen M, Larsen SO. Precision of risk in multidimensional likelihood ratio based screening with first trimester screening for Down’s syndrome as a special case. J Med Screen 2001;8:61-64.[Abstract/Free Full Text]
  35. Wojdemann KR, Larsen SO, Shalmi A, Sundberg K, Christiansen M, Tabor A. First trimester screening for Down syndrome and assisted reproduction: no basis for concern. Prenat Diagn 2001;21:563-565.[ISI][Medline] [Order article via Infotrieve]
  36. Oxvig C, Haaning J, Kristensen L, Wagner JM, Rubin I, Stigbrand T, et al. Identification of angiotensinogen and complement C3dg as novel proteins binding the proform of eosinophil major basic protein in human pregnancy serum and plasma. J Biol Chem 1995;270:13645-13651.[Abstract/Free Full Text]
  37. Bersinger NA, Zakher A, Huber U, Pescia G, Schneider H. A sensitive enzyme immunoassay for pregnancy-associated plasma protein A (PAPP-A): a possible first trimester method of screening for Down syndrome and other trisomies. Arch Gynecol Obstet 1995;256:185-192.[ISI][Medline] [Order article via Infotrieve]
  38. Bueler MR, Bersinger NA. Antiserum to pregnancy-associated plasma protein A (PAPP-A) recognizes human haptoglobin. Br J Obstet Gynaecol 1989;96:867-869.[ISI][Medline] [Order article via Infotrieve]
  39. Qin QP. Maternal serum screening for Down syndrome in the first trimester with special emphasis on pregnancy associated plasma protein A [PhD thesis]. 1998:144pp University of Turku Turku, Finland. .
  40. Sinosich MJ, Teisner B, Davey M, Grudzinskas JG. Pregnancy associated plasma protein-A: interaction with heparin in crossed affinity immunoelectrophoresis. Aust N Z J Med 1981;11:429-433.[ISI][Medline] [Order article via Infotrieve]
  41. Toop KM, Klopper A. Effect of anticoagulants on the measurement of pregnancy-associated plasma protein-A (PAPP-A). Br J Obstet Gynaecol 1983;90:150-155.[ISI][Medline] [Order article via Infotrieve]
  42. Westergaard JG, Hau J, Teisner B, Grudzinskas JG. Specific and reversible interaction between pregnancy-associated plasma protein A and heparin. Placenta 1983;4:13-18.[ISI][Medline] [Order article via Infotrieve]
  43. Pinto Furtado LG, Bolton AE, Grudzinskas JG, Chapman MG, Sinosich MJ, Sharma V. The development and validation of a RIA for human pregnancy-associated plasma protein A (PAPP-A). Arch Gynecol 1984;236:83-91.[ISI][Medline] [Order article via Infotrieve]
  44. Kristensen T, Oxvig C, Sand O, Moller NP, Sottrup-Jensen L. Amino acid sequence of human pregnancy-associated plasma protein-A derived from cloned cDNA. Biochemistry 1994;33:1592-1598.[Medline] [Order article via Infotrieve]
  45. Boldt HB, Overgaard MT, Laursen LS, Weyer K, Sottrup-Jensen L, Oxvig C. Mutational analysis of the proteolytic domain of pregnancy-associated plasma protein-A (PAPP-A): classification as a metzincin. Biochem J 2001;358:359-367.[ISI][Medline] [Order article via Infotrieve]
  46. Bersinger NA, Marguerat P, Pescia G, Schneider H. Pregnancy-associated plasma protein A (PAPP-A): measurement by highly sensitive and specific enzyme immunoassay, importance of first-trimester serum determinations, and stability studies. Reprod Fertil Dev 1995;7:1419-1423.[Medline] [Order article via Infotrieve]
  47. Lövgren T, Pettersson K. Time-resolved fluoroimmunoassay, advantages and limitations. Van Dyke K Van Dyke R eds. Luminescence immunoassay and molecular allocations 1990:233-253 CRC Press Boca Raton, FL. .
  48. Piironen T, Nurmi M, Irjala K, Heinonen O, Lilja H, Lovgren T, et al. Measurement of circulating forms of prostate-specific antigen in whole blood immediately after venipuncture: implications for point-of-care testing. Clin Chem 2001;47:703-711.[Abstract/Free Full Text]
  49. Bayes-Genis A, Conover CA, Overgaard MT, Bailey KR, Christiansen M, Holmes DR, Jr, et al. Pregnancy-associated plasma protein A as a marker of acute coronary syndromes. N Engl J Med 2001;345:1022-1029.[Abstract/Free Full Text]



The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
S. Wittfooth, Q.-P. Qin, J. Lund, I. Tierala, K. Pulkki, H. Takalo, and K. Pettersson
Immunofluorometric Point-of-Care Assays for the Detection of Acute Coronary Syndrome-Related Noncomplexed Pregnancy-Associated Plasma Protein A
Clin. Chem., September 1, 2006; 52(9): 1794 - 1801.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
Q.-P. Qin, S. Kokkala, J. Lund, N. Tamm, X. Qin, M. Lepantalo, and K. Pettersson
Immunoassays Developed for Pregnancy-Associated Plasma Protein-A (PAPP-A) in Pregnancy May Not Recognize PAPP-A in Acute Coronary Syndromes
Clin. Chem., March 1, 2006; 52(3): 398 - 404.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
F. S. Apple, A. H.B. Wu, J. Mair, J. Ravkilde, M. Panteghini, J. Tate, F. Pagani, R. H. Christenson, M. Mockel, O. Danne, et al.
Future Biomarkers for Detection of Ischemia and Risk Stratification in Acute Coronary Syndrome
Clin. Chem., May 1, 2005; 51(5): 810 - 824.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
Q.-P. Qin, S. Kokkala, J. Lund, N. Tamm, L.-M. Voipio-Pulkki, and K. Pettersson
Molecular Distinction of Circulating Pregnancy-Associated Plasma Protein A in Myocardial Infarction and Pregnancy
Clin. Chem., January 1, 2005; 51(1): 75 - 83.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
P. Huhtinen, A.-M. Pelkkikangas, S. Jaakohuhta, T. Lovgren, and H. Harma
Quantitative, Rapid Europium(III) Nanoparticle-Label-Based All-in-One Dry-Reagent Immunoassay for Thyroid-Stimulating Hormone
Clin. Chem., October 1, 2004; 50(10): 1935 - 1936.
[Full Text] [PDF]


Home page
CirculationHome page
J. Lund, Q.-P. Qin, T. Ilva, K. Pettersson, L.-M. Voipio-Pulkki, P. Porela, and K. Pulkki
Circulating Pregnancy-Associated Plasma Protein A Predicts Outcome in Patients With Acute Coronary Syndrome but No Troponin I Elevation
Circulation, October 21, 2003; 108(16): 1924 - 1926.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
Q.-P. Qin, O. Peltola, and K. Pettersson
Time-resolved Fluorescence Resonance Energy Transfer Assay for Point-of-Care Testing of Urinary Albumin
Clin. Chem., July 1, 2003; 49(7): 1105 - 1113.
[Abstract] [Full Text] [PDF]