Clinical Chemistry 51: 270, 2005; 10.1373/clinchem.2004.039370
This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Related Collections
Right arrow Clinical Immunology
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2005;51:270.)
© 2005 American Association for Clinical Chemistry, Inc.


Letters to the Editor

False-Negative Serum Protein Electrophoresis in a Sample with an IgM Monoclonal Protein by Semiautomated Gel Electrophoresis

Xavier Bossuyta and Godelieve Mariën

1 Laboratory Medicine, Immunology, University Hospitals Leuven, Leuven, Belgium

aAddress correspondence to this author at: Laboratory Medicine, Immunology, University Hospitals Leuven, Herestraat 49, B-3000 Leuven, Belgium. Fax 32-16-347931; e-mail Xavier.bossuyt{at}uz.kuleuven.ac.be.


To the Editor:

In a recent report, Keren et al. (1) described a case in which the Sebia semiautomated electrophoresis method failed to demonstrate proteinuria in a 50-fold concentrated urine sample despite the presence of >2 g/24 h total protein. The reason for this failure was attributed to the presence of debris associated with cellular elements that interfered with the wicking of the urine into the microporous membrane applicator. The manufacturer is aware of such problems and recommends centrifugation of turbid urine samples. In our laboratory, all urine samples are systematically centrifuged (10 min at 1124g), and to date, we have not encountered such a failure of the Sebia system.

We did, however, analyze a serum sample for which agarose electrophoresis with the Hydrasys analyzer (Sebia, Inc.), using Hydragel 15 HR gels (Sebia), missed a large (12 g/L determined on a Beckman-Coulter Immage nephelometer) IgM{kappa} monoclonal protein (2) (Fig. 1A ). Similarly, immunofixation with the Sebia system using Hydragel 4 IF (Sebia) gels failed to detect the monoclonal protein (data not shown). The monoclonal protein was detected by capillary zone electrophoresis (Fig. 1B ) and by manual agarose electrophoresis using the Beckman Paragon agarose electrophoresis system. Manual immunofixation (Paragon; Beckman-Coulter) was used to characterize the monoclonal protein (data not shown). We ascribed the failure of the Sebia semiautomated system to the use of plastic applicators to apply the serum and hypothesized that the IgM monoclonal protein remained adhered to the microporous membrane of the applicator and did not diffuse into the gel. Mercaptoethanol treatment [10 µL of a 1:10 dilution of mercaptoethanol (Merck cat. no. 15433) added to 100 µL of serum] of the sample to cleave the disulfide bonds in the IgM pentamer revealed the monoclonal protein with the semiautomated Hydrasys system (Fig. 1C ). The company recommends treating viscous or turbid serum samples (particularly those containing cryoglobulin or cryogel) with Fluidil (prod. no. 4587; Sebia). Such treatment (25 µL of Fluidil added to 75 µL of serum), however, did not resolve the problem. Sebia also recommends use of 10 g/L ß-mercaptoethanol in Fluidil for proteins that can polymerize, which would produce a "monoclonal fraction" in all immunofixed tracks. This, however, was not the case for the sample described. By contrast, the system failed to detect a monoclonal fraction.



View larger version (13K):
[in this window]
[in a new window]
 
Figure 1. Electropherograms of a sample with an IgM{kappa} monoclonal protein.

(A), agarose electrophoresis (Hydragel 15 HR) on the Hydrasys system (Sebia); (B), capillary zone electrophoresis (Paragon 2000; Beckman-Coulter); (C), agarose electrophoresis (Hydragel 15 HR) on the Hydrasys system (Sebia) after mercaptoethanol treatment (10 µL of 1:10 dilution of mercaptoethanol added to 100 µL of serum). The fractions in panels A and C are albumin, {alpha}1-antitrypsin, {alpha}-lipoprotein, {alpha}2-macroglobulin and haptoglobin, ß-lipoprotein, transferrin, C3, and the {gamma}-globulin fraction. The small deflection in the mid-gamma region is an artifact. The arrow indicates the monoclonal protein.

We conclude that the applicator used with the semiautomated electrophoresis system from Sebia may give problems with some rare serum samples that contain an IgM monoclonal protein. Additional techniques to evaluate the presence of monoclonal proteins should therefore be available, such as nephelometry, capillary zone electrophoresis, and treatment with 2-mercaptoethanol.


References

  1. Keren DF, Gulbranson R, Ebrom SJ. False-negative urine protein electrophoresis by semi-automated gel electrophoresis. Clin Chem 2004;50:933-934.[Free Full Text]
  2. Mariën G, Bossuyt X. Response to the comments of K. Day and J. Zakowski on "Clinical capillary zone electrophoresis of serum proteins: balancing high sensitivity and high specificity" [Letter]. Clin Chem 2003;49:1711-1712.[Free Full Text]




This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bossuyt, X.
Right arrow Articles by Mariën, G.
Related Collections
Right arrow Clinical Immunology
Right arrow Proteomics and Protein Markers
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS