Technical Briefs |
1 Bioscientia GmbH, Konrad-Adenauer-Strasse 17, D-55218 Ingelheim, Germany
aauthor for correspondence: fax 49-6132-781-428, e-mail torsten.arndt{at}bioscientia.de
Increased total plasma or serum homocysteine is regarded as a risk factor for occlusive arterial or venous disease [reviewed in Refs. (1), (2)]. Homocysteine is measured by HPLC (3), immunoassays (4), or liquid chromatographymass spectrometry (5)(6)(7)(8)(9). We were faced with a large workload of
200 samples/day, plus high material costs for commercially available immuno- and HPLC assays and laborious sample pretreatment for our HPLC application. We thus wished to establish a high-throughput liquid chromatographytandem mass spectrometry (LC-MS/MS) method. We aimed at a 96-well plate format for sample pretreatment and autosampling without protein precipitation, centrifugation, or derivatization steps.
We pipetted 50 µL of NaF-plasma, calibrator, or quality-control sample into each well of a 96-well plate (well volume, 2.2-mL) and added to each well 1000 µL of 0.4 µmol/L homocystine-d8 (3,3,3',3',4,4,4',4'-d8-homocystine; Cambridge Isotope Laboratories). The plate was sealed with Parafilm (the tightness was tested with colored solutions) and mixed by vigorous shaking. Of this mixture, 50 µL was transferred to each well of a 96-well plate (well volume, 1.2 mL), diluted with 500 µL of dilute NaOH (50 µL of 6 mol/L NaOH in 500 mL of water), and reduced by the addition of 25 µL of 200 mmol/L dithiothreitol (threo-1,4-mercapto-2,3-butandiol; Aldrich). The plate was sealed (see above), shaken to mix the contents of the wells, left at room temperature for 15 min, and again shaken. We removed the Parafilm and placed the 96-well plate in the autosampler (10 °C) for overnight LC-MS/MS analysis. The sample dilution was 1:241.5 (
1:240 as stated throughout this report).
The LC-MS/MS system consisted of a Surveyor MS pump and Autosampler combined with a triple-quadrupole TSQ Quantum mass spectrometer (Thermo Electron). The autosampler injected 2 µL (
0.008 µL of the native plasma sample) into the system. The mobile phase for the chromatography was a mixture of 2 mL/L acetic acid and methanol (50:50 by volume), the flow rate was 0.2 mL/min, run time was 2 min, and the column was a ThermoHypersil BetaMaxAcid [100 x 2.1 mm (i.d.); 5-µm bead size] with a ThermoHypersil Drop-in Guard Cartridge precolumn (Thermo Electron). The mass spectrometer was operated in the electrospray atmospheric pressure ionization, positive ion, and single-reaction-monitoring mode. The collision energy was 12 eV. Ion transitions m/z 136
90 (homocysteine) and 140
94 (homocysteine-d4), both attributable to a neutral loss of formic acid (6), were assessed. The LC-Quan software (Thermo Electron) was used for calculating the homocysteine/homocysteine-d4 peak-area ratios and quantification of homocysteine. Calibration, quality control, and checks for carryover were performed daily. The details and typical chromatograms obtained by our method are given in the Data Supplement that accompanies the online version of this Technical Brief at http://www.clinchem.org/content/vol50/issue4/.
Passing and Bablok regression (10) was used for unbiased method comparison and linear regression for calculating the calibration functions. The significance of differences between mean and median homocysteine concentrations for different groups was tested by parametric ANOVA (means) and nonparametric KruskalWallis ANOVA (medians). All calculations were done with the Analyze-it software for Microsoft Excel (Analyze-it Software Ltd.). The significance level was set at 95%, or P = 0.05.
Calibration curves were constructed with aqueous homocysteine solutions and with homocysteine added to human plasma. Both procedures were found to be not statistically different (see the online Data Supplement). The validation of our procedure included comparison of results obtained with use of 1.2-mL glass vials and the 96-well plates, optimization of the plasma dilution, and evaluation of the stability of the pretreated plasma samples over 3 days. Passing and Bablok regression (1.2-mL glass vial vs 96-well plate format and 1:240 vs 1:570 plasma dilution), box-plot analysis for testing the effects of sample storage of up to 3 days after sample pretreatment, parametric ANOVA, and nonparametric KruskalWallis ANOVA did not detect significant differences between the corresponding sample pretreatment protocols. This confirmed the comparability of the 96-well plate sample pretreatment with the commonly used 1.2-mL glass vial format, of the 1:570 plasma dilution with the 1:240 plasma dilution protocol, and the stability of the pretreated plasma samples over (at least) 3 days (see the online Data Supplement).
The comparability of our LC-MS/MS method with HPLC (Immundiagnostics) was tested with 187 plasma samples. The Passing and Bablok correlation function (Fig. 1A
) was as follows: LC-MS/MS = 1.12 x HPLC + 0.41 µmol/L. The 95% confidence intervals were 1.0751.162 for the slope and -0.049 to 0.856 µmol/L for the intercept; the slope was thus statistically significantly different from 1. BlandAltman analysis (Fig. 1B
) yielded a difference of 1.81 (95% confidence interval, 1.582.04). Thus, our LC-MS/MS gave significantly higher (
1518%) homocysteine results than the HPLC method. Similar results, pointing to the need for a true homocysteine standard (11), were reported by Nelson et al.(8) and, for a comparison between LC-MS/MS and the Abbott IMx assay, by McCann et al.(9).
|
Significant reductions in material and personnel costs were reported by McCann et al. (9) when they used LC-MS/MS instead of HPLC for homocysteine measurement. This was confirmed by our calculations. Material and personnel costs were diminished by 90% and 60%, respectively. The analysis time was reduced to 2 min/analysis or 400 min/200 samples with our LC-MS/MS method, instead of 1215 min/analysis or 24003000 min/200 samples for HPLC. Earlier reports on homocysteine analysis by LC-MS/MS did not address HPLC column life or LC-MS/MS system maintenance, although both can be important parts of the total analysis costs. Our large sample dilution gave a HPLC column life of
10 000 analyses and LC-MS/MS maintenance intervals of
4 months. We consider this an important part of the analysis cost reduction. Most importantly, the 17- to 80-fold greater plasma dilution and the
37-fold smaller injection volume (
0.008 µL of native plasma by our method) compared with earlier reports (5)(6)(7)(8)(9) minimized the matrix load with each injection and thus the ion suppression, despite our exclusion of protein precipitation. A greater plasma dilution (1:570-fold) did not give increased homocysteine and homocysteine-d4 signals, confirming that there was no ion suppression with the 1:240-fold sample dilution protocol under routine conditions. Such an experiment might be a valuable alternative to the procedure described by Streit et al. (12) when analyte-free samples are not available.
Our method is specific, sensitive, reproducible, and accurate (see Table 1 of the online Data Supplement). A 96-well plate format sample pretreatment in combination with LC-MS/MS for homocysteine analysis has been described previously (7)(9). We consider the combination of this format with a large plasma dilution without deproteinization for high-throughput homocysteine analysis the most important aspect of our application.
Acknowledgments
We thank Dr. Hartmut Kirchherr (Labor Bremen) for scientific discussions.