1 The Research and Development Group, Department of Clinical Chemistry, Ullevål University Hospital, 0407 Oslo, Norway.
aAuthor for correspondence. Fax 47-22118189; e-mail reidun.ovstebo{at}ioks.uio.no.
| Abstract |
|---|
|
|
|---|
Methods: Real-time PCR-based calibration curves for a target gene, tissue factor (TF), and a reference gene, ß-actin, generated from PCR amplicons were evaluated by running cDNA controls. In addition, the reverse transcription step was evaluated by running mRNA controls. Amplification efficiencies of calibrators and targets were determined. Variances within and between runs were estimated, and power statistics were applied to determine the concentration differences that could reliably be detected.
Results: Within- and between-run variations (CVs) of cDNA controls (TF and ß-actin), extrapolated from reproducible calibration curves (CVs of slopes, 4.3% and 2.7%, respectively) were 410% (within) and 1538% (between) using both daily and "grand mean" calibration curves. CVs for the ß-actin mRNA controls were 12% (within) and 1928% (between). Estimates of each steps contribution to the total variation were as follows: CVRT-PCR, 28%; CVPCR, 15%; CVRT, 23% (difference between CVRT-PCR and CVPCR). PCR efficiencies were as follows: ß-actin calibrator/target, 1.96/1.95; TF calibrator/target, 1.95/1.93. Duplicate measurements could detect a twofold concentration difference (power, 0.8).
Conclusions: Daily PCR calibration curves generated from PCR amplicons were reproducible, allowing the use of a grand mean calibration curve. The reverse transcription step contributes the most to the total variation. By determining a systems total variance, power analysis may be used to disclose differences that can be reliably detected at a specified power.
| Introduction |
|---|
|
|
|---|
Quantification of mRNAs in a cell lysate requires that the lysate is accompanied by appropriate calibrators and controls (both mRNA and cDNA), which should all be included in the same run (12)(17). Homologous calibrators may be generated either by use of linearized vectors containing the cDNA sequence of interest or obtained by cloning (18)(19) or by use of specific PCR amplicons. Because cloning is more labor-intensive, we have developed calibrators based on purified, specific PCR amplicons.
The aim of this study was to develop and evaluate a protocol to quantify any mRNA in monocyte lysates based on reverse transcription and real-time PCR. For these purposes, we produced specific PCR amplicons to generate calibration curves suitable for quantification of amplified specific cDNAs. Individual PCR efficiencies were determined for targets and calibrators. Reproducibility was evaluated by running individual calibration curves over a period of 1 year and by quantifying specific transcripts in long-term mRNA as well as cDNA controls. The minimum difference that could be reliably detected was determined by estimating the variances (20) and by applying power statistics to the system (21). Variations attributable to mRNA isolation and RT-PCR efficiencies were adjusted by simultaneous quantification of the endogenous housekeeping gene (ß-actin or ß2-microglobulin) in the same lysate. To date, we have successfully produced calibration curves for the quantification of cDNA for tissue factor (TF), tumor necrosis factor-
, urokinase plasminogen activator receptor, interleukin-10 (IL-10), MCP-1, ß2-microglobulin, and ß-actin. The results for ß-actin and TF quantification are presented here.
| Materials and Methods |
|---|
|
|
|---|
Monocyte stimulation.
Human monocytes respond to lipopolysaccharide (LPS) stimulation by the synthesis and release of immunoregulatory and inflammatory mediators such as IL-1, IL-6, IL-8, and tumor necrosis factor-
(23)(24). This approach was used to produce specific cDNA calibrators for specific mRNA quantification from genes that are not constitutively expressed in human monocytes. Monocytes were thawed, resuspended in 50 mL/L acid-treated fetal calf serum in RPMI containing 20 mL/L of a penicillin/streptomycin solution (cat. no. P0906, containing 5000 units/mL penicillin and 5 mg/mL streptomycin; Sigma) and seeded at a density of 5 x 104 cells in microtiter plates (Costar 3595; Corning). Cells were cultured in the absence or presence of 300 pg of Escherichia coli LPS (0.55:B5; BioWhittaker) per well in a total volume of 200 µL for 3 h at 37 °C. Cells were harvested for mRNA isolation by centrifugation (500g for 7 min at 20 °C), the medium was aspirated, and the cells were lysed in 150 µL of lysis/binding buffer according to the instructions accompanying the Dynabeads mRNA Purification Kit® (product no 610.12, Dynal Biotech ASA).
rt-pcr
mRNA isolation.
mRNA was isolated using the Dynabeads mRNA Purification Kit (25)(26).
Reverse transcription reaction.
The enzymatic conversion of a heterogeneous mRNA population to a high-quality cDNA library was performed by first testing various commercially available reverse transcriptases: Omniscript RT Kit (cat. no. 205111; Qiagen); First Strand cDNA Synthesis Kit for RT-PCR (AMV; cat. no. 1 483 188; Roche Diagnostics GmbH); Superscript (Life Technologies); and Ready · To · Go T-primed First Strand Kit (cat. no. 27-9263-01; Amersham Pharmacia Biotech). Their efficiency, sensitivity, parallellism, linearity, and convenience of use were validated. ß-Actin mRNA isolated from diluted solutions of human monocytes and commercially available mRNA from human liver cells (PolyA+ RNA, Adult Liver, Human; cat. no 775013; Stratagene) were quantified using reverse transcription and PCR ß-actin calibration curves. Each reverse transcription enzyme was handled according to the manufacturers instructions. In our systems, Omniscript reverse transcriptase was the best according to the criteria (see above) set out in our work, and reverse transcription was performed as follows: 8 µL of mRNA solution was added to 12 µL of reverse transcription mixture [2.5 mM oligo(dT) primer; 1 unit of RNase inhibitor (PE Applied Biosystems); 10x reverse transcription buffer; 0.5 mM each of dATP, dCTP, dGTP, and dTTP; 4 U of Omniscript Reverse Transcriptase; and RNase-free water (Omniscript RT Kit; cat. no. 205111)]. The reverse transcription step was performed in a gradient cycler (Gene Amp PCR System 2400; PE Applied Biosystems) at 37 °C for 60 min and 95 °C for 5 min. The cDNA libraries were stored at 4 °C.
PCR.
We added 2-µL of a cDNA mixture to 18 µL of PCR Mastermixture (3 mM MgCl2, 0.25 mM primers, 2 µL of LightCyclerTM Fast Start Master SYBR Green), and amplification was performed in a LightCycler (cat. no. 2239264; Roche Diagnostics). PCR primer design and optimizations were carried out with software downloaded from the Internet Center for Genome Research (http://www-genome.wi.mit.edu/genome_software/other/primer3.html). GenBank accession no. J02846 was used for designing TF primers, and D28354 was used for ß-actin. Primers were designed to distinguish cDNA and genomic DNA/pseudogenes (27). The following primers were used: 5'-ggcatcctcaccctgaagta-3' and 5'-ggggtgttgaaggtctcaaa-3' for ß-actin and 5'-gaccgtagaagatgaacggact-3' and 5'-ggagggaatcactgcttgaa-3' for TF. The amplicons were sequenced (CEQ200), and homology with other sequences was examined using the NCBI gene BLASTTM website (www.ncbi.nlm.nih.gov/blast).
LightCycler protocol.
A four-segment LightCycler PCR amplification and melting curve protocol was used for online detection of ß-actin and TF cDNA, using SYBR Green I. The protocol included a 10-min denaturation step at 95 °C to activate the DNA polymerase, followed by 35 amplification cycles at 95 °C for 1 s, 60 °C for 10 s, and 72 °C for 10 s. Melting curve analyses were performed at 95 °C for 0 s, 65 °C for 15 s, and 95 °C for 0 s. Temperature transition rates were set to 20 °C/s in all segments except the final insert in the melting curve analysis, for which the rate was 0.1 °C/s. The protocol ended with a cooling segment (40 °C). LightCycler software version 3.01 was used to estimate the threshold cycle numbers (crossing points; Cp). The reproducibilities of the calibration curves (slopes and intercepts) and controls were estimated by use of the "second derivative maximum method", in which the software based on default values determines the log-linear area. The limiting dilutions were calculated by the "fit points" method, in which the user may decide the log-linear part and thereby increase the sensitivity.
preparation of calibrators
mRNA was isolated from human monocytes incubated with LPS as described above. A cDNA library was synthesized (see above) and used to produce the different calibrators. Specific PCR primers for each calibrator were used in large-scale amplification to obtain quantities of the specific PCR amplicons measurable by ultraviolet detection. Each tube was examined by melting point analysis to assure that only one specific product was present before it was included in the PCR amplicon pool.
Purification of PCR calibrators.
The pools of PCR amplicons were purified conventionally by phenol-chloroform extraction and ethanol precipitation and dissolved in 100 µL of H2O.
Quantification of PCR calibrators by ultraviolet spectrophotometry and limiting dilutions.
We measured the absorbance of 70 µL of the purified PCR calibrators at 260/280 nm and calculated the PCR amplicon concentrations and purity. We checked the amplicon numbers, calculated from the DNA concentration by taking into account the size of each specific PCR calibrator (ß-actin, 203 bp; TF, 211 bp), the mean molecular weight of the nucleotide bases (340.5 g/mol), Avogadros number (6.022 x 1023), and the volume, by analyzing several dilutions containing amplicon numbers down to 10.1, using the fit point method and the melting curve analysis to search for amplification products (28).
Testing diluents: effects of various matrixes.
Known amounts of TF amplicons were added to different solutions [water and MS2 RNA (10 µg/mL; article no. 165948; Roche Diagnostics)] and a nonexpressing TF cDNA library (synthesized from human monocytes and examined for the presence of TF) to examine the matrix effect. PCR was performed, and the Cp values were recorded.
Dilution of PCR calibrators to generate calibration curves.
After ultraviolet quantification, the PCR calibrators were diluted severalfold with carrier nucleic acid solution [MS2 RNA (10 µg/mL)] in siliconized tubes (Corning Costar) until they reached the concentration that became the highest calibrator (1011 to
105 amplicons). We then prepared 10-fold dilutions from the highest calibrator until no theoretical amplicon could be obtained. Generally five calibrators were used for each calibration curve. The calibrators and LightCycler Fast Start Master SYBR Green mixture were prepared in separate rooms. The reproducibilities of the calibration curves were examined by calculating mean and range values for each dilution (Fig. 1
). The Cp values for the calibrators obtained from each run were examined for outliers by use of Grubbs test (GraphPad).
|
Handling and storage of cDNA for calibration curves.
The stock calibrators [1011 copies/µL diluted in MS2 RNA solution (10 µg/mL)] and stored at -70 °C. After thawing, the working calibrator solutions (105) were stored at 4 °C for 1 month. From these working solutions dilution series were made for each run.
controls
Lysate-based cDNA libraries.
The cDNA controls were made by synthesizing cDNA libraries from mRNAs isolated from different numbers (2.510 x 104) of LPS-stimulated human monocytes. The controls were stored at 4 °C. Target values as well as within- and between-run variation were calculated based on both daily calibration curves and a "grand mean" calibration curve. (Because the present LightCycler Software lacks the ability to import a grand mean calibration curve, a representative calibration curve was chosen as the grand mean calibration curve.)
mRNA controls.
PolyA+ RNA from adult human liver cells (cat. no. 775013; Stratagene) was diluted in RNase-free water to a concentration of 1 ng/µL, aliquoted, and stored at -70 °C. An aliquot was thawed and diluted further to concentrations of 200 and 20 pg/µL, and all blends were included each time a reverse transcription reaction was performed.
pcr amplification efficiencies
The PCR efficiencies were determined in calibrators and targets (cDNA libraries obtained from LPS-stimulated human monocytes, containing both the target and housekeeping genes TF and ß-actin, respectively). Serial dilutions of targets and calibrators (in 10 µg/mL MS2 RNA solution) were PCR-amplified, and the Cp values of each dilution were recorded. Cp values were plotted against the log of the initial template concentrations. Slopes were calculated using Statgraphics® Plus (Manugistics). PCR efficiencies were calculated as E = 10-1/slope (29)(30).
Correction for differences in the PCR amplification efficiencies.
The calculation of efficiency-corrected results is based on the equation: T = T0 x ECp, where T is the number of amplicons synthesized when the Cp is at the detection threshold, E is the efficiency of the calibrator or the target, and T0 is the initial number of molecules (4)(29). Assuming that the number of amplicons synthesized at the Cp should be the same in the calibrator and the target reaction, the following equation may be used to correct for the systematic error caused by differences in PCR efficiencies in the calibrators and targets:
![]() |
experimental design for estimation of variances and power statistics in the present rt-pcr assay
Six replicates of three dilutions (giving 8, 2.7, and 0.9 ng of mRNA in the reverse transcription step) of PolyA+ RNA (from adult human liver cells; cat. no. 775013; Stratagene), aliquoted (1 ng/µL) and stored at -70 °C, were analyzed for ß-actin mRNA by RT-PCR on 6 separate days.
Variance components (within- and between-day) and total variance were estimated by nested ANOVA (20). The F-test was used to compare variances in the different sets of concentrations (20). The sample size needed to detect differences with a certain reliability was estimated by computer software (Sample Power Release 2.0 in SPSS 11.0; SPSS Inc.).
| Results |
|---|
|
|
|---|
choice of diluent solutions
The amplification efficiency, which is the fraction of templates replicated during each reaction cycle, is a crucial factor for the reliability of quantitative PCR (15)(16)(17)(31). Because the efficiency of the PCR reaction may be influenced by several factors, such as diluent composition (15), different solutions were examined by amplifying known amounts of TF amplicons in different matrix solutions. It appeared that the Cp values differed depending on diluent composition (data not shown). Highly consistent results, however, were obtained when we diluted the targets in either MS2 RNA solution or a solution similar to our cDNA libraries (nonexpressing TF cDNA library). The usefulness of MS2 RNA (10 µg/mL) as a stabilizer was further evaluated by estimating the ß-actin cDNA numbers in various mixtures of MS2 RNA and cDNA library solution. It appeared that the MS2 RNA solution over a wide range gave similar ß-actin cDNA results on PCR amplification (data not shown). In addition, these results also indicated that there was no inhibition of the PCR.
reproducibility of the calibration curves
Calibration curve performance was evaluated with use of data collected from 57 ß-actin and 14 TF calibration curves over a period of 1 year. Between-day variations, calculated from the y-intercepts (mean Cp values, 36.00 and 37.05) and slopes (mean, -3.46 and -3.43; given by the LightCycler Software after each new run) were 24% and 48% for ß-actin and TF, respectively. Mean and range values for each calibrators Cp values were calculated to show the reproducibility (Fig. 1
). As shown in Fig. 1
, the calibration curves were linear over four orders of magnitude, which indicates constant PCR efficiency over the concentration range studied. We improved reproducibility by increasing the concentration of calibrators in the PCR. A search for outliers in each series of calibrators, using Grubbs test, showed no significant outliers (P >0.05). The lower limits of the method using the "standard deviation method" were 172 copies for ß-actin and 118 for TF (see Materials and Methods).
pcr amplification efficiency
The Cp values of the targets were converted to concentrations with use of our calibration curves. The amplification efficiencies of the calibrators and targets must be the same to make these readings. Similar slopes in plots of dilutions of calibrators and targets indicated equal efficiencies. A linear plot indicated constant efficiency over the concentration range studied. However, because the efficiencies of the calibrators and the targets may differ, we corrected the target results to observe the consequences (Table 1
). The PCR efficiencies were calculated to 1.96 (range, 1.842.07; n = 57) and 1.95 (range, 1.842.03; n = 7) for the ß-actin calibrators and targets and 1.95 (range, 1.842.03; n = 15) and 1.93 (range, 1.921.95; n = 7) for the TF calibrator and target, respectively. We found no significant differences between the PCR efficiencies of ß-actin and TF targets (P = 0.72).
|
evaluation of the pcr calibration curves
The reproducibility of measured ß-actin (n = 57) and TF (n = 15) concentrations in lysate-based cDNA controls at different concentrations were estimated either by use of daily run calibration curves or a grand mean calibration curve (see Materials and Methods). Within-run variation (CV) was 410%, whereas the between-run CV was 1838% for ß-actin and TF. Similar results were obtained when we used a grand mean calibration curve, indicating that the system is sufficiently stable for the same calibration curve to be used over time (Table 1
).
evaluation of rt-pcr
To evaluate the variation in both the reverse transcription and the PCR step, we quantified ß-actin mRNA in dilutions of mRNA (isolated from commercial human liver cells), using both the individual calibration curves and a combined calibration curve (Table 1
). The within-run CV was 12%, and the between-run CV was 1928%, depending on input target concentrations. ß-Actin concentrations obtained from dilutions of mRNA (20 pg/µL, 200 pg/µL, and 1 ng/µL) showed a linear relationship, indicating that the RT-PCR are similar over the concentration range studied.
reliable detection of concentration differences
The minimum difference that can be detected in our system depends on the variances and the power (reliability) chosen. Variance component analysis was performed on data collected from experiments on 6 separate days (see Materials and Methods and Table 2
). There were no statistical differences in the concentration range studied. Using within-run data at a power of 0.8, we needed duplicate measurements to detect differences of 100%. A sample size of four would be needed within a run to detect differences of 40%. The total variation in the system indicated that a sample size of three would be needed to detect differences at 100%; however, to detect a difference as small as 40%, a sample size of seven would be required (Table 3
).
|
|
| Discussion |
|---|
|
|
|---|
Our homologous PCR calibrators consist of purified, specific PCR amplicons, compared with the calibrators used by other authors, which were vectors containing the sequence of interest (18)(32)(34)(35). The choice of purification method is of great importance because of the reliability of the A260 measurements. If a cDNA library is used as target, cDNA content should be as small as possible if phenol-chloroform extraction and ethanol precipitation are used. Another method could be purification of the PCR amplicons from a gel.
To evaluate the between-run reproducibility of the calibration curves, we collected data over a period of 1 year. The between-run variation (CV) of the slopes of the calibration curves was 4% for ß-actin and 8% for TF. The between-run variation (CV) of the y-intercept was 2% and 4%, respectively, and in line with results reported by Kruse et al. (19). This indicates that highly reproducible calibration curves over at least four orders of magnitude may be generated based on the synthesis of PCR amplicons.
To evaluate the LightCycler PCR performances, we ran long-term lysate-based cDNA controls at different concentrations for two cDNAs (ß-actin and TF). The controls were lysate based to closely mimic the matrix of the samples to be quantified. As expected, the concentrations affected the reproducibility. In line with the observations of Rasmussen (29), we found within-run variations (CVs) between 4% and 11%, whereas the between-run variations ranged from 18% to 38% (Table 1
). To the best of our knowledge, very little information has been reported concerning between-run variation when these types of controls and concentrations are used (9)(19).
According to the basic PCR equation, the generated copy number (T) at a certain cycle is a function of the initial copy number (T0), the PCR efficiency (E), and the cycle number (Cp) (17)(29). The efficiency of the PCR describes the kinetics during the reaction, and maximum efficiency (E = 2) is seldom achieved because of various inhibiting factors, such as dilution composition and GC content (17). Separate PCR efficiencies should determined for targets and calibrators to allow for conversion of the Cp values to concentrations based on calibration curves. The PCR efficiencies for targets and calibrators in our systems (1.96/1.95 for ß-actin calibrator/target and 1.95/1.93 for TF calibrator/target) showed small differences that may be attributable to variations in dilution composition and the more complex nature of the cDNA library as a starting material. Slack et al. (32) determined that PCR efficiencies for their calibrators and targets were 98% and 94%, respectively (E = 1.96 and 1.88, using our terms), and "suggested this as little to no decrease in PCR efficiencies". The company Roche (30) advises their customers working with relative quantification and external calibrators to use a common calibration curve for any target as long as the amplification efficiency does not differ more than 0.05. The effect of the differences in PCR efficiencies is exponentially dependent on the cycle number and increases with each cycle. We found no significant difference in PCR efficiencies between the targets presented in this report. However, one should be aware that quantification of other targets may give significant differences in PCR efficiencies compared with the housekeeping gene. The need for an efficiency-correcting calculation will then depend on the accuracy required for a given application. This can be done by multiplying the results by a given factor (Materials and Methods).
The reverse transcription is another critical step in quantitative analyses (4). This step converts a complex mRNA population into a cDNA library that can be analyzed further by either microarrays or PCR. To evaluate this step, we used RT-PCR to quantify the ß-actin mRNA concentrations in solutions containing known amounts of mRNA. Commercially available reverse transcriptases do not always convert the mRNAs to cDNA libraries in a satisfactory way (data not shown). In line with results reported by others (33), we found that Omniscript was the optimal enzyme, based on our criteria (see Materials and Methods). This was used to further evaluate variations in the reverse transcription step. The variations, which include variations both from the reverse transcription and PCR steps, were 12% (within-run CV) and 19% and 25% (between-run CV; Table 1
). Kleiber et al. (6) presented data on the reproducibility in their one-step TaqMan HCV RT-PCR test (within-run CV, 1726%; between-run CV, 2130%; n = 60) which were slightly higher than in our system. Slack et al. (32), who used a two-step system, also reported between-run variations slightly higher than ours (3143%; n = 4). Both the reverse transcription and the PCR steps will contribute to the total variation of the described RT-PCR method (36). On the basis of our CVs, exemplified by the cDNA control with a T0 mean concentration of 6.5 x 103 amplicons and a between-day CV [CVb (PCR)] of 15% and the mRNA control with a T0 mean concentration of 6.2 x 103 amplicons and a CVb (RT-PCR) of 28% (Table 1
), and with use of ANOVA, the reverse transcription reaction appears to contribute the most to the overall variation (CVRT = 23%).
Our CVs are calculated from concentrations of a given transcript in the samples. If CVs instead are calculated from the Cp values (equivalent to a logarithm of the concentration), the CVs necessarily are substantially lower. Thus, our data, using the Cp values for calculation, indicate a CVPCR of 0.7%, CVRT of 1.8%, and CVRT-PCR of 2.0% compared with a CVPCR of 15%, CVRT of 23%, and CVRT-PCR of 28% observed when the calculated concentration of transcripts derived from the calibration curves is used. Because CVs are calculated to predict the stability of the analysis when comparing analyte concentrations, we believe that the CV should be reported accordingly.
The probability of detecting a true difference of a given magnitude can be determined using test power analysis (21). Such analyses give the sample size needed to obtain a specified difference with a certain reliability or power. Determination of the minimum difference in mRNA molecules that could be differentiated in our RT-PCR system was performed by quantifying the ß-actin concentrations in diluted mRNA solutions in an experiment designed to determine the within- and between-run variability (see Materials and Methods and Table 3
). These results indicate that to reliably detect, at a power of 0.8, a twofold difference (100%) between samples, duplicate measurements have to be used. However, to detect a difference as small as 40%, a sample size of four is needed. Gentle et al. (37) performed a similar experiment, testing the PCR step by use of purified PCR product as target molecule, and concluded that a difference of 52% in target molecules can be detected between runs with a sample size of four. Quantitative RT-PCR has been used extensively over the last few years to evaluate hepatitis C virus RNA and HIV-2 RNA load and expression of minimal residual disease (38)(39)(40). The vast majority of reports conclude that despite variations, RT-PCR enables evaluation of the diseases in question.
Finally, we examined the PCR amplification efficiencies of the housekeeping gene ß-actin and TF to allow for normalization of variations in mRNA isolation and RT-PCR efficiencies. The mean slope and PCR efficiency were -3.50 and 1.95 for ß-actin cDNA and -3.43 and 1.93 for TF (difference not significant; Fig. 2
). In our system, ß-actin may be used to normalize for such variations in the range studied without correcting for differences in efficiency. As pointed out, corrections for amplification efficiencies may be made. It should be noted, however, that for every new target gene to be quantified, the amplification efficiencies have to be estimated to qualify for such adjustments.
|
In conclusion, we have developed reproducible PCR-based calibration curves for TF and ß-actin mRNA quantification. The reproducibility of the PCR amplicon-based calibration curves indicates no need to insert the PCR product into a vector to generate calibration curves. If suitable software is available, the reproducibility of the calibration curves indicates that they may be stored between runs, provided that a reference sample is at hand to assure the quality. Differences in PCR efficiencies between targets and calibrators can be compensated with use of suitable software. The evaluation of the present RT-PCR system showed reproducible results that can detect twofold concentration differences. Finally, the present system may be used to produce calibration curves suitable for quantification of any mRNA.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
, and IL-6 in peripheral blood mononuclear cells from atopic dermatitis. Ann Allergy Asthma Immunol 2000;84:553-558.[ISI][Medline]
[Order article via Infotrieve]
and IL-6. J Endotoxin Res 1995;2:27-35.
-synthesis in whole blood. Thromb Haemost 2000;83:868-873.[ISI][Medline]
[Order article via Infotrieve]
a mRNA using real-time reverse-transcription polymerase chain reaction. J Mol Diagn 2001;3:141-149.The following articles in journals at HighWire Press have cited this article:
![]() |
L. Struijk, E. van der Meijden, S. Kazem, J. ter Schegget, F. R. de Gruijl, R. D. M. Steenbergen, and M. C. W. Feltkamp Specific betapapillomaviruses associated with squamous cell carcinoma of the skin inhibit UVB-induced apoptosis of primary human keratinocytes J. Gen. Virol., September 1, 2008; 89(9): 2303 - 2314. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Woelfelschneider, O. Popanda, C. Lilla, J. Linseisen, C. Mayer, O. Celebi, J. Debus, H. Bartsch, J. Chang-Claude, and P. Schmezer A distinct ERCC1 haplotype is associated with mRNA expression levels in prostate cancer patients Carcinogenesis, September 1, 2008; 29(9): 1758 - 1764. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Ovstebo, O. K. Olstad, B. Brusletto, A. S. Moller, A. Aase, K. B. F. Haug, P. Brandtzaeg, and P. Kierulf Identification of Genes Particularly Sensitive to Lipopolysaccharide (LPS) in Human Monocytes Induced by Wild-Type versus LPS-Deficient Neisseria meningitidis Strains Infect. Immun., June 1, 2008; 76(6): 2685 - 2695. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Krop, A. Marz, H. Carlsson, X. Li, N. Bloushtain-Qimron, M. Hu, R. Gelman, M. S. Sabel, S. Schnitt, S. Ramaswamy, et al. A Putative Role for Psoriasin in Breast Tumor Progression Cancer Res., December 15, 2005; 65(24): 11326 - 11334. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Zygalaki, A. Stathopoulou, C. Kroupis, L. Kaklamanis, Z. Kyriakides, D. Kremastinos, and E. S. Lianidou Real-Time Reverse Transcription-PCR Quantification of Vascular Endothelial Growth Factor Splice Variants Clin. Chem., August 1, 2005; 51(8): 1518 - 1520. [Full Text] [PDF] |
||||
![]() |
A.-K. Fjeldheim, P. I. Hovring, O.-P. Loseth, P. W. Johansen, J. C Glover, V. Matre, O. K. Olstad, S. Reppe, J. O Gordeladze, S I. Walaas, et al. Thyrotrophin-releasing hormone receptor 1 and prothyrotrophin-releasing hormone mRNA expression in the central nervous system are regulated by suckling in lactating rats Eur. J. Endocrinol., May 1, 2005; 152(5): 791 - 803. [Abstract] [Full Text] [PDF] |
||||
![]() |
P de Cremoux, I Bieche, C Tran-Perennou, S Vignaud, E Boudou, B Asselain, R Lidereau, H Magdelenat, V Becette, B Sigal-Zafrani, et al. Inter-laboratory quality control for hormone-dependent gene expression in human breast tumors using real-time reverse transcription-polymerase chain reaction Endocr. Relat. Cancer, September 1, 2004; 11(3): 489 - 495. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Ovstebo, P. Brandtzaeg, B. Brusletto, K. B. F. Haug, K. Lande, E. A. Hoiby, and P. Kierulf Use of Robotized DNA Isolation and Real-Time PCR To Quantify and Identify Close Correlation between Levels of Neisseria meningitidis DNA and Lipopolysaccharides in Plasma and Cerebrospinal Fluid from Patients with Systemic Meningococcal Disease J. Clin. Microbiol., July 1, 2004; 42(7): 2980 - 2987. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.-S. W. Moller, R. Ovstebo, A.-B. Westvik, G. B. Joo, K.-B. F. Haug, and P. Kierulf Effects of bacterial cell wall components (PAMPs) on the expression of monocyte chemoattractant protein-1 (MCP-1), macrophage inflammatory protein-1{alpha} (MIP-1{alpha}) and the chemokine receptor CCR2 by purified human blood monocytes Innate Immunity, December 1, 2003; 9(6): 349 - 360. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||